Benchtop Uniform Light Source Q and A 21 Nov 24.pdf
PDF 393 KB Posted
- Attached to
- Benchtop Uniform Light Source
- Solicitation number
- FA860125Q0005
About this file
This document is a set of questions and answers regarding a Request for Quote (RFQ) for a Benchtop Uniform Light Source solicitation by the Department of the Air Force Materiel Command Lifecycle Management Center at Wright-Patterson Air Force Base.
The key details are: The solicitation number is FA860125Q0005 with a quote due date of December 5, 2024 at 12:00 PM EST. The scope is to provide a radiometrically accurate, tunable light source solution covering the visible to SWIR spectrums, in accordance with the attached Statement of Work. The solicitation is a 100% Small Business Set-Aside under NAICS 334516. Quotes must conform to the CLIN structure and include all required representations and certifications. Shipping will be FOB Destination. The government will not accept a 3nm spectral accuracy deviation from the required 2.5nm specification.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 04 Min-Req-Light Source.pdf.pdf | ||
| SOW_Light Source_Rev 2.pdf | ||
| Solicitation - FA860125Q0005.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
AIR FORCE LIFE CYCLE MANAGEMENT CENTER (AFMC)
WRIGHT-PATTERSON AIR FORCE BASE OHIO
21 November 2024
AFIT Benchtop Uniform Light Source Question & Answer
AGENCY AND CONTRACTING ACTIVITY: Operational Contracting Branch, Installation Contracting Division, Directorate of Contracting, Air Force Life Cycle Management Center, Air Force Material Command, Department of the Air Force.
AFLCMC/PZIBB
1940 Allbrook Dr.
Bldg. 1, Room 109 Wright-Patterson AFB, OH 45433-5344
Question: The spectral accuracy requirement in your document is listed as ±2.5nm. However, our device can only achieve ±3nm. Would this deviation be acceptable?
Answer: According to our SME, the below deviation is not acceptable. The application requires small bandwidth light sources and 2.5 nanometers is already pushing the limits for this application.