JMMC-QAT 26 June 2018
STATEMENT OF WORK: Indefinite Delivery Indefinite Quantity (IDIQ) procurement of new complete replacement Perkin Elmer Model XRD 1620 AN20 CS CT Grade digital flat panel x-ray detectors. Replacement detectors are to support MCAAP radiographic inspection operations.
IDIQ contract will be of a three (3) year duration commencing from date of award.
Minimum guaranteed number of detectors for the three (3) year contract will be five (5) detectors each year. Estimated maximum number of detectors required for the three (3) year contract will be twenty (20) detectors total.
1. Scope of Work. Each new complete replacement Perkin Elmer Model XRD 1620
AN20 CS, CT grade digital x-ray detector will have an XRD-FG frame grabber, trigger cable, and XRD-EPS power supply. This is a Brand Name Only purchase. No substitutes allowed.
2. Requirements: Each digital flat panel detector will have the minimum, but not limited to:
a. One (1) Perkin Elmer XRD 1620 AN20 CS, CT grade 16 x 16 inch (41cm) flat panel detector.
1. Panel- Single glass substrate amorphous silicon active two dimensional photodiode array
2. Size- 672mm x 599mm x 44mm
3. Weight- approximately 25 kg.
4. Scintillator- DRZ High with 1mm Cu plate
5. Total Pixel Number- 2048 x 2048
6. Active Pixel Number- 2000 x 2000
7. Pitch- 200 micron
8. Diode capacity- 2.1 pF
9. Dark current- <1pA/Pixel
10. Charge amplifier- 16 x 128 channel ASIC
11. ADC- 16 bit
12. Feedback capacity (Gain)- 0.5 pf, 1 pF, 2 pf, 4 pF, 8 pF
13. Integration time (minimum)- 266.7 ms
14. Non-Linearity- <1% (10% to 90% FSR)
15. Frame Per Second – 3.75
16. Dynamic Range > 75 dB minimum
17. Response Non-Uniformity- <+/- 2% (10% to 90% of FSR)
18. Radiation Energy Range - 40 kV – 15MV
19. Image Lag < 8% 1st frame
b. One (1) XRD-EPS power supply. One (1) DC cable @ 50 ft. each. One (1) data connection cable @ 100 ft. each. One (1) AC cable @ 50 ft. each. One
(1) XRD-FG Frame Grabber.
c. One (1) factory calibration data sheet each. One (1) operating manual each.