This is a Statement of Work for the procurement of a CDS Analytical Pyroprobe 6200 pyrolyzer unit and related accessories to be coupled with a PerkinElmer GCMS 2400 system for molecular analysis of solid samples. The procurement requires purchase through authorized distributors only, with warranties and OEM installation, training, and support provided exclusively by CDS Analytical; third-party support is not acceptable.

The Pyroprobe 6200 must meet specific technical requirements including Drop-In Sample Chamber (DISC) technology, built-in analytical trap supporting PY-TRAP-GC workflows, programmable heating from ambient to 1300°C with ±0.1°C NIST-traceable temperature accuracy, Cryo Trap integration, automated 48-position autosampler compatibility, direct GC-MS interfacing with automated remote start capability, and compatibility with the PerkinElmer GC 2400 unit. The procurement includes fourteen specified accessories, including the 6204-5035 Autosampler Module, 6203-5025 Cryo Trap Option, manual probe option, connectors, calibration components, consumables such as quartz tubes and septa, and replacement parts. Factory-certified installation, calibration, and training must be provided exclusively by a CDS Analytical Technician. The equipment shall be delivered within 4 weeks of purchase order issuance, with installation, commissioning, and training at the end user site to occur shortly after delivery at a date scheduled by the end user.

View the file

Other files

File Type Posted
Tab 07 Capability Statement 7-16-26.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1. Objective/Requirements This procurement is for a CDS Analytical Pyroprobe 6200 pyrolyzer unit and related accessories. The Pyroprobe will be coupled to a PerkinElmer GCMS 2400 system to enable the molecular analysis of solid samples. The hardware purchase must include authorized distributors only, warranties, and OEM installation, training, and support.

Third-party support is not acceptable.

2. Characteristics, Scope, and Specs

The Pyroprobe shall meet the following requirements:

Drop-In Sample Chamber (DISC) technology for reproducible, contamination-free sample introduction Built-in analytical trap supporting PY-TRAP-GC workflows and reactant gas capability Programmable heating from ambient to 1300 °C with ±0.1 °C NIST-traceable temperature accuracy Cryo Trap integration for enhanced volatile compound trapping and analytical flexibility Automated 48-position autosampler compatibility for unattended operation Direct GC-MS interfacing with automated remote start capability Compatibility to PerkinElmer GC 2400 unit CDS software and system integration protocols Factory-certified installation, calibration, and training, provided exclusively by CDS

Analytical Technician The Pyroprobe 6200 Pyrolyzer shall include the following accessories:

a. 6204-5035 6000 Series Autosampler Module for 48 samples

b. 6203-5025 Cryo Trap Option

c. 6204-5053 6000 Series Pyroprobe Manual Probe Option

d. 6203-5069 Vent Free GC column to Mass Spectrometer Connector

e. 6201-5089 Pyroprobe Drop-in Sample Chamber (DISC) temperature calibration

f. 1010-0489 Manual flowmeter 6000 series

g. 6201-3004 Constricted quartz tubes for 6000 DISC and autosampler

h. 2180-0019 Top septum for DISC door/flap

i. 6201-3002 Bottom Kalrez Septum for DISC door/flap

j. 1001-0345 Quartz Wool

k. 1620-0417 Vespel Ferrule 1/8”

l. 9301-5055 1/8” SS trap, vocarb

m. 9301-5056 1/8” SS trap, tenax

n. 6203-5041 DISC Quartz chamber replacement

o. 1620-0508 Kalrez o-ring for DISC

3. Place of Performance N/A

4. Period of Performance Shall be delivered within 4 weeks of date of PO issue. The installation, commissioning, and training at the end user site shall occur shortly after the delivery at a date to be scheduled by the end user.