Spccilicntions for NRL Quenching Dilatomcter

Tcmpcralure range: -150 °C to 1600 QC

I !eating rate: up lo 4000 K/s

Cooling rate: up to 2500 K/s

Maximum Resolution: 0.1:!511111

Measurement Range: adjustable from t 00 11111 to 5000 11111 from 1.cro setting

Acquisition Rate: up to I 000 measun:ments/s

Operating environment: air. inert. reducing. m.:idating. or vuc1111m. The t)pc and tlow rate shall he controllable along with the thermut environment through the programming software

Vacuum by turbomolccular pump for an oil free system achieving a vacuum of less than Io·� mhar

Sample diameter: accommodates samples at least 3 mm in diameter

Sample length: accommodates samples at least IO mm long

In-situ high-speed CCD imaging ol'samptc surface (minimum 3 mm x 3 mm area) for enhanced resolution of transformation temperature

State of the art computer with Windows operating system, di latomcter control software. data acquisition software, analysis sofiwarc. and output software. including:

Prograrnmuble heating and cooling cycles including potential for multiple heating ramps.

hole.I times. und cooling ramps with linear or exponential cooling Calculation or Oct ta L. cocflicienl of thermal expansion, and dilatometry curves with statistical cvaluution or mean and confidence interval Software for detection or nascent melting and prevention of sample melting Software for generntion of CCT. Cl-IT. and TTT curves

---...... -=-ee:--,.._=-- Both �phical nnd tabular output of data ASCII export of data files

Thermocouple welder

Calibration standards

Standard accessories including: i) two spare sets of fused quartz push rods and ii) two spare sets of AbOi push rod�

Installation: lnsla[Jntion will be at the Naval Research Laboratory and shall be conducted no later than sixty (60) days after system delivery

Training

Warranty: One-year standard commercial warranty that covers all parts. labor. and travel expenses for on-site support or the equipment must be included in the syslcm price. The warranty period shall begin on the date or system acceptance. Thc Contractor shall register any equipment requiring separate \\arranty by other manu focturers: such wurranties shalt also he effective from the date of system acceptance.

RFQ # N00173-19-Q-0240