KSC requires the purchase of one VICI Tras Gas Analyzer (OEM Part # T8442W-7T-PP1He) with External Deoxo Provision and one Valco Instruments TGA Install Service. The VICI analyzer is a complete chromatographic unit housed in a 7U chassis with a model 8000 Interface and control system configured for 110 VAC operation. The instrument features dual PDHID Pulsed Discharge Helium Ionization detectors, distributed control between an embedded PC and programmable logic controller, touch screen interface, and Data Apex Clarity chromatographic data acquisition software. The system includes a helium purifier, out-of-box methods for balance gases, and rear panel bulkhead fittings configured for optional external Deoxo trapping apparatus between the injector and precolumn. The analyzer includes methods for helium, hydrogen, argon, and nitrogen bulk gases. The installation service includes up to 2 days of onsite setup, training, and instruction (including 1 day travel), with the customer option to train in Houston, Texas during Factory Acceptance Testing at no additional cost. Additional installation time is billed at $250.00 per hour in 4-hour increments, with installation scheduled a minimum of 3 weeks in advance after delivery and confirmation of onsite utilities. Performance of work will take place at NASA Kennedy Space Center, BOSS Receiving, Kennedy Space Center, Florida 32899, with a period of performance of 20 weeks from receipt of order.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements The NASA Kennedy Space Center (KSC) requires the purchase of VICI Tras Gas Analyzer with External Deoxo Provision and Valco Instruments TGA Install Service which is required to maintain compatibility with the existing analytical infrastructure within the Kennedy Space Center laboratory. Procuring the same manufacturer and platform will ensure seamless integration with existing laboratory operations while minimizing implementation risk.
2. Characteristics, Scope, and Specs
Quantity OEM Part # Product Description
T8442W-7T-PP-
1He
VICI Tras Gas Analyzer with External Deoxo Provision
Description:
Complete chromatographic unit in a 7U chassis, with model 8000 Interface and control, configured for 110 VAC general service electrical area. The chromatograph will be equipped with valves, columns, and 2 PDHID Pulsed Discharge Helium Ionization detectors. The control is distributed between an embedded PC and an onboard programmable logic controller, that together automate the chromatograph and installed options. The PC includes a touch screen interface, all automation control software and chromatographic data acquisition software (Data Apex - Clarity).
Full integration provides total automation of the instrument operation, data collection and reporting for single, batch, and continuous operation. A Helium Purifier is included, along with full analytical testing. This package will provide out of the box methods for specified balance gases shown above.
Rear panel bulkhead fittings configured on the molecular sieve channel to provide optional use of Deoxo or other trapping apparatus. The trap provision to be configured between injector and precolumn. This quote does not provide the external trap.
This instrument includes helium, hydrogen, argon and nitrogen bulk gas methods.
1 TGA-Install
Valco Instruments TGA Install Service
Description:
Onsite training, if required, 2 days of setup and instruction at customer location, (includes 1 day travel). Customer option to train in Houston, TX during FAT at no cost. Additional time billed at $250.00/hr, in 4 hour increments. All installations should be scheduled a minimum of 3 weeks in advance after delivery and confirmation of onsite utilities.
3. Place of Performance NASA Kennedy Space Center (KSC) BOSS Receiving - M6-0744 Kennedy Space Center, FL 32899
4. Period of Performance 20 Weeks ARO