This document is a Request for Quotation (RFQ) for a Nitrogen Gas Generator for Instrument Purge. The USDA, Agricultural Research Service (ARS), Coastal Plains Soil, Water & Plant Research Center requires a nitrogen gas generator that meets the following minimum technical requirements: integrated air compressor, 99.5% nitrogen purity with no water vapor or CO2, -50°C dew point, at least 10 L/min flow rate, 45 psig outlet pressure, and dimensions no greater than 20W x 25H x 37D inches. The solicitation number is 12405B24Q0259, and the NAICS code is 334516 - Analytical Laboratory Instrument Manufacturing. Quotes are due by 4:00 pm CST on August 2, 2024, and must include a quote on company letterhead and a completed FAR 52.212-3 Offeror Representations and Certifications. Award will be based on best value, with technical capability and past performance being more important than price. Payment will be made by electronic funds transfer.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Clauses and Provisions (supply Solicitation).pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Requirements Project: Nitrogen gas generator for instrument purge
GENERAL INFORMATION
1.0 Scope of Work: Delivery of a nitrogen gas generator for instrument purge.
2.0 Background: Benchtop instruments with hygroscopic components are more prone to malfunctions and may require more frequent servicing. Purging these instruments with nitrogen gas reduces the likelihood of failure due to humidity, minimizes necessary maintenance, and is expected to increase the overall life expectancy of the device.
CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
Minimum Requirements:
1) Integrated air compressor
2) Deliver nitrogen at 99.5% purity, with no water vapor or CO2
3) Deliver nitrogen with dew point of -50°C
4) Flow rate of at least 10 L/min
5) Outlet pressure of 45 psig
6) Must be compatible with wall outlet (110-120 V, 60 Hz)
7) Dimensions no greater than (inches) 20W x 25H x 37D
4.0 Government Furnished: Not applicable.
5.0 Deliverables/Schedule: Package must be delivered to USDA-ARS, Coastal Plains Soil, Water, & Plant Research Center, 2611 West Lucas Street, Florence, SC 29501.