Salient Characteristics

Sonosite PX Purpose: Hopi Health Care Center (HHCC) is planning to acquire ultrasound-imaging equipment for portable applications. The equipment must be capable of operating in B Mode, M-Mode, Color Doppler, Color Power Doppler (CPD), Pulsed Wave and Continuous Wave modes. It must support transducers with linear, phased, and curved array formats. Further, it must include a full array of measurement and calculation packages.

Requirements:

POC (Point of Use) for daily ER activity

FIPS 199 and 200 compliant data security

Touch pad and physical controls for efficiency use

Meets infection control guidelines

Unparalleled clarity

DICOM capability to interface with PACS network

Security features to protect ePHI

Brand Name or Equal:

SonoSite PX Ultrasound - Part Number: L25100, QTY 1;

Transducer - Part Number: C5-1 L23119, QTY 1;

Transducer L 12-3 - Part Number: 622916, QTY 1;

Transducer P5-1 - Part Number: L24087, QTY 1:

Sonosite PX stand - Part Number: L25110, QTY 1;

Sonosite PX PowerPark - Part Number: L28623, QTY 1;

Printer BW Hybrid Graphic Sony UP-X898md - Part Number: L25200 (includes User Guide Education Courses: Ultrasound Guide), QTY 1.

System Specifications:

Stand Dimensions width x depth): 28.3 inches x 28.3 inches (72 cm x 72 cm)

Height Maximum (system and stand): 65 inches (165 cm) monitor up

Height Minimum (system and stand): 42.2 inches (107 cm) monitor down

Height Adjustment: 9.8” (24.9 cm) travel

Weight (System only) 16.6 lbs. / 7.5 kg

Weight (System and Stand) 85.1 lbs. / 38.6 kg

Casters: 5 inch Swivel, 4 Locking

User Interface: Projected Capacitive (PCAP) Touch Screen 10 inches x 5.7 inches (25.6 cm x14.5 cm)

User Interface: Physical Buttons

Work Surface Adjusts: 0 to 140 degrees

Monitor Angle: 15 degrees from horizontal

Work Surface Angle: 55 degrees from horizontal

Clinical Monitor: 15.6 inches (39.6 cm)

Clinical Monitor Image Size: 8.1 inches x 6.1 inches (20.7 cm x 15.5 cm)

Clinical Monitor Display Size: 13.5 inches x 7.6 inches (34.4 cm x 19.4 cm)

Screen Resolution (minimum): Display Size: 1920 x 1080 pixels, Image Size: 1152 x 864 pixels

Operating System: Windows 10 IoT Enterprise 2019 LTSC

Architecture: Digital Broadband

System Boot up: <40 secs

System Hard Drive: 128 GB

Gray Scale: 256 shades

Dynamic Range: Up to 183 dB

Transducer connection: 3 x system/stand, 1 x system only

HIPAA Compliance: Meets Federal Information Processing Standards (FIPS) for Moderate security control selections per FIPS199, FIPS200, and NIST 800-53.

The system shall withstand a drop from a height of 1 meter. The system shall be safe and function normally after this drop.

Imaging Modes:

1. 2D/B-mode a. Broadband Imaging

2. Tissue Harmonic Imaging

3. Pulse Inversion Harmonic Imaging

M-mode Imaging:

1. Color Velocity Doppler (CVD)

2. Color Power Doppler (CPD)

3. Color Variance (Cardiac only)

4. Pulsed Wave Doppler (PW)

5. Tissue Doppler Imaging (TDI) (Cardiac Only)

6. Continuous Wave Doppler (CW) (Cardiac Only)

B-mode Imaging:

1. B- mode Controls: Depth, Gain, Time Gain Compensation (TGC), Auto Gain, SonoMB, 2D Mode TH, 2D Image, Optimization (Res, Gen, Pen), Dynamic Range (Live and Frozen), Image Orientation (U/R, U/L, D/L, D/R), Sector width control (Steerable, Cardiac Only), Output

Power Control

Screen depth: 1 – 35cm

Support imaging frame rates up to 100Hz.

2. B-mode Imaging Features:

Imaging Technology: Adjusts imaging parameters based on depth eliminating complicated manipulation of multiple controls

Multi-beam technology to effectively reduce speckle artifact, enhance tissue contrast, and improve border detection.

8x live Read Zoom capability

Write Zoom on C5-1, IC10-3, and P5-1 transducers (Abdomen, OB, Early OB, Gynecology and Cardiac exam types only)

4x Frozen post processing Zoom

Dual Imaging

Centerline graphic on display for linear and curvilinear transducers

On screen needle guide support for IC10-3 and L19-5

Auto Gain adjust and reset

Procedure Mode

Auto Steep Needle Profiling (capabilities):

1. L19-5: Arterial, MSK, Nerve, Superficial, Venous

2. L12-3: Arterial, Breast, Carotid, MSK Nerve, Superficial, Venous

3. L15-4: Arterial, Breast, Carotid, MSK Nerve, Superficial, Venous

4. C5-1: MSK, Nerve, Spine

M-mode Imaging:

M-Mode Controls: Depth, Gain, Time Gain Compensation (TGC), Auto Gain, SonoMB, THI, Image Optimization (Res, Gen, Pen, Dynamic Range

(Live and Frozen), M-Mode Display Format of 1/3 and 2/3 and ½ and ½ and 2/3 and 1/3 side by side and full screen duplex, and M-Mode Sweep

Speed.

Color Imaging:

1. Color modes: Color Velocity Doppler, Color Power Doppler, and Color Variance (Cardiac Only)

2. Color controls: Color Gain, Color Scale, Color Invert, Color Flow Optimization (High, Mid, Low), Color Wall Filter, Color Steer Angle (Linear only), Adjust Color box size and position, Color Baseline, Color Compare, and Hide/Show Color

3. Max PRF used is 12500Hz.

Spectral Doppler Imaging:

1. Doppler modes: Pulsed wave (PW), Tissue Doppler Imaging (TDI) (Cardiac Only), Continuous Wave (CW) (Cardiac Only)

2. Doppler controls: Doppler Gain, Doppler Gate Size (PW/TDI), Doppler Invert, Doppler Auto Trace Type (Peak/Mean/Above/Below/All), Doppler Display Formats of 1/3 and 2/3, 1/2 and 1/2, 2/3 and 1/3, side by side and full screen duplex, Doppler Scale, Doppler Steer Angle

(Linear only), Doppler Sweep Speed, Doppler Baseline, Doppler Angle Correct, Doppler Wall Filter, Simultaneous (Linear Transducers Only), Doppler Volume, 256 gray levels in the spectral

3. Doppler display, and Cine review available

User Interface and Programmable Controls:

1. Projected Capacitive (PCAP) Touch Screen

2. Physical Buttons: TGC (Time Gain Compensation), Gain Wheel, Auto Gain, Depth, Select, Update, Touchpad, Annotations, Calculations, Image Save, Clip Save, Freeze, Color, M-Mode, Doppler, and 2D.

3. Clinical Display Information

4. Configurable Start Screen: Start, Scanning, Transducer/Exam Selection, Patient Information

5. Virtual QWERTY Keyboard

Onboard Education:

1. How-to videos: Imaging Basics, System Setup, Transducers, and System Use

2. Scan-along Educational Tutorials: Over 100 in English, French, German, Spanish, Japanese, Italian, and Mandarin

3. Tutorial Categories: Acute Care, Anesthesia, Pain Management, Obstetrics and Gynecology, Procedures, and Musculoskeletal

Scan-along Educational Tutorials:

1. Anesthesiology & Pain Management

Abdomen: Gastric Contents, II/IH Block, Rectus Sheath, TAP

Thorax: Paravertebral Airway

Lower Extremity: Adductor Canal, Femoral, Popliteal Sciatic, Saphenous, Subgluteal Sciatic

Upper Extremity: Axillary, Infraclavicular, Interscalene, Median & Ulnar, Supraclavicular

2. Obstetrics & Gynecology

Gynecology: Ovarian Cyst-TA, Pelvis-TA

Obstetrics: 1st Trimester OB- TA, 1st Trimester OB- TV, AFI, Amniocentesis, Ectopic-TA, Ectopic-TV, Fetal Biometry, Fetal Viability

Procedures:

1. Vascular Access: Femoral Artery, Femoral Vein IJV Long, IJV Transverse, PICC, Radial Artery, Subclavian Axillary

2. Other Procedures: Abscess Drainage, AFI, Amniocentesis, Foreign Body, Lumbar Puncture, Paracentesis, Pericardiocentesis, Thoracentesis

Acute Care:

EFAST: Lung, Lung Phased, Left Upper Quadrant, Pelvis, Right Upper Quadrant, Subxiphoid

EFAST Pathology: Left Upper Quadrant Free Fluid, Pelvis Free Fluid, Right Upper Quadrant Free Fluid, Subxiphoid Free Fluid

Cardiac: Apical, PSAX Heart, PLAX Heart, Subxiphoid

Cardiac Pathology: Infective Endocarditis, Tamponade

Lung/Thorax + Pathology: Airway Lung, Pleural Evaluation, Pleural Effusion, Pneumothorax, Pulmonary Edema

Abdomen: Aorta, Gallbladder Evaluation, Inferior Vena Cava, IVC Volume Status, Left Kidney, Right Kidney

Abdomen Pathology: AAA, Appendicitis, Cholecystitis, Cholelithiasis, Hydronephrosis, Pyloric, Stenosis, Renal Calculus

Pelvis: Bladder Volume, Pelvis-Transvaginal, Ovarian Cyst-TA, Testicular Torsion

Extremity: DVT, Long Bone Fracture, Tibia

Head/Neck + Pathology: Airway, Ocular, ICP, Retinal Detachment, Thyroid Evaluation

Musculoskeletal:

Shoulder/Rotator Cuff: Infraspinatus, Rotator Cuff, Rotator Cuff Tear, Tendinosis, Subscapularis Tendon, Supraspinatus Tendon

Shoulder - Other: AC Joint, AC Impingement, Biceps Tendon, Glenohumeral Joint, Posterior Groove

Shoulder Injections: AC Joint, Lateral Shoulder, Median Shoulder

Elbow: Anterior Recess, Cubital Tunnel, Distal Biceps Tendon, Elbow Olecranon, Elbow Ulnar Nerve, Lateral Aspect, Median Aspect

Elbow Pathology: Distal Biceps Tendon Tear, Lateral Epicondylitis, Ulnar Nerve Comp

Wrist/Hand + Pathology: Guyon’s Canal, Wrist Ulnar Nerve, Volar, Carpal Tunnel, DeQuervain’s, Median Nerve Entrapment

Hip, Pelvis Injections: Proximal Hamstring, Pubic Symphysis, Sacroiliac Joint

Knee: Iliotibial Band, Infrapatellar Tendon, LCL, MCL, Meniscus, Quadriceps Tendon, Tibia

Lower Extremity Injections: Achilles Tendon, Baker’s Cyst, Lateral Knee, MCL, Medial Knee, Patellar Tendonosis, Achilles Tendon Tear, Plantar Fasciitis

Ankle/Foot + Pathology: Ankle Ext Tendons, Ankle Flexor Tendons, MTP Joint, Tarsal Joint, Achilles Tendon Tear, Plantar Fasciitis

Onboard Image and Clip Storage Review:

2D Cine Review – minimum 10 seconds (or until number of frames fill up the 384 MB cine memory)

PW, CW, M Mode Cine Review – minimum 2 cine frames (slow sweep 23 seconds, fast sweep 6 seconds)

Internal Memory – 128 GB

700 maximum number of studies (average of 30 clips/images per study)

500 maximum number of clips/images per study

Thumbnail review of saved images and clips

Prospective and Retrospective Clip Store

Video clip playback at 1, ½ or ¼ of the captured rate

Video Clip Save Lengths: 2, 3, 4, 6, 10, 15, 30 and 60 seconds. (Retrospective/Prospective)

Image/Clip export format: JPEG, BMP and MP4

Report export Format: PDF

DICOM export to USB: JPEG, BMP and MP4

JPEG compression options: High, Medium and Low

Warranty & Service (minimum):

Five-year standard warranty on the system.

Five-year standard warranty on the stand head.

Five-year standard warranty on C5-1, IC10-3, L19-5, L12-3, L15-4, P5-1, transducers.

One-year standard warranty on battery, stand and accessories carrying the manufacturer’s label excluding the system and stand head.

Provide a replacement system within 24hrs.