Vaisala PTB210 Specs.pdf

PDF 1 MB Posted

Attached to
Vaisala Humidity and Temperature Equipment Federal contract opportunity
Solicitation number
145034240076SM
Issued by
Department of Commerce National Oceanic and Atmospheric Administration

About this file

This document is a Sources Sought Synopsis for the National Oceanic and Atmospheric Administration (NOAA) to identify potential qualified U.S. Domestic businesses, including small businesses, to provide Vaisala's humidity, temperature, barometric pressure, and related meteorological instrumentation. The NOAA, Oceanic and Atmospheric Research, Earth Systems Research Laboratories, Global Monitoring Laboratory manages atmospheric baseline observatories and a tall tower atmospheric measurement network that requires the Vaisala HMP155 probes, PTB330 barometers, MI70 handheld indicators, PTB210 barometers, and SPH20 static pressure heads to maintain long-term historical records of atmospheric chemistry and meteorological data. The government is seeking instrumentation that meets specific technical requirements detailed in the document, with delivery required within 60 days of award. Interested firms are requested to provide a brief narrative addressing their capability and experience, as well as company information, by August 2, 2024. This is a market research exercise and not a solicitation for bids or proposals.

View the file

Other files for this federal contract opportunity

Other files attached to Vaisala Humidity and Temperature Equipment, newest first.
File Type Posted
Vaisala SPH20 Specs.pdf PDF
Vaisala PTB330 Specs.pdf PDF
Vaisala MI70 Specs.pdf PDF
Vaisala HMP155 Specs.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

BAROCAPâ Digital Barometer

PTB210

Features

• 500 … 1100 hPa or 50 … 1100 hPa pressure ranges with serial output

• Different scalings between 500 … 1100 hPa with analog output

• Electronics housing IP65 protected against sprayed water

• Accurate and stable measurement

• Traceable calibration (certificate included)

Vaisala BAROCAPâ Digital Barometer PTB210 is a reliable outdoor barometer for harsh conditions.

For harsh environments PTB210 is ideal for outdoor installations and harsh environments. PTB210 is designed to operate in a wide temperature range, and the electronics housing provides IP65 (NEMA 4) standardized protection against sprayed water.

PTB210 is ideal for use in applications such as weather stations, data buoys, ships, airports, and agrology. It is also an excellent solution for monitoring barometric pressure in industrial equipment such as laser interferometers and engine test benches.

Several pressure ranges PTB210 is designed for various pressure ranges. It is available in 2 pressure ranges in 3 configurations:

• Serial output for 500 … 1100 hPa

• Serial output for 50 … 1100 hPa

• Analog output with different scalings between 500 … 1100 hPa

Accurate and stable measurement PTB210 is digitally adjusted and calibrated by using electronic working standards. A higher accuracy barometer, which is fine-tuned and calibrated against a high-precision pressure calibrator, is available for the 500 … 1100 hPa pressure range.

In addition, PTB210 integrates directly with Vaisala Static Pressure Head Series SPH10/20. This pairing offers accurate measurement in all wind conditions.

Vaisala BAROCAP technology PTB210 uses the Vaisala BAROCAP sensor, a silicon capacitive absolute pressure sensor developed by Vaisala for barometric pressure applications.

The Vaisala BAROCAP sensor provides excellent hysteresis and repeatability characteristics and outstanding temperature and long-term stability.

PTB210 is delivered with a traceable factory calibration certificate.

PTB210 paired with SPH10 static pressure head

Technical Data

Measurement performance Pressure range

Serial output 500 … 1100 hPa 50 … 1100 hPa

Analog output 500 … 1100 hPa 600 … 1060 hPa 800 … 1060 hPa 900 … 1100 hPa

Serial output, accuracy (hPa)

Pressure range 500 … 1100 50 … 1100

Class A Class B

Non-linearity 1) ± 0.10 ± 0.15 ± 0.20

Hysteresis 1) ± 0.05 ± 0.05 ± 0.10

Repeatability 1) ± 0.05 ± 0.05 ± 0.10

Calibration uncertainty 2)

± 0.07 ± 0.15 ± 0.20

Accuracy at +20 °C (+68 °F) 3)

± 0.15 ± 0.20 ± 0.35

Temperature dependency 4)

± 0.20 ± 0.20 ± 0.40

Total accuracy -40 … +60 °C (-40 … +140 °F) 3)

± 0.25 ± 0.30 ± 0.50

Long-term stability (hPa/ year)

± 0.10 ± 0.10 ± 0.20

Analog output, accuracy

Non-linearity 1) ± 0.20 hPa

Hysteresis 1) ± 0.05 hPa

Repeatability 1) ± 0.05 hPa

Calibration uncertainty 2) ± 0.15 hPa

Accuracy at +20 °C (+68 °F) 3) ± 0.30 hPa

Temperature dependency 4) ± 0.50 hPa

Total accuracy -40 … +60 °C (-40 … +140 °F) 3)

± 0.60 hPa

Long-term stability ± 0.10 hPa/year

1) Defined as the ±2 standard deviation limits of end point non-linearity, hysteresis error, or repeatability error.

2) Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to international standards.

3) Defined as the root sum of the squares (RSS) of end point non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature.

4) Defined as ±2 standard deviation limits of temperature dependence over the operating temperature range.

Operating environment Operating temperature −40 … +60 °C (−40 … +140 °F)

Operating humidity 0 … 100 %RH, non‑condensing

Mechanical specifications Housing material PC plastic

IP rating, electronics IP65 (NEMA 4)

IP rating, sensor IP53

Weight, sensor 110 g (3.9 oz)

Cable weight 28 g/m (1.0 oz)

Inputs and outputs Serial output

Shutdown ON/OFF

Settling time at startup 2 s

Serial I/O RS-232C RS-232C / TTL (optional) RS-485, non-isolated (optional)

Parity None, even, odd

Data bits 7, 8

Stop bits 1, 2

Baud rate 1200, 2400, 4800, 9600, 19200

Response time 1 s

Resolution 0.01 hPa (1 measurement/s)

0.03 hPa (10 measurements/s)

Current consumption, normal mode < 15 mA (factory setting)

Current consumption, shutdown mode 0.2 mA

Analog output

Outputs 0 … 5 V DC, 0 … 2.5 V DC (order specified)

Shutdown ON/OFF

Response time 500 ms

Resolution 300 μV

Measurement rate 3 measurements/s

Current consumption, normal mode < 8 mA

Current consumption, shutdown mode 0.2 mA

All models

Max. pressure 5 000 hPa absolute

Pressure connector M5 (10-32) internal thread

Pressure fitting Barbed fitting for 1/8 in I.D. tubing

Supply voltage (reverse polarity protected), with RS-232/TTL output

5 … 28 V DC

Supply voltage (reverse polarity protected), with RS-485 or analog output

8 … 18 V DC

Compliance Directives EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU)

EMC compatibility EN / IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirement;

Basic environment CISPR 32 / EN 55032, Class B

Compliance CE, FCC, UKCA

Published by Vaisala | B210942EN-E © 2021

All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners.

Any reproduction, transfer, distribution or storage of information contained in this document is strictly prohibited. All specifications — technical included — are subject to change without notice.

File details come from the government source that posted it. Updated .