Sole Source Justification for MS VA Tester.pdf

PDF 157 KB Posted

Attached to
MonCvONE perimeter update with electrophysiology software Federal contract opportunity
Solicitation number
NIH-SS-6870298
Issued by
Department of Health and Human Services National Institutes of Health Office of Logisitics and Acquisition Operations

About this file

This document includes a sole source justification and technical specifications for a MonCvONE perimeter update with electrophysiology software. The justification seeks a perimeter with a 30 cm radius hemispheric bowl, scotopic and photopic luminance capabilities, fixation and near-infrared light sources, and bioelectric amplifiers with specific input ranges, resolutions, and filter options. The related federal contract opportunity is for the same perimeter update and software from the National Institutes of Health Office of Logistics and Acquisition Operations. The specifications require a perimeter, software, and accessories as described in the attached specification document to meet the needs of the agency.

View the file

Other files for this federal contract opportunity

Other files attached to MonCvONE perimeter update with electrophysiology software, newest first.
File Type Posted
Specs.pdf PDF
Notice of Intent to Sole Source.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Specifications

MONCVONE STIMULATOR

Description of the supplies and services required to meet the Agency’s needs.

MonCvONE perimeter update with electrophysiology software

• hemisphere with 30 cm radius

• white background luminance: scotopic dynamic range of 7.0 log units up to 1400 cd/m2

• amber background (590 nm): up to 350 cd/m2

• white flash stimulation: up to 10 cd.s.m-2 with 5 ms duration

• blue flash stimulation (447nm) up to 60 photopic cd/m2

• red flash stimulation (655nm) up to 160 photopic cd/m2

• Fixation light sources: red 625 nm central and 15 degrees right and left

• Near infra-red illumination for video monitoring: 940 nm

BIOELECTRIC AMPLIFIERS

The amplifiers include controls for electrode impedance and for calibration.

They are designed to be placed close to the patient so as to minimize risks of interference.

Input impedance 1000 GOhms // 220 pF Input noise level from 1 to 800 Hz: 2.6 μV peak-peak from 1 to 25 Hz: 0.5 μV peak-peak

CMRR @ 50 Hz

115 dB

Input dynamic range

- 3.2 mV to + 3.2 mV

Admissible input DC offset

- 1200 mV to +1200 mV (electrode polarization)

Test of electrode impedance application of a current of 20 nA with a 25 Hz modulation (

Test of calibration injection of an input signal of calibrated amplitude

Digital resolution

0.4 μV 16 bits

High pass filter Programmable: 0.1Hz 1.0 Hz 10Hz

Low pass filter Programmable from 25 Hz up to 1200 Hz

Band reject filter Programmable 50Hz or 60Hz or none

Number of channels

HVM-AMPLI2: 2, HVM-AMPLI4:

4, HVM-AMPLI5: 5