Vaisala MI70 Specs.pdf

PDF 5 MB Posted

Attached to
Vaisala Humidity and Temperature Equipment Federal contract opportunity
Solicitation number
145034240076SM
Issued by
Department of Commerce National Oceanic and Atmospheric Administration

About this file

This document is a Sources Sought Synopsis issued by the National Oceanic and Atmospheric Administration (NOAA) to identify potential qualified U.S. Domestic businesses, including small businesses, that can provide Vaisala humidity and temperature equipment to maintain the agency's atmospheric baseline observatories and measurement network.

The key requirements include Vaisala HMP155 humidity and temperature probes, PTB330 barometers, MI70 handheld measurement indicators, PTB210 barometers, and SPH20 static pressure heads. Detailed specifications are provided for each product. The intended use is to integrate the equipment into NOAA's existing systems to continue long-term historical records of atmospheric chemistry and meteorological data. Responses are due by August 2, 2024, and delivery is required within 60 days of award. The North American Industry Classification Code is 334519 with a 600-employee size standard. Interested firms must provide a brief narrative and supporting documentation demonstrating their capability and experience to meet the requirements.

View the file

Other files for this federal contract opportunity

Other files attached to Vaisala Humidity and Temperature Equipment, newest first.
File Type Posted
Vaisala PTB210 Specs.pdf PDF
Vaisala SPH20 Specs.pdf PDF
Vaisala PTB330 Specs.pdf PDF
Vaisala HMP155 Specs.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

MI70 Handheld Measurement Indicator

Features

• Wide selection of measurement probes – HM70, DM70, MM70, GM70

• Multi-probe operation - combine any two probes

• Can be connected to selected Vaisala's fixed transmitters

• Intuitive interface

• Nine language options

• Suitable for field checking of fixed instruments

Vaisala Handheld Measurement Indicator MI70 is a user-friendly indicator. When combined with a measurement probe it is ideal for field checking and calibration of Vaisala's fixed instruments.

MI70 has probes for measuring humidity and temperature, dewpoint, moisture in oil, and CO2. In addition, MI70 can be used as a display when connected to selected Vaisala transmitters.

Multi-Probe Operation The MI70 indicator has two cable ports by which a combination of any two probes (or transmitters) can be connected to it simultaneously. For example, a humidity and temperature probe could be connected to the one port, and a CO2 probe to the other. The indicator can display the measurements made by both probes, so humidity, temperature, and CO2 data could be displayed simultaneously.

Intuitive User Interface MI70 features a multilingual, menu-based user interface and a clear graphical LCD display. All in all three measurement parameters can be displayed and simultaneously logged in the memory. The display shows measurement trends graphically and numerically. Users can select the measurement parameters and units.

MI70's intuitive user interface operates with menus and keypads to change settings, logging, and display preferences.

Multiple Language Options There are nine language options in MI70:

Chinese, English, Finnish, French, German, Japanese, Russian, Spanish, and Swedish.

Optional Connection to PC The optional MI70 Link Windowsâ software and the USB connection cable form a practical tool for transferring logged data as well as real time measurement data from the MI70 to a PC

Users can select and display up to three measurement parameters at the same time. The graphic display shows the stabilization of measurement and process trends.

Technical Data

MI70 Measurement Indicator Operating Environment

Operating temperature -10 ... +40 °C (+14 … +104 °F)

Operating humidity 0 ... 100 % RH, non-condensing

Storage temperature -40 ... +70 °C (-40 ... +158 °F)

Inputs and Outputs

Max. no of probes 2

Power supply Rechargeable NiMH battery pack with AC adapter or 4xAA size alkalines, type IEC LR6

PC interface MI70 Link software with USB or serial port cable

Analog Output

Scale 0...1 VDC

Output resolution 0.6 mV

Accuracy 0.2 % full scale

Temperature dependence 0.002 %/˚C full scale

Minimum load resistor 10 kΩ to ground

Mechanical Specifications

Housing classification IP54

Housing materials ABS/PC blend

Weight 400 g

Compatibility

EMC compliance EN61326-1, Portable Equipment

Other

Menu languages English, Chinese, Spanish, Russian, French, Japanese, German, Swedish, Finnish

Display • LCD with backlight

• Graphic trend display of any parameter

• Character height up to 16 mm

Alarm Audible alarm function

Data logging capacity 2700 real time data points

Logging interval 1 s to 12 h

Logging duration 1 min ... memory full

Resolution 0.01 %RH, 0.01 °C/°F, 0.01 hPa,

0.01 aw, 10 ppm / 0.01 %CO2

Battery Operation Time Typical charging time 4 hours

Operation Times

Continuous use (one probe of HM70, DM70 and MM70 series)

48 h typical at +20 ˚C (68 °F)

Continuous use (GM70 handle) better than 8 h at +20 °C (+68 °F)

Continuous use (GM70 pump) better than 5 h at +20 °C (+68 °F)

Data logging use up to a month, depending on logging interval

Measurement Probes for the MI70 HM70 Series Humidity and Temperature Probes

HMP75(B) General purpose probe

HMP76(B) Long, stainless steel probe

HMP77(B) Small probe, with 5 m cable

DM70 Series Dewpoint Probes

DMP74A, DMP74B Pressure tight probes for dry air measurements

DMP74C Probe for SF6 gas measurements

MM70 Moisture in Oil Probe

MMP78 Pressure tight, stainless steel probe

GM70 Carbon Dioxide Probes

GMH70 Handle for probe for diffusion sampling

GM70PUMP Sampling pump for GM70

For full specifications of the measurement probes see relevant brochures/ User's Guides (use the relevant order form to define ordered probe)

Spare Parts and Accessories Carrying Cases

A weatherproof carrying case for HM70, DM70 and GM70 short probes

MI70CASE3

A weatherproof carrying case for HM70 and MM70 long probes

MI70CASE4

A soft carrying case for HM70, DM70 and GM70 short probes

MI70SOFTCASE

Transmitter Connection Cables Small probe, with 5 m cable

MMT/HMT330, HMT120/130, DMT340,

PTB330, PTU300 Series

211339

HMD60/70 Series HMA6070

DMT242 27160ZZ

DMT143, DMT143L, DMT152, MMT162

HMW90 Series, HMDW110 Series and GMW90 Series

219980SP

MMT/HMT310 Series, GMP343 DRW216050SP

GMT220, GMD20 GMA70

HMP155 221801

Software

MI70 Link software with USB cable 219687

MI70 Link software with serial port cable

MI70LINK

Other

Analog output cable 27168ZZ

10 m (32.81 ft) extension cable for probe

213107SP

Battery, NiMH 4.8 V 26755

Published by Vaisala | B210830EN-E © Vaisala 2017

All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. Any reproduction, transfer, distribution or storage of information contained in this document is strictly prohibited. All specifications — technical included — are subject to change without notice.www.vaisala.com

File details come from the government source that posted it. Updated .