Salient Characteristics .docx
DOCX document 33 KB Posted
- Attached to
- 6515--Hearing Aid Analyzer - FARGO Federal contract opportunity
- Solicitation number
- 36C26324Q0648
About this file
This document is a Salient Characteristics document that describes the detailed technical specifications for a hearing aid analyzer required by the Fargo VA Medical Center. The agency is seeking to purchase 4 units of the Audioscan Verifit2 Binaural Real Ear Hearing Aid Analyzer, or an equivalent brand name or equal product. The required analyzer must meet specific capabilities such as binaural real-ear measurement, wideband verification, speech-activated directional tests, and wireless networking. The salient characteristics provide granular details on the required technical specifications, stimuli, and measurement accuracy.
The related federal contract opportunity is Solicitation Number 36C26324Q0648 issued by the Department of Veterans Affairs for the "6515--Hearing Aid Analyzer - FARGO" requirement. This is a combined synopsis/solicitation for commercial items set aside for small businesses. Quotes are due by 10:00am CST on 05-28-2024 and should be submitted via email to the Contracting Specialist. The government anticipates awarding a single firm-fixed price contract to provide the hearing aid analyzers to the Fargo VA Health Care System.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFQ 36C26324Q0648.pdf | ||
| 36C26324Q0648_1.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics Description & Salient Characteristics: The Fargo VA Medical Center has a need for the purchase of hearing aid analyzers for Audiology.
Salient Characteristics Brand Name or Equal to: Audioscan Verifit2 Binaural Real Ear Hearing Aid Analyzer
Includes Binaural REM and Test Box Wideband verification minimum 12.5 kHz Calibrated real speech & percentile analysis DSL, NAL, & CAMFIT targets Wideband RECD & REDD measurement ANSI/IEC test box measurement Binaural monitor headphones real-time Binaural Sound Field Assist for simultaneous binaural REM Speech-activated directional test Noise reduction, feedback, occlusion tests SII target ranges in DSL adult & child RMS error from target Ambient noise level check for REM FM fitting protocol Frequency lowering fitting protocol CROS/BiCROS fitting Noah module via wired or wireless networking PC control/display capability (On-top mode) Puttyless coupling in the test box Quick-connect couplers Context sensitive help with instructive pictures Stereo headphone monitor amplifier 250 mW into 16 ohms, L/R Simultaneous stimulus channels 2 Simultaneous measurement channels 2 Connectors
| 1 - Ethernet (RJ45) |
| 5 - USB |
| 1 - HDMI |
| 2 - External speakers (RCA) |
| 1- Display unit/Test box interconnection (HDMI style) |
| 1 - Probe dock (Mini-din) |
| 2 - Probe microphone (3.5 mm TRS) |
| 1 - WRECD transducer (3.5 mm TRS) |
| 1 - Teletest handset (3.5 mm TRS) |
| 1 - Binaural monitor headphone (6.3 mm TRS) 2 - Test box ref. mic. (3.5 mm TRS) |
| 1 - Binaural coupler microphone (3.5 mm TRS) 1 - Battery substitute (3.5 mm TRS |
REM
REM Speakers dual 5 x 9 cm w/ ducted ports Probe modules dual probe and ref. microphones Probe microphone tube silicone Maximum 1 mm diameter x 75 mm Analyzer frequency range 200 – 16000Hz Analyzer filter parameters tones, warble 1/12 octave speech, noise 1/3 octave Analyzer display range 200 – 12500 Hz Probe microphone noise floor (200 – 12500 Hz) Measurement dynamic range 30 – 135 dB SPL (200 – 2500 Hz) 30 – 140 dB SPL (2500 – 12500 Hz) Speech-like stimuli calibrated speech (level and spectrum), ISTS, band-limited and s/sh for verifying frequency lowering, live speech Broadband noise stimuli tone burst, pseudo-pink noise Narrowband stimuli warble sawtooth modulated +/- 3% over 128 ms User supplied stimuli WAV file, auto level Stimulus level range 40 – 85 dB SPL in 5 dB steps Measurement accuracy at 1 kHz (CAL position) +/- 1 dB Measurement accuracy re 1 kHz 200 – 2000 Hz +/- 1.5 dB SPL 2000 – 8000 Hz +/- 2.5 dB SPL 8000 – 12500 Hz +/- 4 dB SPL Equalization modified pressure method – concurrent and stored
TEST BOX
Test box speakers 1 front, 2 side, 30 mm independent Isolation @1kHz >25dB Analyzer frequency range 200 – 16000 Hz Analyzer filter parameters tones 1/12 octave speech, noise 1/3 octave Analyzer display range 200 – 12500 Hz Coupler microphone noise floor (200 – 12500 Hz) Frequency accuracy 1% Measurement dynamic range 30 – 135 dB SPL (200 – 2500 Hz) 30 – 140 dB SPL (2500 – 12500 Hz) Speech-like stimuli calibrated speech (level and spectrum), ISTS, band-limited and s/sh for verifying frequency lowering, live speech Broadband noise stimuli tone burst, pseudo-pink noise Narrowband stimuli pure tone User supplied stimuli WAV file, auto level Test stimulus levels 40 – 90 dB in 5 dB Battery drain range (+/- .01 mA) 0 – 20 m Battery drain accuracy +/- 5%
| Part Number |
| Description |
| Quantity |
| VF-2 |
| Audioscan Verifit2 Binaural Real Ear Hearing Aid Analyzer |
Includes Mouse, Monitor Headset, 4 Battery Pills, RECD Transducer, Telecoil Test Capabilities and Cables.
2 Years Parts & Labor Warranty
| VA-201 |
| NOAH Module, 3.6 or higher |
| 1 |
| VA-201-ADD |
| Additional Noah Module |
| 3 |
| Installation Required |
| 1 |
Delivery Location:
Fargo Health Care System 2101 N Elm St.
Fargo, ND 58102
Delivery and Installation
1. The vendor will be responsible to deliver the materials 90 days ARO Materials of all pieces to which make the equipment fully functional, including all accessories must be present upon delivery.
2. Delivery Location:
Fargo Health Care System 2101 N Elm St.
Fargo, ND 58102
Contacts:
| Sam Hammerlund |
| Chief Audiology |
| Sarah.hammerlund@va.gov |
| 701-239-3700 ext. 3174 |
| Trevor Nitschke |
| Deputy Chief Supply Chain Officer |
| 701-239-3700 ext. 3333 |
| Trevor Nitschke |
| Jennifer Ernst |
| Chief Supply Chain Officer |
| 701-239-3700 ext. 3337 |
| Jennifer.ernst@va.gov |
3. All items to be shipped F.O.B. destination. Shipping shall be included in the product price and shall not be billed separately.
4. No tools during deinstallation or installation will be supplied/provided by the VA.
File details come from the government source that posted it. Updated .