This document is a Sources Sought notice for Vaisala humidity and temperature equipment required by the National Oceanic and Atmospheric Administration (NOAA). NOAA's Global Monitoring Laboratory manages atmospheric baseline observatories and an atmospheric measurement network, and needs to maintain existing Vaisala instruments, including HMP155 humidity and temperature probes, PTB330 digital and analog barometers, MI70 handheld measurement indicators, PTB210 digital barometers, and SPH20 static pressure heads. The government is seeking sources that can provide these Vaisala products to integrate into their current systems and maintain long-term historical data records. The government requires delivery within 60 days after award, and the NAICS code is 334519 with a 600-employee size standard. Interested firms should provide a brief narrative and documentation supporting their capability and experience to meet the government's requirements.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Vaisala SPH20 Specs.pdf | ||
| Vaisala MI70 Specs.pdf | ||
| Vaisala HMP155 Specs.pdf | ||
| Vaisala PTB210 Specs.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
BAROCAP Digital Barometer
PTB330
For professional meteorology, aviation, and industrial users
Features
• Vaisala BAROCAP sensor
• Accurate measurement
• Excellent long-term stability
• Added reliability through redundancy
• Graphical trend display with 1-year history data
• Height and altitude corrected pressure (QFE, QNH)
• For aviation, professional meteorology, laboratories, and demanding industrial applications
• Corrosion-resistant IP65/IP66 housing, suitable for outdoor and marine environment
Vaisala BAROCAPâ Digital Barometer PTB330 is designed for a wide range of high-end atmospheric pressure measurements. The pressure measurement of PTB330 is based on the Vaisala silicon capacitive, absolute pressure sensor - the Vaisala BAROCAP sensor. It provides high measurement accuracy and excellent long-term stability.
Highly accurate The PTB330 series is highly accurate.
Class A barometers for the most demanding applications are fine-tuned and calibrated against a high-precision pressure calibrator. Class B barometers are adjusted and calibrated using an electronic working standard. All PTB330 barometers come with a traceable factory calibration certificate.
Reliability through redundancy According to your choice, PTB330 can incorporate 1, 2, or 3 BAROCAP sensors.
When 2 or 3 sensors are used, the barometer continuously compares the readings of the pressure sensors against one another and reports if they are within the set internal difference criteria.
This unique feature provides redundancy in pressure measurement.
Users also get a stable and reliable pressure reading at all times, as well as a pre-indication of when to service or recalibrate the barometer.
QNH and QFE PTB330 can be set to compensate for the QNH and QFE pressure used especially in aviation. The QNH represents the pressure reduced to sea level, based on the altitude and temperature of the observation site.
The QFE represents the height-corrected pressure of small differences in altitude, for example, the air pressure at the airfield elevation.
Graphical display PTB330 features a multilingual, graphical display allowing users to monitor measurement trends. PTB330 updates the graph automatically during measurement and it provides a 1-year measurement history. In addition to instant pressure, PTB330 provides the WMO pressure trend and tendency codes.
Applications PTB330 can be used successfully for aviation, professional meteorology, and for demanding industrial pressure measurement applications such as accurate laser interferometric measurement and exhaust gas analysis in engine test benches.
Technical data
Measurement performance
Property Class A Class B
Barometric pressure range 500 … 1100 hPA
Linearity 1) ±0.05 hPa ±0.10 hPa
Hysteresis 1) ±0.03 hPa ±0.03 hPa
Repeatability 1) ±0.03 hPa ±0.03 hPa
Calibration uncertainty 2) ±0.07 hPa ±0.15 hPa
Accuracy at +20 °C (+68 °F) 3)
±0.10 hPa ±0.20 hPa
Barometric pressure range 50 … 1100 hPA
Linearity 1) – ±0.20 hPa
Hysteresis 1) – ±0.08 hPa
Repeatability 1) – ±0.08 hPa
Calibration uncertainty 2) – ±0.15 hPa
Accuracy at +20 °C (+68 °F) 3)
– ±0.20 hPa
Temperature dependence 4)
500 … 1100 hPa ±0.1 hPa ±0.1 hPa
50 … 1100 hPa ±0.3 hPa ±0.3 hPa
Total accuracy −40 … +60 °C (−40 … +140 °F)
500 … 1100 hPa ±0.15 hPa ±0.25 hPa
50 … 1100 hPa – ±0.45 hPa
Long-term stability
500 … 1100 hPa ±0.1 hPa/year ±0.1 hPa/year
50 … 1100 hPa ±0.2 hPa/year ±0.2 hPa/year
Miscellaneous
Pressure units hPa, mbar, kPa, Pa inHg, mmH20, mmHg, torr, psia hPa, mbar, kPa, Pa inHg, mmH20, mmHg, torr, psia
Resolution 0.01 hPa 0.1 hPa
Settling time at startup (1 sensor)
4 s 3 s
Response time (1 sensor) 2 s 1 s
Acceleration sensitivity – Negligible
Maximum pressure limit – 5000 hPa absolute
Maximum measurement rate 5)
– 10 Hz
1) Defined as ±2 standard deviation limits of endpoint non-linearity, hysteresis, or repeatability error.
2) Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to international standards.
3) Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature.
4) Defined as ±2 standard deviation limits of temperature dependence over the operating temperature range.
5) For class A you need a longer averaging time or measurement interval.
Operating environment Pressure range 500 … 1100 hPa, 50 … 1100 hPa
Operating temperature −40 … +60 °C (−40 … +140 °F)
Operating temperature with local display
0 … +60 °C (+32 … +140 °F)
IP rating IP66 IP65 (NEMA4) with local display 1)
1) IP rating of PTB330AWS is IP40.
Mechanical specifications Pressure fitting Barbed fitting for 1/8-inch (inside diameter) tubing or quick connector with shutoff valve for 1/8‑inch hose
Pressure connector M5 (10‑32) internal thread
Housing material G AlSi10 Mg (DIN 1725)
Weight 1 … 1.5 kg (2.2 … 3.3 lb)
Inputs and outputs Supply voltage 10 … 35 V DC
Supply voltage sensitivity Negligible
Typical power consumption at +20 °C (+68 °F) (voltage at 24 V DC with 1 pressure sensor)
RS‑232: 25 mA RS‑485: 40 mA Output voltage Uout: 25 mA Output current Iout: 40 mA Display and backlight: +20 mA
Serial communication RS-232, RS‑485, RS‑422
Analog output (optional) Current output 0 … 20 mA, 4 … 20 mA
Voltage output 0 … 1 V, 0 … 5 V, 0 … 10 V
Accuracy at pressure range
500 … 1100 hPa 50 … 1100 hPa
At +20 °C (+68 °F) ±0.30 hPa ±0.40 hPa
At −40 … +60 °C (−40 … 140 °F)
±0.60 hPa ±0.75 hPa
Data transfer software MI70 Link Interface software requirements
Microsoftâ Windows OS Microsoftâ Excel
Accessories Modules
Relay module RELAY-1L
Temperature-compensated analog output module
AOUT-1T
Isolated RS-485 module RS485-1
Power supply module POWER-1
AC adapters for devices already equipped with an external AC adapter connector
AC adapter, EU MI70EUROADAPTER
AC adapter, USA MI70USADAPTER
AC adapter, UK MI70UKADAPTER
AC adapter, AUS MI70AUSDAPTER
Static pressure head
Static pressure head SPH10
Static pressure head with heating SPH20
Barometer mounting accessories
Junction box ASM211113
Wall mounting kit 214829
Installation kit for pole or pipeline 215108
Outdoor installation kit (weather shield)
215109
DIN rail clips with installation plate 215094
Panel mounting frame 216038
Connection cables
Connection cable for PTB330 and MI70 handheld meters for spot check or calibration and adjustment
211339
Service cables
USB-RJ45 serial connection cable 219685
D9-RJ45 serial connection cable 215005
Output cables for 8-pin connector
Connection cable 5 m with 8-pin M12 female, black
212142
Female connector 8-pin M12 with screw terminals
212416
Cable bushings
PTB330/220/PTU200 DC adapter and RS-232 cable for PC
213019
PTB330/PTB220/PTU200 DC adapter cable
213026
Others
Dust filter 237018SP
Barbed pressure fitting 1/8-inch 19498SP
Quick connector 1/8‑inch 220186
Compliance
Property Value
EU directives and regulations RoHS Directive (2011/65/EU) amended by 2015/863 EMC Directive (2014/30/EU) Low Voltage Directive (2014/35/EU), applies to units equipped with single-phase AC power supply Power-1
Electrical safety EN 61010-1:2010 + A1:2019, applies to units equipped with single-phase AC power supply Power-1
Electromagnetic compatibility (EMC) EN 61326‑1, industrial environment
EN 55011:2009 + A1:2010
Environmental EN IEC 63000:2018
183 [7.20]
119 [4.69]
143 [5.63]
77 [3.03] mm [in]
PTB330 dimensions
145 [5.71]
[4.72] mm [in]
141 [5.55]
75 [2.95]
PTB330AWS dimensions
Published by Vaisala | B210708EN-J © Vaisala Oyj 2023
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.
Any reproduction, transfer, distribution or storage of information contained in this document is strictly prohibited. All specifications — technical included — are subject to change without notice.
www.vaisala.com
File details come from the government source that posted it. Updated .