This statement of requirements document outlines specifications for a live cell imaging microscope system needed by the National Institute of Standards and Technology. The microscope must have capabilities including fluorescence, brightfield, phase contrast, and color brightfield imaging modes. It must include high-sensitivity cameras, automated focus, time-lapse and Z-stacking functions, and be able to scan multi-well plates. The system must accommodate four independent LED light cube illuminators and filters, and include software for image acquisition and analysis. Installation and training are required within 30 days of delivery, which must occur within 60 days of award of the single microscope. A one-year warranty is also required.

View the file

Other files

File Type Posted
Provisions and Clauses Final.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF REQUIREMENTS

Live Cell Imaging Microscope System

I. BACKGROUND

The Biomarker and Genomic Sciences Group within the Biosystems and Biomaterials Division

(644) needs to acquire a live cell imaging microscope with fluorescent capabilities. This microscope is needed for precision medicine/regenerative medicine projects requiring detection of fluorescence expressing chimeric antigen receptor on T cells (CAR-T), chimeric antigen receptor-transduced natural killer cells (CAR-NK) and for visualizing green fluorescent protein

(GFP) expressing adenovirus serotype 5 (Ad5) plaques for the viral vector reference material

(RM) characterization.

II. SPECIFICATIONS

The Contractor shall provide one (1) live cell imaging microscope system that meets the following specifications:

• System must be a standalone integrated unit that includes microscope, digital camera, computer, fluorescence lighting system and LCD display.

• The system must be capable of acquiring images in the following modes: fluorescence, brightfield, color brightfield, and phase contrast.

• The system must be capable of acquiring images with the following methods:

monochrome, single color fluorescence, multi‐color fluorescence, area scan with montage or tile‐stitch, time lapse, Z‐stacking, movie capture.

• The system must include two (2) cameras with auto‐switching capabilities

- An integrated high‐sensitivity CMOS 3.2 MP color camera with (2048 ×1536) pixels with 3.45 μm pixel resolution.

- An integrated high‐sensitivity CMOS 3.2 MP monochrome camera with (2048

×1536) pixels with 3.45 μm pixel resolution.

• The system must include an automated focus mechanism with sub‐micron resolution.

• The system must have time-lapse and Z-stacking capabilities.

• The system must be capable of automated full plate scanning in time‐lapse and/or area scan modes.

• Instrument must have a minimum of a five-position objective turret.

• Instrument must include the following three phase contrast objectives: 4X, 10X, and 20X

• The microscope must have at least four-color fluorescent capabilities and include light cubes for the following three dyes: 4′,6-diamidino-2-phenylindole DAPI (Excitation:

357nm Emission: 447nm), GFP (Excitation: 482nm Emission: 524nm), Texas Red

(Excitation: 585nm Emission: 628nm)

• A universal stage must be included so that a variety of multi-well plates, flasks, or microscope slides can be visualized.

• A digital camera readout must be displayed and incorporated with the microscope.

• The system shall simultaneously accommodate 4 independent high output LED light cube illuminators with integrated hard coated fluorescence band‐pass excitation and emission filters for each cube plus a dedicated transmitted light slot.

• System must automatically recognize fluorescence cube installed and adjust software configuration accordingly.

• Software must be included for image acquisition and analysis.

• The system must simultaneously acquire images with four separate fluorescence channels plus transmitted light.

• The system must include a motorized X/Y scanning stage.

• The system must include one automated objective turret, filter set and condenser.

• The fluorescence LED illuminators must have a lifetime of no less than 50,000 hours at

100% power.

• The LED illuminators must have independent intensity control.

• The system must have a stage designed for a fully integrated onstage incubator for control of temperature, humidity and gases including CO2 and O2.

• The Z‐plane single‐step must have an accuracy of less than or equal to 0.150 μm and the random movement accuracy of less than or equal to 0.750μm.

• Instrument must include an onstage incubator which controls humidity, temperature, and gases (ex. CO2) needed for live cell microscopy and time lapse studies.

• Onstage incubator temperature range must be 22°C to 40°C.

• Onstage incubator CO2 range must be 0 to 20%.

• Onstage humidity range must be greater than or equal to 80% relative humidity at 37°C.

• The time-lapse function must have the capability to be combined with autofocus and or

Z-stack.

• The system must include a scale bar to determine size of sample.

• The system must include automated cell counting and cell confluence calculation capabilities.

• The system must include automated transfection efficiency calculation capability.

• The system must include a white LED light for transmitted illumination.

• The system must support direct output to USB and networked storage.

• The system must provide outputs in the following file formats.

- Images: jpg, tif, and png.

- Movies and time-lapses: avi and wmv.

• The system must include the ability to perform cell counting, confluence and transfection efficiency calculations of multiple images using the same analysis parameters in batch mode.

• An operations and maintenance manual (electronic, hardcopy, or both) shall be provided.

III. DELIVERY

Delivery shall occur within sixty (60) calendar days of award. Delivery shall take place during normal business hours, between 8:30 am and 5:00 pm Eastern Time, Monday through Friday except Federal Holidays, and shall be coordinated with the Contracting Officer Representative

(COR) and Subject Matter Expert (SME). Delivery shall be made to the NIST campus at 100

Bureau Drive, Gaithersburg, MD, 20899, USA.

IV. INSTALLATION

The Contractor shall install the instrument at NIST, Gaithersburg, MD. Installation shall include, at a minimum, uncrating/unpacking of all equipment, set-up and necessary connections for operation, hardware and software installation, basic functionality, instrument start-up, demonstration of instrument performance with the specifications and removal of any trash.

Installation shall take place between 8:30 AM and 5:00 PM Monday through Friday, with the exception of holidays. Installation of the system shall occur within thirty (30) days of the delivery date.

V. TRAINING

In-person training for up to two (2) people shall be coordinated through the COR and SME and shall be completed within 30 calendar days of installation Training shall include operation of the instrument and all features including camera, optics, stage, basic data analysis and file saving and exporting.

VI. WARRANTY

The Contractor shall provide, at minimum, a one (1) year warranty to include all parts, materials, labor, and travel for the instrument to keep the instrument operating to manufacturer's specifications. The warranty period shall commence upon installation of the instrument.

VII. ACCEPTANCE

The Contractor shall demonstrate that performance specifications are met and tested using NIST cell samples for compliance with the parameters of Section II. SPECIFICATIONS. Acceptance shall occur within 7 calendar days of installation.

VIII. DELIVERABLES

Description Quantity Recipient Due Date

Live cell imaging microscope system One COR and SME 60 calendar days after award.

Installation Once COR and SME 30 calendar days after delivery.

Manual (electronic, hardcopy, or both) One COR and SME 30 calendar days after installation.

Training Once COR and SME 30 calendar days after installation.

IX. GENERAL INFORMATION

Safety: The Contractor employee shall be responsible for knowing and complying with commercial installation safety prevention regulations. Such regulations include, but are not limited to, general safety, fire prevention, and waste disposal.

Security: NIST is a restricted campus. An identification badge is required for access for entry into buildings and also is shown to the armed Security Police when entering the campus.

Identification Badges: Contractor employees shall comply with NIST identification and access requirements. The Contractor shall provide the name and citizenship information of employees for on-site visits to the COR prior to arrival. The COR will submit the name(s) and citizenship information to the NIST Office of Security in order for a given Contractor employee to enter the

NIST campus. The Contractor employee is responsible for absences due to missing or expired identification and access documents. Each Contractor employee shall wear a visible identification badge provided by the NIST Security Office.

Vehicle Registration: All Contractor employees must register their vehicles with the NIST

Security Office to gain access to the campus. A valid driver’s license, Government-furnished civilian ID (which in most cases can be a valid driver's license), proof of insurance and current registration must be presented to the NIST Security Office.

File details come from the government source that posted it. Updated .