This is a Sources Sought Notice (RFI) issued by the Department of Veterans Affairs, Network Contracting Office 22, seeking information about potential suppliers for motorized surgical tables used in fluoroscopic imaging procedures. The notice is issued solely for information and planning purposes and does not constitute a solicitation or binding contract. Responses must be submitted via email to Debby.Abraham@va.gov by June 19, 2026 at 4:00 PM EST.

The VA Greater Los Angeles Healthcare System seeks motorized surgical C-arm tables with specific technical specifications, including a maximum table top length of 97 inches, radiolucent area of 38.5 inches, carbon fiber tabletop material, 2-inch maximum memory foam mattress, attenuation ratings (.7mm Al Eq top, .5mm Al Eq mattress), Z-motion capability of 30-40 inches, Trendelenberg of ±20°, lateral roll of ±15°, hand and foot controls, accessory clamp attachments, carbon fiber arm board for two arms, and maximum 500-pound patient capacity. The reference product is the Capintec Table, Surgical C-Arm model 870 with component part numbers specified. Respondents must provide detailed capability statements, business size classifications under NAICS code 339112 (1,000 employee threshold), small business certifications, manufacturer or distributor status, lead times, FDA clearance information, pricing information for market research purposes, SAM.gov Unique Entity ID/Cage Code numbers, and comprehensive Buy American Act compliance documentation including country of origin, domestic content percentages, and availability of domestic alternatives. Responses will be used for market research and acquisition planning only and do not obligate the government or result in automatic inclusion on prospective bidders lists.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice

*= Required Field Sources Sought Notice

SUBJECT* Pain management C arm table- Motorized surgical table for imaging during fluoroscopic procedures

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE* 90815

SOLICITATION NUMBER* 36C26226Q0983

RESPONSE DATE/TIME/ZONE 06-19-2026 4PM PACIFIC TIME, LOS ANGELES, USA

ARCHIVE 30 DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS N

SET-ASIDE

PRODUCT SERVICE CODE* 6515

NAICS CODE* 339112

CONTRACTING OFFICE ADDRESS Department of Veterans Affairs

Network Contracting Office 22 4811 Airport Plaza Drive Suite 600 Long Beach CA 90815

POINT OF CONTACT*

CONTRACTING OFFICER

DEBBY ABRAHAM

DEBBY.ABRAHAM@VA.GOV

PLACE OF PERFORMANCE

ADDRESS VA Greater Los Angeles Healthcare System 11301 Wilshire Blvd

Los Angeles CA

POSTAL CODE 90073

COUNTRY USA

THIS REQUEST FOR INFORMATION (RFI) / SOURCES SOUGHT NOTICE IS ISSUED

SOLELY FOR INFORMATION AND PLANNING PURPOSES. THIS IS NOT A

SOLICITATION.

SUBMISSION OF INFORMATION ABOUT PRICING, DELIVERY, THE MARKET,

AND CAPABILITIES IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI

FOR PLANNING PURPOSES IN ACCORDANCE WITH (IAW) FAR 15.201(e).

DISCLAIMER

This RFI is issued solely for information and planning purposes and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. IAW FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.

SOURCES SOUGHT/RFI DESCRIPTION

This is NOT a solicitation announcement. This is a Sources Sought Notice / RFI only. The purpose of this Sources Sought Notice / RFI is to gain information about potential qualified sources and their size classification relative to NAICS 339112 (size standard of 1,000 Employees). Responses to this Sources Sought Notice / RFI will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Notice / RFI, further

RFIs and/or a solicitation or other announcements may be published.

The Department of Veterans Affairs (VA) is seeking sources capable of providing Motorized surgical table for imaging during fluoroscopic procedures and components that can address the attached list of DRAFT requirements for multiple VA Greater Los Angeles.

The list of DRAFT characteristics is intended to be descriptive, not restrictive, of the supplies/services that are required.

BRAND NAME/PART #/QUANTITY

• CAPINTEC TABLE, SURGICAL C-ARM - 870, 115VAC, REC/# 058-870-10/1 EA

• CAPINTEC CLAMP, ACCESSORY ATTACHMENT/# 058-849/4 EA

• CAPINTEC FOOT CONTROLLER, SURGICAL C-ARM TABLE/# 058-877/ 1 EA

• CAPINTEC ARM BOARD, CARBON FIBER, TWO ARMS/# 056-866/ 1 EA

EQUAL TO PRODUCT

Motorized surgical table for imaging during fluoroscopic procedures

SALIENT CHARACTERISTICS

This table will be used to treat patients in Chiropractic Clinics at GLA for studies/treatments.

• Must have a table top length of maximum 97”.

• Must have Radiolucent area maximum of 38.5” unobstructed.

• Must have Tabletop material of carbon fiber.

• Must have mattress with a maximum of 2” memory foam.

• Must have Attenuation Top with a maximum of .7mm Al Eq

• Must have Attenuation Mattress with a maximum of .5mm Al Eq.

• Must have Z motion with a minimum of 30” and maximum of 40”.

• Must have Trendelenberg of ±20°.

• Must have Lateral Roll of ±15°.

• Must have Hand Control to include Height, roll, tren

• Must have foot controller to include Height, roll, tren

• Must have Clamp, Accessory Attachment-4

• Must have ARM Board, Carbone Fiber, Two Arms-1

• Must have patient capacity of maximum of 500 lbs.

• Must have warranty

If your company is interested and capable of providing the required supplies/services, please provide the information indicated below. Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions:

(1) Submit your capabilities statement illustrating how your organization can/cannot meet the list of DRAFT/SALIENT CHARACTERISTICS/SOW requirements. For instances where your company cannot meet the DRAFT/SALIENT CHARACTERISTICS/SOW requirement(s), please explain. For instances where your company can meet the

DRAFT/SALIENT CHARACTERISTICS/SOW requirement(s), please show how your company meets/exceeds each requirement.

(2) Please review the list of DRAFT/SALIENT CHARACTERISTICS/SOW requirements and provide any additional feedback or suggestions. If none, please reply as N/A.

(3) Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran-Owned Small Business (SDVOSB), Veteran-Owned Small

Business (VOSB), HUBZone, Woman Owned Small Business (WOSB), Large Business, etc.

(4) Is your company considered small under the NAICS code identified in this RFI?

(5) Are you the manufacturer, authorized distributor, and/or can your company provide a solution to the required supplies/services described in the list of DRAFT/SALIENT

CHARACTERISTICS/SOW?

(6) If you are a large business, do you have any designated/authorized distributors? If so, please provide their company name, telephone, point of contact and size status (if available).

(7) If you’re a small business and you are an authorized distributor/reseller for the items identified above, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified.

(8) Limitations on Subcontracting: How does your business ensure compliance with the limitations on subcontracting as outlined in 13 CFR § 125.6?

(9) Are the items you are identifying/providing considered Commercial of the Shelf (COTS) items as defined in FAR Part 2.101 under commercial items?

(10) Non-Manufacturer Rule: If applicable, can you confirm your business complies with the Non-Manufacturer rule? Specifically, does your company: Provide a product from a small business manufacturer or processor? Not exceed 500 employees? Primarily engage in the retail or wholesale trade and normally sell the type of item being supplied? Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice?

(11) What is your lead time to deliver a single unit with all components? Is there scale in lead time with greater quantities? Please elaborate.

(12) What is estimated life span of your solution? What support/services does that entail?

(13) Does your proposed equipment have FDA clearance? Please specify what FDA clearance(s) have been obtained.

(14) Does your organization offer a leasing solution? Please elaborate.

(15) Does your company have a Federal Supply Schedule (FSS) GSA/NAC/SAC/BPA/NASA

SEWP or any other Federal Government contract? If so, please provide the contract number(s).

(16) If you are an FSS GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract holder, are all the items/solutions you are providing information about available on your schedule/contract? Please state if all or some items are available on the contract.

(17) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.

(18) Please provide your SAM.gov Unique Entity ID/Cage Code number.

Buy America Compliance

Respondents shall provide sufficient detail to determine compliance with the Buy American

Act, including country of origin, domestic content, and availability of domestic alternatives.

Responses should be complete and supported with verifiable information.

1. Please identify the country of origin for the end product(s) being offered. Include the location of final assembly/manufacture.

2. Please provide the percentage of domestic content (by cost) for each product(s).

Describe the origin of the major components.

3. Does your product qualify as a “domestic end product” in accordance with the Buy

American Act?

4. If the proposed product is not compliant, are you able to provide a BAA-compliant domestic alternative?

5. Please identify the manufacturer name and location for the product(s). Confirm whether the manufacturer is U.S.-based.

6. Identify the country of origin of each product and indicate whether the product qualifies as a U.S.-made or designated-country end product in accordance with the

Trade Agreements Act

Responses to this notice shall be submitted via email to Debby.Abraham@va.gov. Telephone responses will not be accepted. Responses must be received no later than Friday, June 19, 2026 by 4:00 PM EST.

All responses to this Sources Sought/RFI will be used for planning purposes only. Responses to this Sources Sought Notice / RFI are not considered a request to be added to a prospective bidders list or to receive a copy of the solicitation. If further RFIs and/or a solicitation or other announcement is issued as a result of the information provided from this RFI, all interested parties must respond to the specific posting separately IAW the specifications of that announcement.