Requirement/Justification
The APSL requirement is to procure a 250 gpm flow loop to allow the secondary and/or transfer laboratories to perform secondary liquid flow calibrations.
The 250 gpm prover must have the following specifications:
1) The flow loop must use a pump to circulate liquid in a closed path.
2) The flow loop must have a full scale flow rate of 250 gpm and a lowest flow rate of 0.2 gpm
3) The flow loop must control the fluid temperature and pressure
4) The flow loop must include software to read the reference and device under test flow rate.
Background and justification are given below that result in a 250 gpm flow loop manufactured by CompuFlow Solutions as the only material solution to the APSL requirement.