Purchase Description 20M-279A-MA Rev.1.pdf
PDF 677 KB Posted
- Attached to
- High Range Balance (50 lb) Federal contract opportunity
- Solicitation number
- FA226320Q0010
About this file
This combined synopsis/solicitation seeks offers for a high range balance to strict specifications. Offerors must propose a commercial balance meeting requirements in the attached purchase description, including a display resolution of at least 10 milligrams, a continuous measuring range of 3 kilograms to 26 kilograms, and repeatability not exceeding 15 milligrams throughout the range. The solicitation provides detailed functional and technical requirements for controls, draft shielding, calibration adjustment capabilities, mechanical properties, and operating environment tolerances. Offers are due by August 24, 2020 and must be submitted electronically through DoD SAFE. The award will be a fixed-price contract under simplified acquisition procedures by the Air Force Materiel Command Lifecycle Management Center at Wright Patterson Air Force Base.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Instructions to Offerors.pdf | ||
| Instructions to Offerors.pdf | ||
| Anticipated Delivery Locations.pdf | ||
| Evaluation Factors.pdf | ||
| Solicitation - FA226320Q0010.pdf | ||
| TEP WORKSHEET.xlsx | XLSX spreadsheet | |
| 3-Year AFMETCAL ML Warrranty dated 11 August 2008.pdf | ||
| Data Requirements.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
20M-279A-MA
30 January 2020
Revision 1
PURCHASE DESCRIPTION
FOR
High Range Balance
1.0 SCOPE: This purchase description specifies a High Range Balance to be used in the calibration of mass standards by selected Precision Measurement Equipment Laboratories (PMELs) to calibrate both
Avoirdupois and Metric field masses and torque calibration mass sets in the range of 3 kilograms to 50 pounds and also seraphins in the range of 10 to 26 kilograms.
2.0 PERFORMANCE REQUIREMENTS:
2.1 Display Resolution: The Balance shall be readable to at least 10 milligrams.
2.2 Measuring Range: The Balance shall have a continuous metric range, which covers 3 kilograms to 26 kilograms minimum.
2.3 Repeatability: The repeatability of the Balance shall not exceed 15 milligrams throughout the
Measuring Range (see PD paragraph 2.2 above).
2.3.1 Repeatability Definition: Repeatability for this instrument is defined as the standard deviation of 6 sets of 2 pairs of differences between two master weights of the same nominal value (6 runs using ABBA method).
2.4 Reliability: The Balance shall maintain its repeatability specification over a period of 12 months, minimum.
2.5 Shift Test/Eccentricity: The Shift Test error of the Balance shall not exceed 250 milligrams throughout the Measuring Range (see PD paragraph 2.2 above).
2.6 Span: The Span error of the Balance shall not exceed 275 milligrams at 25 pounds and 575 mg at 26 kilograms.
2.7 Stabilization Time: The stabilization time for an individual reading shall be no more than 20 seconds.
3.0 FUNCTIONAL REQUIREMENTS:
3.1 Manual Control: The Balance shall be self-contained and manually controlled via push button(s) or switch.
3.1.1 Internal/External Adjustment: The Balance shall have an internal, motorized substitution weight(s) and the ability to use an external substitution weight(s). Item’s internal adjustment function shall have the option to be both automated (settings configured by user) or user-initiated.
3.1.1.1 Internal Substitution Weight: When using the internal substitution weight adjustment function, operator shall have the ability to edit the programmed mass value of the internal weight.
3.1.1.2 External Substitution Weight: When using an external substitution weight for adjustment, operator shall have the ability to enter/edit the actual mass value of the external weight(s).
Attachment 1
Revision 1
3.2 Ancillary Equipment: The Balance shall have a rigid, separate, external draft shield, which provides adequate space around the Balance and easy access to the weighing chamber.
3.2.1 Materials: External draft shield shall be composed primarily of at least 5 clear panes of
Lexan, glass, or equivalent transparent material with ¼ +/- 1/16 inch thickness. Shield shall not be composed of Plexiglas due to inherent static properties.
3.2.2 Assembly: External draft shield shall come fully assembled, or be assembled in-situ by the manufacturer’s trained technician.
3.2.3 Functionality: External draft shield shall be placed on a table and shield entire weighing portion of the Balance, and allow for cords, such as power cords, to be positioned as needed. External shield shall function such that the operator is able to access the Balance’s weighing pan through the internal draft shield in order to add and remove weights as necessary to perform measurements through the use of a sliding door or hinge mechanism.
3.2.4 Surface Area: Assembled closed draft shield shall not cover a surface area of more than 6 square feet.
3.3 Calibration Adjustment: Any special equipment, fixtures, tools, standards, etc., required for calibration and are not common to a secondary calibration laboratory must be supplied with the
Balance. A unit that fails calibration and is recommended to be aligned or adjusted in order to meet all requirements of para 2.0, either manually or using alignment/adjustment software, shall include any special hardware, software, software licenses needed to perform the alignment/adjustment process.
4.0 MECHANICAL & PHYSICAL REQUIREMENTS:
4.1 Components: Commercial parts and materials that are used shall have properties consistent with a precision laboratory instrument.
4.1.1 Weighing Pan: The Balance’s weighing pan/surface shall be made of stainless steel, or similar material, which resists rusting, chipping, flaking, or scraping. Weighing pan/surface shall not be composed of, or covered in, a soft material, such as cork, which may flake, disintegrate, or rub off over time.
4.2 Use of Best Commercial Practices: The equipment shall be designed and constructed of materials that meet the requirements of this specification and in accordance with the best commercial practices for industry approved equipment. It shall be designed and manufactured to maximize reliability, maintainability, accuracy of operation, and ease of operation. All parts (which include, but are not limited to, shafts, bearings, regulators, switches, controls) shall have proper clearance and adjustments. They shall work together so that the equipment will supply the rated requirements without unnecessary strains, vibrations, or overheating. They shall be able to withstand the conditions met in shipping, storage, installation, and service. All materials and parts of this equipment shall be capable of withstanding the environmental requirements of this specification.
4.3 Matched and Selected Components: These shall be kept to an absolute minimum. Use of specially built, nonstandard components shall be avoided.
4.4 Model and Serial Number Identification: Balance’s Part Number, Manufacturer, and Serial Number shall be identified clearly on the item in a location visible to the operator. Unique item identification shall follow Item Identification and Valuation requirements as stated in the contract.
Revision 1
4.5 Operating Environment/Storage:
4.5.1 Operating Temperature: The Balance shall be capable of operating in an ambient temperature of 67 to 79 degrees F without loss of required performance.
4.5.2 Operating Humidity: The Balance shall be capable of operating in an atmosphere from 20% to 50% relative humidity without loss of required performance.
4.6 Facility Electrical Power and Connections: The Balance shall be capable of operating on standard
110 V AC 60 Hz electrical power.
4.7 Safety: Documentation will be provided to support the means by which electrical safety considerations have been addressed. The safety of all personnel will be assured in conductive or nonconductive contact with the outside of the equipment, and during the changing of the fuse with the power to the system and its components are turned on. This documentation shall consist of a minimum of a citation of a UL or ANSI identified standard or equivalent United States Standard to which the equipment conforms.
4.8 Working Surface Area and Height: The Balance’s weighing pan surface shall be located no more than
8 inches higher than the Balance’s supporting table. In addition, when moving a 20 inch tall weight from the Balance’s supporting table and onto the weighing pan, the weight shall be lifted no more than 9 inches vertically, with no obstructions from the External Draft Shield (see PD paragraph 3.2 above). The
Balance shall accommodate a hanger weight with a surface area of 19 x 24 inches.
5.0 OPERATOR’S MANUAL: Manual shall be IAW DD1423s and Data Item Descriptions, and shall include, at a minimum, the following information:
A. Setup/installation of balance including external draft shield
B. Procedure on how to adjust the balance per sections 2.5 and 3
C. Drawing/Diagrams needed for adjustment
D. A list of peripheral support equipment required to accomplish the adjustments
E. Information on any internal adjustments made during operation
F. Definitions on changeable settings
G. Procedure on how to package, store and transport the balance
6.0 TESTING: This item will be inspected and accepted at destination as stated in the contract. The item will be checked with test equipment with parameters equal to or better than those required to assure that specified performance parameters are met. Test equipment used shall have first echelon certification directly traceable to the National Institute of Standards Technology (NIST) or other AFMETCAL-approved equivalent international laboratory. Tests will be performed to assure that the item meets the requirements of this purchase description. Such testing at destination does not relieve the contractor of performing all inspections and quality checks at the point of manufacture as necessary to assure performance as specified herein.
Attachment:
Data Requirements
Attachment to
Purchase Description 20M-279A-MA
High Range Balance
DATA REQUIREMENTS
MANUALS: A complete user and service manual and calibration procedure shall be provided with each unit in contractor format. Manuals shall be on CD/DVD-ROM in Indexed Portable Document Format
(iPDF). The manuals shall comply with Data Item Description (DID) DI-TMSS-80527C and the Contract
Data Requirements List (DD Form 1423).
File details come from the government source that posted it. Updated .