Ozone TechSpecsSOW rev2.pdf
PDF 133 KB Posted
- Attached to
- Presolicitation Synopsis for Ozone Testing Chamber Federal contract opportunity
- Solicitation number
- SP470624Q0016
- Issued by
- Defense Logistics Agency
About this file
This document is a Technical Specifications Statement of Work for an EIA-364-14 compliant Ozone Chamber required by the Defense Logistics Agency (DLA) Product Test Center in Columbus, Ohio. The key details are:
The contractor shall provide a new, stainless steel Ozone Chamber with a 5 ft3 test volume, ozone exposure range of 0-1000 ppm, and ability to control temperature from ambient to 70°C ±5°C. The chamber must meet EIA-364-14B standards, have a plexiglass window, integrated air dryer, and onboard data logging. Required accessories include a Windows 11 PC with software to control the chamber and 8 hours of on-site training. The instrument must be delivered within 30 days of award to the DLA Product Test Center in Columbus, Ohio. All shipping costs are included in the contracted price. Invoices shall be submitted through the Wide Area Workflow (WAWF) system. The government points of contact are the Contract Officer Representative Kyle Palmer and the Technical Representative Larry Baas.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Rev 04/12/2024
Technical Specifications
Ozone Chamber
DLA Product Test Center, VTP
I. The contractor shall provide all labor, materials, tools, transportation, supplies, supervision, etc. (except as noted), necessary to provide all components for an EIA-364-14 compliant Ozone Chamber to the Defense Supply Center Columbus (DSCC), Columbus, Ohio. The salient characteristics of the EIA-364-14 compliant Ozone Chamber are as listed below:
A. EIA-364-14B compliant Ozone Chamber with support equipment and with the following salient characteristics:
1. Shall be new, not refurbished.
2. Shall be supplied with Operating manual.
3. Shall be powered by standard US 120V, 15A AC power outlet.
4. Shall be compliant to perform testing described in accordance with EIA-364-14B.
5. Shall be constructed of stainless steel.
6. Shall have plexiglass window for easy viewing.
7. Shall have an ozone exposure range of 0-1000 ppm.
8. Shall have a chamber test volume of 5 ft3.
9. Shall include integrated air dryer.
10. Shall have means to control temperature from ambient to 70 °C ±5 °C.
11. Shall have means to remove exhaust air, containing ozone, and vent it through ductless self-contained filtered vented system.
12. Shall provide adequate internal circulation with velocity of at least 2 ft/sec.
13. Shall not use silent arc discharge.
14. Shall provide ozone measuring devices capable of measuring the concentration within a tolerance of ±3%.
15. Shall have on-board data logging with ability to transfer data to standard external source such as USB flash drive or SD card.
16. Shall be mounted on wheeled cart.
17. Optional extended warranty and multi-year service contract must be offered.
18. Shall have: NIST Traceable Calibration Certificate and ISO 17025 accredited calibration certificate with data and uncertainties
19. Shall at a minimum have: Included Accessories: All necessary electrical cables, communication cables, and hoses to install and operate machine.
B. Required Upgrades and Accessories - Brand Name or Equal:
1. Windows 11 PC with all necessary peripherals, hardware, ports or on-board controller, with software and permanent license to reliably control ozone chamber.
2. 8 hours of on-site training and setup.
C. Place of Performance
The instrument shall be delivered to:
Defense Supply Center Columbus
300 North James Road, Bldg 17-3
Mark for: DLA Product Test Center, Bldg 11-7 (TE Lab)
Attn: Eric Kelly (eric.kelly@dla.mil / 614-692-9862)
Columbus, Ohio 43213 mailto:eric.kelly@dla.mil
DLA Product Test Center, VTP
D. Travel (if applicable)
N/A
E. Invoicing
Payment of invoices will be accomplished by payment through the Defense Finance and Accounting
Service. Invoices will be submitted to the Wide Area Workflow (WAWF) system in accordance with
DFARS 252.232-7003 Electronic Submission of Payment Requests and Receiving Reports and DFARS
252.232-7006 Wide Area Workflow Payment Instructions within 10 workdays after the conclusion of each training session.
F. WAWF Acceptance DODAAC#
1. TA Lab is SL0700.
G. Other Requirements:
1. Delivery: 30 days from date of the award.
2. Contractor shall give notification to Eric Kelly (eric.kelly@dla.mil /614-692-9862) when the instrument is ready to be delivered.
3. All shipping costs are to be included in the contracted price.
4. The instrument shall be delivered to:
Defense Supply Center Columbus 300 North James Road, Bldg 17-3 Mark for: DLA Product Test Center, Bldg 11-7 (TE Lab) Attn: Eric Kelly (eric.kelly@dla.mil / 614-692-9862) Columbus, Ohio 43213
5. Invoices shall be submitted through Wide Area Workflow (WAWF), See DFARS Clause 252.232-7006 Wide
Area Workflow Payment Instructions for detailed instructions on how to submit invoices.
H. Points of Contact The Government Contract Officer:
TBD
The Government Contract Officer Representative (COR):
Kyle Palmer Kyle.palmer@dla.mil
(614) 692-8860
DLA Land & Maritime Product Test Center, 3990 E. Broad St., Columbus OH, Bldg. 11, Section 7 Larry Baas (614) 692-9893
The Government Technical Representative Name: Larry Baas Phone #: (614) 692-9893 mailto:eric.kelly@dla.mil mailto:eric.kelly@dla.mil mailto:Kyle.palmer@dla.mil
DLA Product Test Center, VTP
Email: larry.baas@dla.mil
File details come from the government source that posted it. Updated .