Attachment A - Salient Characteristics V-Shell Blender OR EQUAL.pdf
PDF 644 KB Posted
- Attached to
- Globepharma V-Shell Blender; OR EQUAL
- Solicitation number
- 36C25821Q0381
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment C - CONTRACT CLAUSES.docx | DOCX document | |
| Attachment B FMS Vendor File Update Request Form.docx | DOCX document | |
| Attachment D - Price and Delivery Schedule.docx | DOCX document | |
| 36C25821Q0381_1 COMBO.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Supply Item Brand Name or Equal to Description:
Brand Name or Equal to: Globepharma Description: Simpleblend 3 Lab/Scale Blender Salient Characteristics In order to be considered for award the V-Shell Blender Requirement must be equivalent to a Globepharma 3 Lab-Scale Blender and must have the following salient characteristics.
One (1) EA- SIMPLEBLEND – 3 Lab-Scale Blender Features:
o All stainless steel construction and on casters o Control box with Touch Screen HMI to set times and start-stop the system o Direct Drive o Optional intensifier bar o Optional integrated SIFT-N-BLEND attachment for eliminating the pre-screening step before blending o Optional Safety-interlocked guard o Sanitary butterfly valve o (1) EA, Wire mesh screen (316 SS, #30 mesh) o (1) EA, Wire mesh screen (316 SS, #40 mesh) o 1 year warranty
TECHNICAL SPECS:
o Shell Working capacity – 3 cu.ft. (86.4 L), Total Volume: 5.31 cu. ft.
o Standard Shell fixed speed – 20 rpm, 1 hp o Intensifier bar tip speed – 2900 fpm o Intensifier bar blade diameter – 9.25 inches o Electrical – 208 V, 3 phase, 60 Hz o Stand, drive shaft, and caster material – 304 stainless steel o Shell material – 316 stainless steel o Shell finish – Ra #4 finish outside, Ra #4 - #7 finish inside o cGMP valve material – 316L Stainless steel (#4 finish on valve body and #7 finish on vane, valve lining is made with FDA-approved platinum-cured silicone rubber) o Shell gasket material – FDA-approved white nitrile rubber o Safety guard material – Aluminum frame, ½” acrylic panels
One (1) EA – Documentation and Testing:
o Installation Qualification (IQ) and Operational Qualification (OQ) protocol testing o Factory Acceptance Testing (FAT) o Site Acceptance Testing (SAT)
One (1) EA – Delivery
Period of Performance Information
Period of performance will be 150 days from contract award.
Vendor must schedule installation and training with SME or COR within 30 days after delivery of equipment.
Delivery Information All items shall be shipped to the customer at the following location:
Address: VA Cooperative Studies Program Clinical Research Pharmacy Coordinating Center
2401 Centre Avenue, SE ATTN: CSP Warehouse (151-I) Albuquerque, NM 87106-4180
File details come from the government source that posted it. Updated .