About this file

This document includes specifications for a MonCvONE perimeter device and related federal contract opportunity. The specifications call for a hemispherical cupola perimeter with a 30cm radius capable of full field projection and Goldmann test sizes I through V. Additional required features include dimmensions of 62x35cm footprint and 74cm height, a weight of 23kg without accessories, and a power supply of 110-240V at 3.6-1.8A and 50-60Hz. The device must support static automated perimetry tests ranging from 60 to 12 seconds, as well as mixed kinetic and static tests, binocular tests for drivers and low vision patients, and single vision and diplopia tests. It must also allow programmable photopic and mesopic backgrounds and testing with user-defined color filters.

The related federal contract opportunity is solicitation number NIH-SS-6870298 from the National Institutes of Health seeking a perimeter update for the MonCvONE device that adds electrophysiology and dark adaptation software. The specification document provides details on the requirements.

View the file

Other files for this federal contract opportunity

Other files attached to MonCvONE perimeter update with electrophysiology software, newest first.
File Type Posted
Sole Source Justification for MS VA Tester.pdf PDF
Notice of Intent to Sole Source.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

MonCvONE CR Specs

• Hemispherical cupola with 30 cm radius;

• Full field projection perimeter, up to the true limits of the visual field;

• Test sizes: Goldmann I, II, III, IV, V, ganzfeld;

• Dimensions: footprint=62x35cm, height=74cm;

• Weight: 23 kg (without PC, printer and electric table);

• Power supply: 110-240V, 3.6-1.8A , 50-60Hz.

Photopic background

(10 cd/

Standard static automated perimetry

FAST60, FAST30, FAST24, FAST12

Mixed perimetry

(kinetic +

Binocular tests for drivers (Group 1 and 2)

Binocular test for low vision

(Large field lenses for correction of

Blue/Yellow perimetry (SWAP)

Manual perimetry

(interactive Video imaging

Attraction perimetry

(young Single vision test diplopia

Programmable background

(mesopic, photopic)

Dark adapted chromatic perimetry

Perimetry (up to 5 user defined color filters)

Dark adaptometry

Full field stimulus threshold (FST)

Chromatic pupillometry

Eye movement recording

Vision electrophysiology

Electrophysiology (flash ERG and VEP, EOG)

File details come from the government source that posted it. Updated .