Statement of Work - Hologic 3D Quorum Imaging Workstation - RFI 36C26226Q0373.pdf
PDF 129 KB Posted
- Solicitation number
- 36C26226Q0373
About this file
This is a draft Statement of Work for the Veterans Health Administration (VHA) to procure and install a Hologic 3D Quorum Imaging Technology system or equivalent add-on to their current Hologic 3D Mammography unit. The system will be installed in the Mammography suite in Building 126, Room E126. The contractor must deliver the system within 60 calendar days after receipt of order, with work performed during duty hours of 8:00 AM to 12:00 PM on regular business days, excluding 12 recognized federal holidays. The contractor is responsible for providing technical assistance, development, and generation of final specifications in coordination with the government.
The system must meet specific technical requirements, including full compatibility with the existing Hologic 3 Dimension Mammography unit, the latest IT and security patches, and the capability to reconstruct high-resolution 3D data by combining six 1-millimeter images into a single 6-millimeter thick slice with 70-micron pixel resolution. The equipment must reduce file size to 50 percent or less for easier PACS storage and faster file transfer, identify and preserve clinically relevant regions of interest during reconstruction, and maintain a minimum of 70-micron pixel resolution. The contractor must submit an authorization letter or documentation from the Original Equipment Manufacturer dated within 90 calendar days confirming authorized dealer, distributor, or reseller status for the proposed equipment.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26226Q0373.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK (DRAFT)
Statement of Work is tentative and subject to change based on responses received for the RFI.
1. GENERAL SCOPE
The Veterans Health Administration (VHA) has a requirement to add Hologic 3D Quorum Imaging Technology system OR EQUAL to the current Hologic 3D Mammography unit to produce highest 3D resolution images in the fastest time possible. This upgrade will be installed in Mammography suite in Bldg. 126, Room E126.
2. PERFORMANCE REQUIREMENTS
The contractor shall provide and be responsible for the technical assistance, development, and generation for final specifications.
Deliver by 60 Calendar Days after Receipt of Order (ARO).
Duty hours will be 8am – 12pm. The Government recognized US holidays are: New Year’s Day, Martin Luther King, Jr. Day, Presidents Day, Memorial Day, Juneteenth, Independence Day, Labor Day, Columbus Day, Veterans Day, Thanksgiving Day and Christmas Day. If the holiday falls on Sunday, it is observed on Monday.
3. PRODUCT/SERIVCE INFO
The use of breast tomosynthesis in breast imaging has demonstrated an improvement in cancer detection and a reduction in false positives compared to digital mammography.
Challenges with the routine use of breast tomosynthesis has been the increase in the number of images that need to be reviewed by radiologists which increased the average interpretation time. 2/3 less images to read which yield in saving 1 hour of interpretation time per day. This can increase workload. Hologic 3DQuorum project OR EQUAL one 6-mm thick slices 3D image by combining (6) 1mm images into one image. This image has 70-micron pixel resolution and utilizes artificial intelligence and machine learning. In return, the radiologist will read and interpret less images which results a faster average interpretation time. Hologic 3D Quorum OR EQUAL will also reduce the file size, which reduces required space, for PACS storage and reduces high network bandwidth requirements.
4. OTHER REQUIRED DOCUMENTATION
4.1 Submit an authorization letter or other document(s) from Original Equipment Manufacturer dated within 90 calendar days stating you are an Authorized Dealer, Authorized Distributor or Authorized Reseller for the proposed equipment/system.
5. SALIENT CHARACTERISTICS
5.1. Must be fully compatible with our current Hologic 3 Dimension Mammography unit.
5.2. Must have latest IT and security patches
5.3. Must be able to reconstruct high resolution 3D data to produce overlap slices to 1 image.
5.4. Must be able to reduce file size to 50% or less to make it easier to store in PACS and faster file transfer.
5.5. Must be able to identify clinically relevant and to preserve regions of interest during reconstruction process.
5.6. Maintains a minimum of 70-micron pixel resolution on the reconstructed images.