SATPC0041294 Tab 04 4 SOW_Redacted.pdf
PDF 117 KB Posted
- Solicitation number
- 80NSSC26934740Q-R1
About this file
This is a Statement of Work for NASA Glenn Research Center's procurement of Chell microDAQ3 Pressure Acquisition Systems and associated support hardware.
The requirement specifies procurement of three microDAQ3 64-channel pressure acquisition systems with varying scanner ranges: two units (64MD3-07010201) with 35 kPa scanner range, one unit (64MD3-09010201) with 103 kPa differential scanner range, and one unit (64MD3-11010201) with 300 kPa scanner range. All three systems include features such as absolute or differential measurement capability, 1 mm tubulations, TCP/UDP interfaces, IEEE 1588 PTPv2 time stamping, embedded web servers, DC power with hardware triggers, CAN output, 400 Hz per-channel maximum acquisition speed, and 0 to 90°C compensated operating range. Supporting hardware includes one 8-way interface (microDAQ-INT-8F), one 4-way interface (microDAQ-INT-4-02), five 3-meter interface cables (MD3-CAB-3), one 3-meter nanoDAQ/microDAQ3 cable (nanoDAQ-CAB-RJ-3), and two calibration interface modules (microDAQ-PCU). The calibration module is a single-board PC capable of automatic scanner calibration for up to 32 scanners through pressure controller module connections, with re-zero, span calibration, and 5-point validation automation accessible via webserver or command line with PoE interface support. Performance is specified at NASA Glenn Research Center, Building 6, Room 21, Cleveland, Ohio, with an 8-10 week delivery timeline as requested.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SATPC0041294 Tab 10 RFQ Open Market.pdf | ||
| SATPC0041294 Tab 10 BNJ_Redacted.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The NASA Glenn Research Center (GRC) has a requirement to purchase a Chell microDAQ3 Pressure Acquisition System and associated support hardware.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 64MD3-07010201 microDAQ3 64 pressure acquisition system.
- Scanner range: 35 kPa (5 psi)
- Absolute range: 0.5 to 115 kPa.
- Fitted with valve for purge and re-zero.
- Absolute or differential measurement.
- 1 mm tubulations.
- Includes TCP and UDP interface.
- With IEEE 1588 PTPv2 time stamping and Embedded web server.
- DC powered version with hardware trigger and CAN output.
- 400Hz per channel maximum acquisition speed.
- 0 to 90C compensated range.
2 64MD3-09010201 microDAQ3, 64 pressure acquisition system.
- Scanner range: 103kPa (15psi) differential.
- Absolute range 0.5 to 400kPa
- Absolute or differential measurement.
- With cal / purge valve.
- 1 mm tubulations.
- Includes TCP and UDP interface.
- With IEEE 1588 PTPv2 time stamping and embedded web server.
- DC powered version with hardware trigger and CAN output.
- 400Hz per channel maximum acquisition speed.
- 0 to 90C compensated range.
3 64MD3-11010201 microDAQ3 64 pressure acquisition system.
- Scanner range: 300 kPa (43.5 psi)
- Absolute range : 0.5 to 400 kPa.
- With valve.
- Absolute or differential measurement.
- 1 mm tubulations.
- Includes TCP and UDP interface.
- With IEEE 1588 PTPv2 time stamping and embedded web server.
- DC powered with hardware trigger and CAN output.
- 400Hz per channel maximum acquisition speed.
- 0 to 90C compensated range.
3 microDAQ-INT-8F microDAQ 8-way interface.
- Connects to 8 x MD3 / nanoDAQ-LT via MD3-CAB-1 cable.
- Features 8 x 9 way miniature D type connectors for MD3 / nanoDAQ-LT.
- Features SFP slot for fibre cable (module not supplied).
- 1 x miniature D type for Ethernet input and DC supply.
- Internal CAN hub.
- Buffered trigger outputs.
4 microDAQ-INT-4-02 microDAQ interface.
- Connects to 4 x MD3 / 24XX / nanoDAQ-
LT
- Features 4 x 9 way miniature D type connectors.
- 1 x miniature D type for Ethernet input and DC supply.
- Internal CAN Hub.
- Buffered Trigger Outputs.
- Shipped with mains powered DC supply
- Optional RJ45 and DC supply input for bench top use.
- Input supports PoE 802.3at.
5 MD3-CAB-3 microDAQ / nanoDAQ interface cable.
- Connects nanoDAQ-LT / microDAQ3 to microDAQ-INT-4/8.
- 3M length.
6 nanoDAQ-CAB-RJ-3 nanoDAQ / microDAQ3 cable.
- End 1: Mates with nanoDAQ / microDAQ3 microDAQ-INT4/8 with micro-D connector
- End 2: Male RJ45 for Ethernet. Bare ends for power, CAN and hardware trigger.
- 3m length
7 microDAQ-PCU microDAQ-PCU calibration interface module.
- Single board PC for automatic scanner calibration.
- Facilitates automatic calibration of MD3, CD5, 2416 and 2432 products.
- Connects to up to 6 pressure controller modules.
- Connects to up to 32 scanners
- Automates, re-zero, span calibration and 5-point validation.
- Can be interfaced through webserver or over command line.
- PoE interface.
3. Place of Performance
NASA Glenn Research Center
Building 6, Room 21 21000 Brookpark Rd Cleveland OH 44135-3127
4. Period of Performance
8-10 weeks ARO.