Specification
SPEC. QTY
FORUMSPEC
****CHARACTERISTICS OF FORUM DATA MANAGEMENT****
Image and Data Management System must be FDA 510k approved and listed as a Medical Device.
Image and Data Management System must provide “Post Processing Analysis, Dynamic Review/Manipulation of RAW data” functionality for Cirrus HD-OCT. If Glaucoma Workplace is procured you can do with the HFA Instruments as well.
Image and Data Management System must be able to store and access raw instrument data from Field Analyzer, Cirrus HD-OCT, Visucam Fundus Camera and FF450/Visupac Fundus Camera systems.
Image and Data Management System must have functionality to enable real-time and dynamic editing/manipulation of all prior Visual Field Test data for all Glaucoma patients and allow for Baseline changes, real-time report generation for any VFI date, display RealEye images and data from the new HFA 3 instruments.
Image and Data Management System must support DICOM standards and Work-List functionality in order to create and send patient demographics to a specific or all instrument’s via a DICOM work-list within the eye clinic’s network.
Combined Report: HFA-Cirrus reports for Structure Function analyses in Single Display with OD-OS on one (1) page print out from data of Cirrus HD OCT and a Humphrey Visucal Field. If Combined Report is procured as option of FORUM.
Image and Data Management System must have Bi-directional DICOM capabilities to send both DICOM Modality Work-lists containing Consult/Order details to instrument’s within the eye-clinics network and receive back the diagnostic test results and RAW data for storage within the Image and Data Management System for access.
Image and Data Management System must connect to any brand of non-DICOM diagnostic instrument with export capabilities, and convert the diagnostic test results into a DICOM compatible file containing all patient demographical information, UID’s, Accession number, Device name and type, laterality and other key pieces of data with the least amount of user interaction.
This must be then stored within the Image and Data Management System and also forwarded to the NAVY ECIA (Enterprise Clinical Imaging Archive)
QTY
Image and Data Management System must not require the use of an external PC or import computer for a DICOM compatible modality.
Image and Data Management System must be able to download all past RAW data from one or more Field Analyzers, Cirrus HD-OCT’s, Visucam Fundus Camera and FF450/Visupac Fundus Cameras acquired within the main clinic and all satellite clinics upon installation of the Image and Data Management System.
Image and Data Management System must be able to display Field Analyzer Glaucoma Progression Analysis (GPA) and enable the clinician to manipulate data and store baseline changes at the workstation within the Image and Data Management System if Glaucoma Workplace is procured as a part of FORUM
Image and Data Management System must be able to display and allow the clinician to scroll through the various Cirrus Macula Cube B-scan slices, within in the Image and Data Management System viewer.
Image and Data Management System must be able to display images in pre-formatted clinical display groupings for Glaucoma, Retina and Cataract cases.
000000-2131-312/01FOBP42
FORUM BASIC LICENSE PACKAGE (V.4.2) | 1/EA
306601-8530-111/30FSFOL3
SCHEDULED, FORWARDING LICENSE (V3)
306601-8530-110/30FMSAL3
MULTI-SITE ACCESS LICENSE (V3)
306601-8530-100/20FVIEW3
FORUM VIEWER (V3) | 1/EA
000000-2259-029/20GW31US
GLAUCOMA WORKPLACE 3.1
000000-2020-237/30FGWPU3
LICENSE FORUM GLAUCOMA WORKPLACE 3.1 | 1/EA
000000-2259-031/30RW201U
RETINA WORKPLACE 2.0.1
000000-2127-546/30FRWPUL
RETINA WORKPLACE USER LICENSE X1 | 1/EA
306601-8530-101/40FCZDIC
FORUM-DICOM INTERFACE LICENSE TO CZM INSTRUMENT (Vx.xx)
306601-8530-103/40F3PDL3
FORUM INTERFACE LICENSE TO 3RD PARTY DICOM INSTRUMENT (V3)
000000-2076-772-PRO
UPGRADE CIRRUS INSTRUMENT REVIEW SOFTWARE TO CIRRUS
REVIEW SOFTWARE 8.1 LOCAL DATABASE (INCLUDES RPE AND GC
ANALYSIS) (10-PACK) WITH REMOTE INSTALLATION | EA
266002-1131-909-PRO
FORUM ON-SITE INSTALLATION & TRAINING
266002-1150-893-PRO
Enterprise Architecture Design (per hour)
266002-1130-065-PRO
CONNECTIVITY SERVICE PLAN, ANNUAL
3PWRNTY
WARRANTY