TECH SPECS.SOW - Keithley DAQ6510 9.17.24.pdf

PDF 283 KB Posted

Attached to
Precision Data Acquisition Systems (Solicitation) (Keithley P/N DAQ6510/7700) Federal contract opportunity
Solicitation number
SP4706-24-Q-0040
Issued by
Defense Logistics Agency

About this file

This document is a Technical Specifications document for a Precision Data Acquisition System solicitation. It specifies the requirements for four (4) Keithley DAQ6510/7700 datalog scanners that the Defense Logistics Agency (DLA) is seeking to procure for the Defense Supply Center Columbus (DSCC) in Columbus, Ohio. The system must have 40 isolated universal input channels, the ability to measure thermocouples, thermistors, DC voltage, DC current, and resistance, 6-digit display resolution, internal data storage, and two independent alarms per channel, among other technical specifications. The solicitation number is SP4706-24-Q-0040, and quotes should be submitted directly to Blair.Hartley@dla.mil using the subject "Data Acquisition Systems".

View the file

Other files for this federal contract opportunity

Other files attached to Precision Data Acquisition Systems (Solicitation) (Keithley P/N DAQ6510/7700), newest first.
File Type Posted
SOW.Specs - Fluke 8.27.24.pdf PDF
SP4706-24-Q-0040.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Rev 0

Technical Specifications Precision Data Acquisition Systems

DLA Land and Maritime Engineering & Technical Support Directorate Product Test Center Division, Electronics Lab, VTP

I. The contractor shall supply four (4) Multichannel Data Acquisition Systems to the Defense Supply Center Columbus (DSCC), DLA Land and Maritime, Product Test Center Division, Electronics Lab, Columbus, Ohio.

A. Keithley DAQ6510/7700 datalog scanners (DAQ6510 with two 7700 20-channel modules installed on each DAQ6510 mainframe), with the following salient characteristics:

1. Shall be new, not refurbished.

2. Shall have 40 isolated universal input channels.

3. Shall have the capability to measure thermocouples, thermistors, DC voltage, DC current, and Resistance.

4. Shall be configured with Internal High-Capacity Module.

5. Shall have 6 ½ digit display resolution for voltage, current, and resistance.

6. Shall have internal data storage with ability to store approximately 7 million readings.

7. Shall have minimum of two independent, user defined alarms per channel.

8. Voltage Input Requirement: 120V (108V to 132V), 60 Hz.

9. Operating Temperature Range: 0C to 50C.

10. Channel Capacity:

a) Analog Input Channels: 40.

b) Voltage/Resistance Channels: 40.

c) Current Channels: 4.

d) Totalizer: 1.

e) Alarm Outputs: 6.

f) Trigger Input: 1.

11. Connectivity:

a) USB Host Port – Type A, FAT32, 32GB Capacity.

b) USB Device Port – Type B, Command Protocol – SCPI.

c) LAN – Ethernet 10/100, TCP/IP, Command Protocol – SCPI.

12. Dimensions:

a) Height: 107.2 mm.

b) Width: 224 mm.

c) Depth: 387.4 mm.

13. Measurement Capabilities:

a) Thermocouple:

1) Temperature Range: -200 C to 1820 C.

2) Temperature Coefficient: 0.03 °C/°C.

3) Accuracy: 0.20 C.

b) DC Voltage:

1) 1000 V.

Rev 0

Technical Specifications Precision Data Acquisition Systems

DLA Land and Maritime Engineering & Technical Support Directorate Product Test Center Division, Electronics Lab, VTP

2) A/D Linearity: 0.0001% of measurement of 10V range.

3) Input Bias Current: <50 pA at 23C.

4) Accuracy in +/- 100 mV Range: 0.0015% + 0.0030%.

5) Accuracy in +/- 1V Range: 0.0015% + 0.0006%.

6) Accuracy in +/- 10V Range: 0.0010% + 0.0004%.

7) Accuracy in +/- 100V Range: 0.0015% + 0.0006%.

8) Accuracy in +/- 1000V Range: 0.0020% + 0.0006%.

c) Digital I/O:

1) Absolute Voltage Range: -0.25 V to 5.25 V.

2) Input Minimum Logic Low: 3.7 V.

3) Input Maximum Logic High: 0.7 V.

4) Maximum Sink Current: -50 mA.

5) Maximum Source Current: 2 mA.

14. High-Capacity Input Module Specifications:

a) Maximum Input: 300 V.

b) Offset Current: < 100 pA.

c) Contact Resistance: <1Ω.

d) Contact Life: 108 operations.

e) Contact Potential: <±500nV.

f) Automatic CJC.

B. Other Requirements:

1. Cost for Shipping shall be included.

2. Shall have Calibration Certificate valid for a minimum of 1 year from delivery date.

3. Contractor shall contact PTC POC (614-692-9893, larry.baas@dla.mil) or (614-692-

9862, eric.kelly@dla.mil) to schedule the delivery.

4. The system shall be delivered to:

Defense Supply Center Columbus (DSCC) 300 North James Road, Bldg 17-3 Mark For: DLA Product Test Center, Building 11-7 (VTP Lab) Attn: Larry Baas (614-692-9893; Larry.Baas@dla.mil) Columbus, Ohio 43218-3990

5. Payment of invoices will be accomplished by payment through the Defense Finance and Accounting Service. Invoices will be submitted to the Wide Area Workflow system in accordance with DFARS 252.232-7003 Electronic Submission of Payment Requests and Receiving Reports and DFARS 252.232-7006 Wide Area Workflow Payment Instructions within 10 workdays after the conclusion of work performed.

Invoices shall be submitted through Wide Area Workflow (WAWF), See DFARS Clause 252.232-7006 Wide Area Workflow Payment Instructions for detailed instructions on how to submit invoices.

6. DODAAC: SL0700.

STATEMENT OF WORK

September 17, 2024

1. Title

Precision Data Acquisition Systems with High-Capacity Modules

Requiring Activity: DLA Product Test Center Division, Electronics Lab, located at Cols, Ohio.

Contracting Activity: DLA Contracting Services located at Cols, Ohio.

2. Introduction The Electronics Product Test Center (EPTC) is looking to acquire four precision data acquisitions systems (Keithley DAQ6510/7700 with two 7700 High-Capacity module is considered a single system), with a minimum of 40 measurement channels. These systems, along with the current system, will enhance testing capabilities within the EPTC. The current system is used for battery testing in one location. Adding data acquisition systems provide testing support for several areas within the EPTC. These new systems will also support the expanding battery test lab.

3. Background The current data acquisition system is used to measure the temperature at which the thermal overloads trip on several batteries in a single testing setup. These systems must be accurate to ensure the safe configuration of Lithium based batteries. Currently, only one system exists within the EPTC. By adding these 4 systems, it will enhance readiness to test.

During calibration, the system is out of the lab for up to a month at a time. Adding systems will allow for a staggered calibration cycle. These systems also support testing throughout the lab for components such as connectors, switches, and other types of electronics. These systems provide automated precision data capture, increasing the reliability of testing results.

4. Scope Four (4) Precision Data Acquisition Systems with High-Capacity testing channels.

5. Applicable Directives

None.

6. Performance Requirements The Contractor shall supply four Precision Data Acquisition Systems with High-Capacity testing channels, with calibration certificates valid for one year from delivery date.

June 27th, 2024

7. Deliverables

• Four Precision Data Acquisitions systems with High-Capacity Input testing channels.

• Calibration Certificates for each system.

8. Contractor Information: please list what type of Experience, Past Performance, Certifications, and/or Scheduling Needs that are required from the Contractor

None.

9. Place of Performance

DSCC Columbus 3990 East Broad Street Whitehall, OH 43213

ATTN: Larry Baas Work to be performed will take place at:

Building 11, Section 07 & 08.

POC: Larry Baas, 614-692-9893.

10. Travel (if applicable)

None.

11. Invoicing Payment of invoices will be accomplished by payment through the Defense Finance and Accounting Service. Invoices will be submitted to the Wide Area Workflow system in accordance with DFARS 252.232-7003 Electronic Submission of Payment Requests and Receiving Reports and DFARS 252.232-7006 Wide Area Workflow Payment Instructions within 10 workdays after the conclusion of each training session.

11. WAWF Acceptance DODAAC#

SL0700.

11. Point of Contacts

The Government Contract Officer:

TBD

June 27th, 2024

The Government Contract Officer Representative (COR):

Kyle Palmer Kyle.Palmer@dla.mil 614-692-8860

DLA Land and Maritime Product Test Center, Electronics Office 3990 E. Broad St., Columbus, OH Bldg. 11, Section 07 Larry Baas, 614-692-9862

The Government Technical Representative/Project Manager:

Name: Larry Baas Phone Number: (614) 692-9893 Email Address: larry.baas@dla.mil.

I. The contractor shall supply four (4) Multichannel Data Acquisition Systems to the Defense Supply Center Columbus (DSCC), DLA Land and Maritime, Product Test Center Division, Electronics Lab, Columbus, Ohio.

File details come from the government source that posted it. Updated .