X-Ray_Requirements_Document_PTPLF_13_May_19.pdf
PDF 371 KB Posted
- Attached to
- Wavelength Dispersive X-Ray Fluorescence Spectrometer Federal contract opportunity
- Solicitation number
- FA860119QA059
About this file
UPDATED STATEMENT OF WORK
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Combined_solicitation-synopsitation.pdf | ||
| X-Ray_Requirements_Document_30_May_2019.pdf | ||
| Combined_solicitation-synopsitation.pdf | ||
| X-RAY_SOW.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Wavelength Dispersive X-Ray Fluorescence Spectrometer (WDXRF)
Requirement
20 February 2019
The procurement is for one benchtop Wavelength Dispersive X-Ray Fluorescence Spectrometer (WDXRF) instrument that will be used at AFPET laboratories. Aerospace Fuels Laboratories perform quality control testing for
Aviation Gasoline 100LL grade aviation fuel.
WDXRF instrument shall at a minimum:
1. Be a benchtop instrument with approximate size of 24” Width x 28” Depth x 28” Height and an input voltage requirement of 220 VAC.
2. Be a fully automated sequential WDXRF.
3. Does not require Helium or P-10 gas for the operation of the instrument.
4. Provide a stable and rapid vacuum system capable of maintaining optics in vacuum while the sample chamber remains in air.
5. Meets all the requirements in ASTM D5059-14, Standard Test Methods for Lead in Gasoline by
X-Ray Spectroscopy, Test Method A and/or B.
i. In accordance with T.O. 42B-1-1, “Quality Control Of Fuels and Lubricants”, page 1.1, paragraph 1.1; MIL-STD-3004D, “Quality Assurance/Surveillance For Fuels, Lubricants and Related Products” and ASTM D910-17 “Standard Specification for Leaded Aviation Gasolines”, 100LL grade.
6. Meets all the requirements in ASTM D2622-16, Standard Test Methods for Sulfur in Petroleum Products by Wavelength Dispersive X-Ray Fluorescence Spectrometry.
i. In accordance with T.O. 42B-1-1, “Quality Control Of Fuels and Lubricants”, page 1.1, paragraph 1.1; MIL-STD-3004D, “Quality Assurance/Surveillance For Fuels, Lubricants and Related Products” and the following fuel specifications:
1. ASTM D910-17 “Standard Specification for Leaded Aviation Gasolines”, 100LL grade.
2. ASTM D1655-18, “Standard Specification for Aviation Fuels.”
3. MIL-DTL-83133K, “Turbine Fuel, Aviation, Kerosene Type, JP-8 (NATO F-34), NATO F-35, and JP-8+100 (NATO F-37).”
4. MIL-DTL-25524F, “Turbine Fuel, Aviation, Thermally Stable.”
5. ASTM D975-19,”Standard Specification for Diesel Fuel Oils”, Paragraph 5.1.8.
7. Include measurement, quantification, statistical analysis and reporting software with simplified user interfaces that maximize instrument performance and quality of data output.
i. Shall have a dedicated quantification method for the analysis of Tetraethyl Lead (TEL) in Aviation Gasoline that meets the requirements in ASTM D5059-14 and ASTM D910-17.
ii. Shall have a dedicated quantification method for the analysis of Sulfur in Aviation Gasoline that meets the requirements in ASTM D2622-16 and ASTM D910-17, ASTM D975-19, ASTM D1655-18, MIL-DTL-83133K and MIL-DTL-25524F.
8. Utilize a scintillation counter detector for heavy elements and a sealed proportional counter detector for light element analysis.
9. Have the ability to utilize two or more crystals and provide the equivalent of a LiF200 crystal for Ti-U and a PET crystal for Al-Sc.
10. Utilizes an air-cooled 200 watt x-ray tube with Pd target.
i. Does not require water or liquid nitrogen for cooling.
ii. Includes automatic bean filter to remove Pd-K spectral interference lines.
11. Provide an automatic sample changer with multiple sample positions.
12. Includes sample spinning functionality for optimized sampling statistics.
13. Provide a multipurpose sample cell designed to hold liquids, powders and beads.
14. Provide all required x-ray safety and warning signs, stickers and handbooks.
15. Include a minimum of two days on-site for:
i. Instrument set-up and calibration by technician.
ii. Training for both instrument operation and method development.
16. Include vendor-supplied classroom training for x-ray theory as well as specific training consistent with aviation and ground fuel applications.
17. Include the option for a post-warranty annual service contract that includes:
i. All parts, labor, shipping, travel and travel expenses for one preventative maintenance service per year and emergency repair service as required (consumable items including x-ray tubes, o-rings, seals and roughing pump oil are not included).
ii. Technical telephone support 24 hours a day, 365 days/year
iii. Emergency on-site service within 72 hours.
18. Delivery within 3 months after receipt of award to the following address:
AFPET Labs
RAF Mildenhall BLDG 1546
Bury St. Edmunds GB IP28 8NF
Note: ASTM documents are generally available at University libraries or contact ASTM by writing; ASTM
International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959; by telephone at 610-
832-9255; by e-mail at service@astm.org; or through the ASTM website at www.astm.org mailto:service@astm.org http://www.astm.org/
File details come from the government source that posted it.