This is a Statement of Work (SOW) for the procurement of a replacement residual gas analyzer (RGA) system for NASA Glenn Research Center's Small-scale Multipurpose Research Facility (SMiRF).
The objective is to replace an existing RGA system that suffers from software compatibility limitations preventing integration with current operating systems and data management requirements. The replacement system must provide continuous monitoring and characterization of gaseous environments within vacuum-based systems to diagnose leaks, detect contamination, monitor outgassing, and verify vacuum quality. The required deliverables include: one HAL 201 RC system with mass range to 200 amu and Faraday/Channeltron electron multiplier (Part No. RC 1.2 / 144100); one filament kit with three oxide-coated iridium twin filament assemblies and accessories (Part No. RS 6.1.4 / 201600); and one set of 15-meter RF/Controller and Multiplier Supply cables (Part No. RCB 7.1.15/RCB 7.3.15). Engineering requirements include atomic mass range of 1–200 AMU, detection limit ≤10⁻¹² Torr, electron multiplier and Faraday detector, operating pressure to at least 10⁻⁴ Torr, dynamic range ≥10⁷, UHV-compatible construction, connection port size ≤3-3/8" Conflat, and separated measurement head from power supply/controller. Software must include perpetual licensing with user-installable upgrades, single-seat capability with multiple terminal access, leak detection mode for individual gas species identification with operator alerts, and data export functionality in raw or engineering units. Additional requirements include a calibration report at delivery, vendor proof of existing product line with similar scope and technical requirements with client references, NASA facility engineer approval before procurement, detailed installation instructions for NASA technicians, and service and support for unit commissioning, training, and troubleshooting. Delivery must be to NASA Glenn Research Center at 21000 Brookpark Road, Cleveland, OH 44135, with acceptance by Jonathan Kubiak. The period of performance is 12 weeks as soon as possible (ARO).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SATPC0042184 Tab 10 Open Market RFQ.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The Small-scale Multipurpose Research Facility (SMiRF) operates vacuum-based systems that require continuous monitoring and characterization of the gaseous environment to ensure system performance, process reliability, and equipment integrity. Accurate identification and quantification of residual gases within these vacuum systems are critical for diagnosing leaks, detecting contamination, monitoring outgassing, and verifying vacuum quality during operations and maintenance activities.
The existing residual gas analyzer (RGA) is being replaced due to software compatibility limitations that prevent effective integration with current operating systems and data management requirements. These compatibility issues have reduced the system’s reliability restricted access to vendor support and software updates, and limited the facility’s ability to efficiently collect, analyze, and archive vacuum system data. Replacement of the RGA will restore full functionality, improve system maintainability, and ensure continued support for vacuum diagnostics and monitoring activities.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 RC 1.2 / 144100 HAL 201 RC system, with mass range to 200 amu and with Faraday/Channeltron electron multiplier.
2 RS 6.1.4 / 201600 Filament Kit. Includes:- 3x Oxide coated Iridium twin filament assembly, 1x jeweller's screwdriver, 1x pair of tweezers, 1x eyeglass and gloves All provided in a compact carrycase.
3 RCB 7.1.15/RCB 7.3.15 15m RF/Controller Cable (type 250100/15) and 15m Multiplier Supply Cable (type 270100/15).
Scope of Work Description
A spare filament and replacement kit is required as well.
Service and support for unit commissioning, training, and troubleshooting shall be included.
• Engineering Requirements:
o Atomic Mass Range: 1-200AMU o Detection Limit: ≤10^(-12) Torr o Electron Multiplier + Faraday Detector o Operating pressure to at least 10^(-4) Torr o Dynamic Range ≥ 10^7 o UHV-compatible construction o Leak-detection mode for single gas species.
o Connection port size ≤3-3/8” Conflat o Measurement head must be separated from power supply/controller. RGA cabinet is space limited and not environmentally controlled.
• Include spare filament and replacement kit.
o Software Requirements:
Including software component with perpetual and included licensing with necessary future upgrades made user installable.
Software tools can be single seat but allow for multiple computer terminal access.
Must include a leak detection mode that allows the operator a way to single out a particular gas species and alerts the operator to its presence.
Ability to export measurement data in raw or engineering units.
o Testing & Quality Assurance Calibration Report is required with delivery of hardware Vendor shall provide proof of existing product line with similar scope, size, complexity, and technical requirements. Such evidence shall include product description and client references. Vendor will need to be approved by NASA facility engineers before procurement can be made.
Delivery & Installation All materials must be delivered to NASA Glenn Research Center:
o 21000 Brookpark Road, Cleveland OH 44135 Detailed instructions on installation of hardware and software for NASA technicians to install
Government Acceptance
Jonathan Kubiak shall accept the procurement
3. Place of Performance
Glenn Research Center
4. Period of Performance
12 weeks ARO