Specifications

Specifications

The vapor sources require the following specifications for Mass Flow Controllers (MFC) and Absolute Pressure Controllers (APC):

The MFC shall have a 10 slpm full scale flow rate, and an uncertainty equal to or better than +/- (0.8% reading + 0.2 % full scale).

The APC shall have a full scale pressure of 30 psia, with an uncertainty equal to or better than +/- 0.25% full scale.

The APC shall operate by being plumbed into the branch of a tee located between the MFC and a flow restriction, and by venting flow from the MFC to achieve the set point pressure upstream of the APC.

MFC and APC shared specifications

- The MFC and APC shall have a remote display that can be panel mounted

- The remote display shall be less than or equal to 4 inches wide by 5 inches tall

- The remote display shall provide full instrument control from the front panel

- The remote display shall have a color display