- SMALL BUSINESS SET-ASIDE FOR SOLE SOURCE: one (1) Elemental Analyzer (EA), including components
- Solicitation number
- W911QX24Q0220
About this file
This document is a Salient Characteristics Commercial Availability Certification related to a Federal Contract Opportunity for one (1) combustion based elemental analyzer (EA) that meets specific requirements. The system must be capable of determining carbon, hydrogen, and nitrogen (CHN) in organic materials, with the ability to also analyze oxygen (O), and carbon, hydrogen, nitrogen and sulfur (CHNS). Key requirements include separate combustion and reduction tubes, furnace temperature of at least 1200 C, temperature programmed desorption column, 50-sample autosampler, and 2-year preventive maintenance. The contract opportunity, Solicitation Number W911QX24Q0220, is a Small Business Set-Aside for Sole Source with Elementar Americas Inc. as the intended awardee. Delivery is required 12 weeks after Receipt of Order to Aberdeen Proving Ground, Maryland. Offers are due within 5 business days.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Provisions and Clauses Full Text.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature:
Rose A. Pesce-Rodriguez Technical Point of Contact (TPOC)
Salient Characteristics The Government requires one (1) combustion based elemental analyzer (EA) that meets the following minimum requirements:
1. The system as delivered shall be capable of determining carbon, hydrogen and nitrogen (CHN) in organic materials.
2. The system shall be configurable to also analyze oxygen (O), and carbon, hydrogen, nitrogen and sulfur (CHNS).
3. The system shall have separate combustion and reduction tubes not to be combined into a single tube .
4. The furnace shall be capable of operating at or above a temperature of 1200 °C.
5. The system shall allow independent control of combustion and reduction tube temperatures.
6. The system shall use a temperature programmed desorption (TPD) column, which traps and holds gases until the appropriate temperature is achieved for complete baseline separation of evolved gases.
7. The system shall include an autosampler with a minimum of 50 sample holders.
8. The system shall include at least one (1) electronic controller and all necessary software for data collection and analysis.
9. The system shall include all items necessary for installation and operation, includng expendible components sufficient for 4000 CHN analyses.
10. The systems shall include standard materials for CHN analysis.
11. The system shall be configured for operation with both argon and helium carrier gases.
12. The system shall come with an additional 2-year preventive maintence and repair service.
13. The contractor shall deliver the system to Aberdeen Providing Ground (APG), MD 21005, FOB Destination preferred.