B08_ATCH_1_SALIENTS.pdf
PDF 188 KB Posted
- Attached to
- PE Churn Sample Splitters (8L) IAW Salients
- Solicitation number
- 140G0125Q0043
About this file
This document is a Salient Characteristics specification for an 8-liter (2-gallon) churn sample splitter required by the USGS Hydrologic Instrumentation Facility (HIF) in Tuscaloosa, AL. The equipment will be used at water data collection sites nationwide to combine water and sediment samples from stream flow to create composite representative samples for analysis.
The technical specifications require a polyethylene cylindrical vessel that meets USGS water sampling protocol, with an 8-liter minimum capacity, disc churn with holes and orienting notch for optimal mixing, and a spring-loaded spigot with full open lever stop. The unit must include a lid that allows the handle to pass through, replaceable components (lid, churn, and spigot), and a heavy-duty swing handle capable of supporting the churn when full or empty. Size requirements specify maximum dimensions of 22cm (8-1/2 inches) diameter and height of 28cm (11 inches), extending to 41cm (16-1/8 inches) with the handle upright.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sol_140G0125Q0043.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
The Hydrologic Instrumentation Facility (HIF) of the USGS located Tuscaloosa, AL needs to procure churn sample splitters. This equipment will be used at water data collection sites all over the country. They will be used to combine water and sediment samples collected from the stream flow to make a composite representative sample of the water for analysis.
CHURN SAMPLER SPLITTER, 8 LITER (2 GALLONS)
SALIENT CHARACTERISTICS:
1. Shall be a polyethylene cylindrical vessel.
2. Shall meet USGS water sampling protocol.
3. Shall be capable of holding a minimum of eight (8) liters (2 gallons).
4. Shall include disc churn with holes and orienting notch for optimal mixing of samples.
5. Shall include lid and allow handle of churn to pass through.
6. Shall include spring-loaded spigot with full open lever stop for dispensing subsamples.
7. Lid, churn, and spigot shall be replaceable.
8. Shall have a heavy duty, swing type handle capable of supporting full or empty churn.
9. Shall be no larger than 22cm (8-1/2 inches) diameter.
10. Shall be no higher than 28cm (11 inches) and 41cm (16-1/8 inches) with handle in upright position.