This document is a Specifications document related to a federal contract opportunity for a Real Time PCR instrument. The key details are:

The required minimum specifications for the Real Time PCR instrument include a 384-well block format with interchangeable blocks, ability to set up and monitor experiments without a computer, maximum dimensions to fit the available lab space, remote monitoring capabilities, ability to perform multiplexing reactions of at least 5 targets, compatible analysis software for PC and Mac, and specific temperature and ramp rate requirements.

The associated federal contract opportunity is Solicitation Number 12905B24Q0101 for "Real Time PCR for Pullman WA", issued as a Request for Quotation (RFQ) by the Department of Agriculture Agricultural Research Service. The delivery location is 410 SE Dairy Road Vogl 325 Pullman, WA 99164, with a required delivery time of 60 days after award. The solicitation is unrestricted with the NAICS code 334516 and a small business size standard of 1,000 employees. Offers are due by 05/16/2024 at 12:00pm PST via email to Brian.Fien@usda.gov.

View the file

Other files

File Type Posted
Terms and Conditions.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Required Minimum Specifications:

Delivery to: 410 SE Dairy Road

Vogl 325
Pullman, WA 99164

· Power source: 120V

· 384-well block format

· Interchangeable blocks to accommodate 96-well (0.2 mL) plates, 96-well (0.1 mL) plates, and TaqMan array cards.

· Must be able to set up , start, and monitor experiments and run updates without use of an attached computer

· Maximum Dimensions 112.5 x 90.7 x 74.2cm (H x W x D) ) (to accommodate space available in lab)

· Remote monitoring capabilities

· The ability to perform multiplexing reactions of at least 5 targets

· Analysis software that is compatible with PC and Mac operating systems

· Compatible dyes: FAM™, SYBR™ Green, VIC™, JOE™, TET™, ABY™, NED™, TAMRA™, Cy®3, JUN™, ROX™, Texas Red™, Mustang Purple™, Cy®5, LIZ™, Cy®5.5

· Excitation/emission range 450–680 nm/500–730 nm

· Average sample ramp rate should be at least 3.66°C/sec or faster

· Temperature uniformity should be at least ±0.4°C

· Temperature range should be at least 4–95.0°C

· Compatible with automation