ATTACHMENT A - SALIENT CHARACTERISTICS - SPECTROPHOTOMETER.docx
DOCX document 17 KB Posted
- Attached to
- 6640--SPECTROPHOTOMETER
- Solicitation number
- 36C26224Q1804
About this file
This document is a Request for Quotation (RFQ) issued by the Department of Veterans Affairs (VA) Veterans Health Administration for a spectrophotometer (fluorescence and absorbance spectrometer) for the VA San Diego Healthcare System.
The RFQ includes the salient characteristics for the required equipment, which must have both absorbance and fluorescence capabilities, detect wavelengths from 250-1,100 nm, allow for wavelength filtering, and transmit results to a computer with 3D visualization software. The contractor must also provide a laptop and technical support/training. The RFQ is a one-time purchase with a 30-day delivery timeframe after receipt of order. Offers must be received by September 6, 2024 and demonstrate that any proposed "equal" products meet all the salient characteristics. The award will be made to the responsible offeror submitting the lowest priced quotation that meets all requirements.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| ATTACHMENT B - SCHEDULE OF PRICING - SPECTROPHOTOMETER.xlsx | XLSX spreadsheet | |
| 36C26224Q1804.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA San Diego Healthcare System Spectrophotometer (Fluorescence and Absorbance Spectrometer)
1. BRAND NAME OR EQUAL
| BRAND |
| PRODUCT DESCRIPTION |
| PRODUCT # |
| QTY |
| Horiba Scientific |
| Fluorescence and Absorbance Spectrometer |
Duetta
| Horiba Scientific |
| Duetta Laptop |
Duetta
| Horiba Scientific |
| Service (Technical Support and Training) |
N/A
As Needed
2. SALIENT CHARACTERISTICS
1. The equipment must have both absorbance and fluorescence capabilities that can be performed independently and simultaneously.
2. The equipment must be capable of detecting fluorescence wave lengths from 250 to 1,100 nm.
3. The equipment must be capable of detecting excitation and absorbance wave lengths from 250 to 1,000 nm.
4. The equipment must allow the user to program for the filtering of wavelengths.
5. The equipment must be capable of transmitting the results of antibody conjugation test to a computer.
a. The software/program used to display the result must be capable of displaying them in a 3-dimensional format.
6. The contractor must provide a laptop that will be used to support the spectrophotometer and required testing activities.
a. The software/program required to view and analyze test results must be pre-installed on the computer.
b. The laptop must run on window 7, 8, or 10 operating system
7. The equipment must have a charge coupled device (CCD).
8. The equipment must be compliant with FDA requirements for laboratory equipment used for animal testing.
9. The contractor shall provide technical support and on-site training on an as-needed basis.
10. The contractor shall provide a 1-year or manufacturer warranty, which ever is greater, for all parts and labor.