Text version

USAMRIID Minimal Essential Characteristics (MEC) Requirements For Laboratory Imager Workstation

PURPOSE: This MEC has been composed to specify parameters for the purchase of a Sector S 600 Imager Workstation.

DESCRIPTION: This Imager Workstation will be used for the Molecular and Translational Sciences (MTS) high-throughput screening programs. This instrument has the capabilities that we need:

a. Must have Electrochemiluminescence detection.

b. Must be Stable, non-radioactive labels that could be easily conjugated to biological molecules.

c. Must have 96-well and 384-well plate capabilities with stacker or deck layout capability to handle multiple plates.

d. Must be Cooled Charge Couple Device (CCD) camera with telecentric lenses for rapid detection using all multi-array or multi-spot plates.

e. Must have Capability to simultaneously measure multiple analytes in the same well.

f. Must have Rapid read times (70 seconds/plate) .

g. Must be Highly sensitive imaging detection systems.

h. Must have Six-log dynamic range.

i. Must not have No fluidics.

j. Must be Robotic integration software.

k. Must have Custom data export for integration with external databases.

l. Must have 3 one year service maintenance option years from MSD to include maintenance and calibration.