This document summarizes a federal solicitation for a scanning transmission electron microscope. The U.S. Army Corps of Engineers Engineer Research and Development Center Geotechnical and Structures Laboratory requires a scanning transmission electron microscope to meet specified technical requirements for research in nanoscale particle analysis as part of its Ground-based Higher Order Geophysics program. Quotes are due by email to the point of contact by the date listed. The award will be a firm-fixed-price contract for the microscope equipment, twelve months of software licensing and support, installation, training, and warranty. The evaluation will consider price and technical acceptability of the quote to meet the stated requirements.

View the file

Other files

File Type Posted
Quote Submission Form.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Description of Requirement (Scanning) Transmission Electron Microscope

SPECIFICATIONS:

The Engineer Research and Development Center (ERDC), Geotechnical and Structures Lab (GSL) requires a (Scanning) Transmission Electron Microscope with the following features/capabilities:

The core (S)TEM instrument must meet the following minimum specifications:

Electron emitter source

• 200kV FEG STEM/TEM with flexible high tension (20, 40, 80, 120, 160, 200 kV)

• Electron emitter should be of a Schottky high brightness long stability type and should have brightness of 1.8x109 A/cm2sr at 200kV

• Probe current should be >2nA/1nm probe with maximum beam current >50nA available at 200kV

• The microscope should have a fixed pole piece with a gap of at least 5mm that is used to satisfy all of the resolution requirements as well as the tilt specifications

STEM

• System should have a scanning transmission attachment (STEM unit, BF and HAADF detectors) with resolution of HAADF 0.16nm or better, 3-fold condenser astigmatism correction should be available

• At least 12 parallel imaging channels should be available to acquire 12 or more images at the same time from different STEM detectors

• Drift Corrected Image Integration (DCFI) option should be available as part of the standard microscope user interface

• At least one of the DF STEM detectors should have 8 segments with capability to read each segment individually.

• The STEM acquisition software should be written by the column manufacturer, not an accessory supplier.

TEM

• Information limit should be ≤ 0.12nm

• Camera should be manufactured and supported by the column manufacturer.

• Camera should have ≥ 16M pixels

• Camera speed should be 25fps at 512x512pixels frame. Camera dynamic range should be 16 bit

• Camera should be usable at 20-200kV HT range

• Camera should be retractable

Goniometer

• Minimum specimen tilt angle (α/β) ±30 deg./± 30 deg with a double tilt holder

• Microscope should have a fully eucentric goniometer with all 5 axis motorized TEM stage should be stable with drift 0.5nm/min or better

• Goniometer maximum tilt should be +/-90 degrees, tilt speed from -90 degrees to +90 degrees should be less than 5 seconds.

• X movement range – 2mm in total or more Y movement range – 2mm in total or more

• Z movement range – 0.75mm in total or more

• Maximum tilt for tomography application should be achieved with a standard 3.0 mm grid.

• STEM Tomography option should be automated with dynamic focusing capability

Vacuum

• Microscope should have a completely oil free vacuum system.

• FEG gun vacuum should be < 10-7Pa

• TEM column vacuum should be <10-6Pa

• The column should have a differential pumping aperture not exceeding 200 micron between the camera chamber and the column to keep the vacuum clean in the column.

Diffraction

• Minimum spot size (in Nano-Diffraction mode) should be 0.3nm or smaller Camera length range (Selected area diffraction) 14 - 5700 mm

• Maximum convergence angle should be >100mrad or more

• Maximum diffraction acceptance angle should be 26 degree or more

EDX system

• EDX system should have more than 4 detector crystals placed symmetrically around the sample rod axis. EDX solid angle should be 0.9sr or more

• EDX detector should be windowless to increase the sensitivity for light elements EDX resolution should be at least 136eV at 10,000 spectra/s or better

• EDX speed for output count rate should be 1,000,000 cps or better

• EDX mapping speed should have pixel dwell time 10 micro-seconds or better

• Anti-contamination device (LN2) should be provided with the time without refill for 4 days or longer EDX should have a Fiori number >4000

• EDX 3D tomography operation should have an option to be AUTOMATED and UNATTENDED during the series acquisition

• Tomography series tilt range should be +/- 75 degrees

• EDS tomography acquisition software should be written by the column manufacturer, and include the ability to do on the fly absorption correction and holder shadowing for accurate quantification.

Operation:

• Microscope should be operated remotely to minimize the interference from operator and maximize operator efficiency;

• Microscope should have the option for AUTOMATED update on the operation status and history sent to the user regularly with the chosen frequency in a range from minute – to day or week

• Microscope should have an enclosure to protect hardware from environment, and reduce system sensitivity to drafts.

Software:

• All software used to operate the instrument, acquire and process the data should be based on 64-bit platform such as Win 10 or compatible.

• The image and spectrum acquisition software should be written by the microscope column manufacturer, not a third-party accessory manufacturer.

Additional items must include (at a minimum):

• Workstation with 24’’ LCD Monitor

• 115 V, 50/60 Hz Compressor:

o Silent, automatic compressor with start relay and capacitor, thermal protection and intake filter. Provides clean, oil-free and water-free compressed air at 6 Bar (max. 37 Liter/minute), with a noise level of 40 dB(A) at one meter.

o Direct-drive 0.34 kW motor mounted on a 25-liter reservoir. Complete with manometer, control valve, pressure switch, air admittance valve and safety valve.

o Dimensions 375 x 375 x 540 mm (l x w x h) o Weight 30.5 kg o Power consumption 0.34 kW

• Cooling Unit (Water):

o Cooling capacity: 3.9 kW (at 20 °C supply water temp) o Temperature stability: ±0.05 °C o Mains supply: 208-230 V- 1 Phase - 60 Hz : 230 V - 1 phase - 50 Hz o Water pump capacity: 13.6 l/min at 4 bar o Refrigerant: R407C o Dimensions (WxDxH): 15 x 30 x 27 (inch) or 39 x 77 x 69 (cm) o Manufacturer and model: Haskris, Water-cooling unit, LX2-W5-7XXXXX-4-M144

• Mains Matching Transformer for TEM : 190 - 440 volt, 50 - 60 Hz

• Single-tilt Holder

• High-visibility Low-background Double-tilt Holder

• Tomography Software should include (mimimum):

o TEM Tomography Data Acquisition o STEM Tomography Data Acquisition o EDX Tomography Data Acquisition o A user-friendly solution for alignment of a tomographic series of images, 3D reconstruction of the data, as well as basic visualization functionality.

• Must include visualization and analysis software for advanced visualization and analysis of reconstructed tomography data, acquired using electron and correlative microscopy systems.

• Manufacturer must have a proven history (minimum of 5-yrs) of (S)TEM development, installation and support, to include 24/7 customer support/operation centers.

Quote shall include shipping, installation and warranty, software licensing (12 months) and training required to ensure the equipment is fully functional.

DELIVERY DATE:

Equipment shall be delivered and installed no later than 6 months after award. Training and installation shall be provided within 60 days after delivery.

DELIVERY LOCATION:

Attn: Oliver-Denzil S. Taylor Engineer Research and Development Center Geotechnical and Structures Laboratory 3909 Halls Ferry Road Vicksburg MS, 39180