LABORATORY UV-Vis HYPERSPECTRAL CAMERA

Salient Characteristics

1 ea. UV-Visible Hyperspectral Camera:

1. The system shall be a hyperspectral camera to measure radiance, irradiance, or scalar irradiance in VIS ranges.

2. Shall utilize USB 3.0 communications for interfacing to a laptop or PC.

3. Shall include a CS-mount lens mount.

4. Shall have at least 255 spectral channels.

5. Shall have a pixel size of 5.86 µm or smaller.

6. Shall have a Full Width Half Maximum (FWHM) resolution less than or equal to 2.8 nm or better.

7. Shall have a bandwidth of less than or equal to 1.8 nm

8. Shall be capable of operating in a laboratory environment.

9. Shall be a near-UV-Visible range system capable of operating in the spectral range from

330 to 800 nm range or better.

10. Shall include all accessories required for operation when attached to a microscope such as, but not limited to, communication cable(s), software program(s) to operate the camera and analyze data, calibration data, and manual.