FDA-SOL-1168378
TruScan RM with Discovery Library
Salient Characteristics
General Requirements:
· Portable Raman spectrometer
· Battery powered
· Laser source of either 780 or 785 nm
· Spectrometer containing no moving parts
· USB-/cable-ready to transfer data to laboratory computers
Additional Specifications:
· Spectrometer: Must cover at least the range of 400 to 2500 cm-1 and have spectral resolution of at least 20 cm-1 (FWHM)
· Laser: 780 or 785 nm with fully adjustable power (e.g., 0 to 250 mW)
· Sampling: Must allow direct focusing as well as have a vial holder
· Integrated Computer/Software that:
· Allows the user to build their own standard sample libraries
· Can output data in .csv and/or .txt formats
· On-board Chemometric Package: Allows more control over instrument functions and measurement/quantitation of complex samples
· Portability: Less than 5 lbs. for entire instrument/integrated computer