Point of Care Ultrasound System shall meet or exceed the following salient characteristics:
· Shall include an embedded metal jacket armored cables to protect transducers from common abuse accident scenarios, like being rolled over, stepped on, or twisted, and help maintain image quality over the life of transducers.
· Touchscreen-based user interface
· Stand with storage basket, tray with lockable drawer, gel bottle holder, an AC cord retainer, casters, and cups.
· Video output (HDMI) for external video monitor connection and allows printing to a thermal printer.
· This device shall be able to accurately conduct Tissue Harmonic Imaging (THI)
· Operating System: Windows 10 IoT Enterprise 2019 LTSC
· Gray Scale: 256 shades
· Dynamic Range: Up to 183 dB
· Transducer connection: 3 x connections
· HIPAA Compliance: Meets Federal Information Processing Standards (FIPS) for Moderate security control selections per FIPS199, FIPS200, and NIST 800- 53.
· The transducers shall withstand a drop from a height of 1 meter and be safe and function normally after this drop.
· Imaging Modes:
· 2D/B-mode
· Broadband Imaging
· Tissue Harmonic Imaging
· Pulse Inversion Harmonic Imaging
· M-mode
· Color Velocity Doppler (CVD)
· Color Power Doppler (CPD)
· Color Variance (Cardiac only)
· Pulsed Wave Doppler (PW)
· Tissue Doppler Imaging (TDI) (Cardiac Only)
· Continuous Wave Doppler (CW) (Cardiac Only)
· B-Mode Imaging:
· B- mode Controls:
· Depth
· Gain
· Time Gain Compensation (TGC)
· Auto Gain
· 2D Mode THI
· 2D Image Optimization (Res, Gen, Pen)
· Dynamic Range (Live and Frozen)
· Image Orientation (U/R, U/L, D/L, D/R)
· Sector width control (Steerable, Cardiac Only)
· Thermal Index (TIS/TIB/TIC)
· LVO
· Output Power Control
· Support imaging frame rates up to 100Hz.
· Needle visualization without image degradation (Shallow, Medium and Steep settings)
· Shall include the following transducers:
· C5-1: 5-1 MHz multi-frequency, broadband curved transducer. Maximum Scan Depth: 30 CM. Exams: Nerve, Musculoskeletal, Spine
· L12-3: 12-3 MHz multi-frequency, broadband, linear transducer. Maximum Scan Depth: 9 cm. Exams: Arterial, Breast, Carotid, MSK Nerve, Superficial, Venous, PIV
· L19-5: 19-5 MHz multi-frequency, broadband, linear transducer Maximum Scan Depth: 6 cm. Exams: Arterial, MSK, Nerve, Superficial, Venous, PIV
· P5-1: 5-1 MHz multi-frequency, broadband, curved transducer Maximum Scan Depth: 35 cm. Exams: Abdomen, Cardiac, Focused Cardiac, Lung, Obstetrics, Orbital, and Trans Cranial Doppler (TCD)
· Spectral Doppler Imaging:
· Doppler modes:
· Pulsed wave (PW)
· Tissue Doppler Imaging (TDI) (Cardiac Only)
· Continuous Wave (CW) (Cardiac Only)
· Doppler controls:
· Doppler Gain
· Doppler Gate Size (PW/TDI)
· Doppler Invert
· Doppler Auto Trace Type (Peak/Mean/Above/Below/All)
· Doppler Display Format - 1/3 and 2/3, 1/2 and 1/2, 2/3 and 1/3, side by side and full screen duplex
· Doppler Scale
· Doppler Steer Angle (Linear only)
· Doppler Sweep Speed
· Doppler Baseline
· Doppler Angle Correct
· Doppler Wall Filter
· Simultaneous (Linear Transducers Only)
· 8x live Read Zoom capability
· Write Zoom on C5-1, C10-3, IC10-3, T8-3 and P5-1 transducers (Abdomen, OB, Early OB, Gynecology, Prostate and Cardiac exam types only)
· 4x Frozen post processing Zoom
· Dual Imaging
· Centerline graphic on display for linear and curvilinear transducers
· On screen needle guide support for IC10-3 and L19-5 Attachment A Salient Characteristics
RFQ # 36C24823Q0861
· Auto Gain adjust and reset
· Procedure Mode
Maneuverable Point of Care Ultrasound System shall meet or exceed the following salient characteristics:
| • | Shall include an embedded metal jacket armored cables to protect transducers from common abuse accident scenarios, like being rolled over, stepped on, or twisted, and help maintain image quality over the life of transducers |
| • | Touchscreen-based user interface |
| • | Stand is with storage basket, tray with lockable drawer, gel bottle holder, an AC cord retainer to prevent accidental removal of the AC power cord, casters, and cups. |
| • | Has a video output (HDMI) for external video monitor connection and allows printing to a thermal printer. |
| • | This device shall be able to accurately conduct Tissue Harmonic Imaging (THI), a capability in which the system transmits at one frequency and receives at a higher harmonic frequency to reduce noise and clutter and improve vessel and tissue resolution. |
| • | Needle visualization without image degradation (Shallow, Medium and Steep settings) |
| • | Shall include the following probes: |
| o | L25x – 13-6 MHz multi-frequency, broadband, 25 mm linear array |
transducer. Biopsy compatible Exams: Nerve, Musculoskeletal, Arterial, Venous B-mode Imaging
| o | B-Mode Controls include: |
| • | Depth, 2D Mode (Near, Far, and Overall Gain), Autogain, 2D Mode THI, 2D optimization (Res, Gen, Pen), Dynamic Range (7), Image Orientation (U/R, U/L, D/L, D/R) |
| o | Support acquisition frame rate up to 150Hz. |
| o | Tissue harmonic imaging (THI) transmit at a low frequency and receive at a higher harmonic frequency to reduce clutter, improve lateral resolution and enhance tissue contrast. |
| o | Up to 2x pan/zoom |
| o | Supports up to three focal zones to improve resolution uniformity |
| o | Support individual live scanning on each side of the dual imaging display |
| o | Support seamless dual imaging display as extended field of view |
| • | It shall be optimized for cardiac, lung and abdominal imaging and include B-mode, M- mode and Color Doppler imaging. |
| • | Power docking station for the Universal stand. This stand module and docking station allows the ultrasound systems to charge without plugging in the power |
| • | Shall have the ability to include CW and PW Doppler. |
| • | Shall have a Cardiac Calculations Package. |
| • | Due to COVID-19 procedures for cleaning, this device shall withstand harsh cleaning chemicals and sustain a one-meter (3-feet) drop for durability. |