STATEMENT OF WORK
Digital Camera Intraoral Dental
· Introduction: The 88TH Medical Treatment Facility (MTF) located at Wright–Patterson AFB, OH requires Digital Camera Intraoral Dental Qty 10.
· Background: This camera will be used for residency training, specifically for intra-oral photography, comprehensive treatment planning, documentation of patient care, patient education, and case presentations.
· Delivery Date:
· Requested delivery date - 14 business days after award date
· Building address: ATTN:MEMO 4881 Sugar Maple Drive Bldg. 830 Area A
WPAFB OH, 45433-5529
· Minimum Requirements: Digital Camera Intraoral Dental Qty 10
Size – Hand held
Design – Water, chemical, and scratch resistant to allow for proper cleaning Weight – No greater than three (3) lbs.
Performance Requirements –
· Photography capabilities to include:
· Standard and distance intra-oral photography;
· Ability to photograph with use of mirrors (i.e. image can be reversed);
· Ability to photograph teeth for optimal shade matching;
· Ability to shoot in high resolution and close up magnification (macro mode or manual mode);
· Macro lens between 49 mm and 100 mm;
· Flash diffuser or similar function;
· Shutter speed between 1/1000 and 1/3000;
· ISO between 100-800;
· SD card capabilities; and
· Image size a minimum of 8 megapixel.
Accessories –
· Camera cases Qty 10 The following items will be included per camera:
· Required batteries
· User manual
Warranty – Minimum 1 year warranty for replacement or repair of main equipment (camera) if it becomes defective. The warranty does not have to include accessories or normal wear and tear.