This document is a Request for Quotes (RFQ) for a Centese Digital Pleurovac system (brand name or equal) for the National Institutes of Health (NIH) Clinical Center. The government is seeking to procure 2 Centese Thoraguard Drainage Kits and 6 Centese Digital Pleurovac Modules. The required system must meet 11 salient requirements, including digital chest drain capabilities, notification features, water seal, air leak measurement, self-clearing options, onsite training, technical support, and security features. The RFQ is being competed on a full and open basis, with a NAICS code of 339112 - Surgical and Medical Instrument Manufacturing. Quotes are due by July 11, 2024 at 4:00 PM EST and will be evaluated based on technical capability, price, and past performance to determine the best value for the government. The anticipated delivery timeframe is within 4 weeks of award.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
List of Salient Requirements
1. Digital chest drain capable of allowing therapy targets to be set (This improves and increases patient safety via continuous monitoring)
2. Notify users when targets are achieved. (This shortens the time that chest tubes are in place and therefore benefits the patient from a pain standpoint.)
3. Has a digital water seal equivalent (Without water seal setting, this limits the options for management of patients)
4. Measures air leak with a resolution of 1 ml/min from 0-15 ml/min (improves patient safety by ensuring that the leak is truly resolved prior to tube removal)
5. Self-clearing chest tube option. (Clot formation within the tube impedes drainage. Self-cleaning provides on demand clearing of the tube to prevent a blocked line-patient complications)
6. Self-cleaning capabilities across multiple lines. (Post operative thoracic surgery patients can have multiple chest drains in place. The self-cleaning technology allows the user to maintain drain patency and avoid urgent unplanned surgeries as a result of compromised drains)
7. Onsite training
8. 24 hour technical support
9. Able to hang from hospital bed securely
10. Features to prevent unit from tipping over
11. Access port to retrieve samples