About this file
This document is a Technical Specifications sheet for the procurement of an Oxygen/Carbon Dioxide Gas Analyzer for the Grand Forks Human Nutrition Research Center (GFHNRC). The gas analyzer is a direct replacement unit for an existing room calorimeter and must maintain continuity of ongoing research, data comparability, and integration with current measurement and control systems. The specified delivery location is 2420 2nd Avenue North, Grand Forks, ND 58203, with delivery required within 60 days of award. The facility operates Monday through Friday from 6:00 AM to 6:00 PM, excluding federal holidays.
The analyzer must simultaneously measure both oxygen and carbon dioxide using a single unit equipped with a double-layer detector, optical coupler, and dual-beam Non-Dispersive Infrared (NDIR) sensors with paramagnetic alternating pressure methods. Carbon dioxide measuring ranges must support a minimum range of 0–1,000 vpm and maximum range of 0–10,000 vpm with an operating range of 0–1 percent, while oxygen measuring ranges must support a minimum range of 0–0.5 percent and maximum range of 0–100 percent with an operating range of 20–21 percent. The unit must maintain absolute linearity and feature 6 mm or 0.25-inch inlet and outlet connections compatible with stainless steel Swagelok connections. The analyzer must be configured as a 19-inch rack unit with RS-485, PROFIBUS, and SIPROM GA interface compatibility and powered by a 100–120 AC, 48–68 Hz power supply.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 1 - Solicitation Terms and Conditions (0001).pdf | ||
| Attachment 3 - Quote Submission Form.pdf | ||
| Attachment 1 - Solicitation Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical Specifications Oxygen/Carbon Dioxide Gas Analyzer
GENERAL INFORMATION
A requirement has been made for the purchase of a new gas analyzer for a room calorimeter at the Grand Forks Human Nutrition Research Center (GFHNRC). This is a direct replacement gas analyzer for the existing room calorimeter to maintain continuity of ongoing research, data comparability, and integration with the current measurement/control architecture.
REQUIREMENTS
Technical Requirements:
• Must be able to measure both oxygen and carbon dioxide with the same unit.
• Must have high selectivity with a double-layer detector and optical coupler.
• Must have dual-beam Non-Dispersive Infrared (NDIR) sensors with paramagnetic alternating pressure methods.
• The smallest measuring range for carbon dioxide must be between 0 - 1000 vpm and largest range must be between 0 - 10,000 vpm with an operating range between 0 - 1%.
• The smallest measuring range for oxygen must be between 0 - 0.5% and the largest range between 0 - 100% with an operating range between 20 - 21%.
• Must be able to maintain absolute linearity.
• Inlet and outlet connections for sample gas must be 6 mm or 0.25 inch and compatible with stainless steel Swagelok connections.
• Interface must be RS-485, PROFIBUS, and SIPROM GA compatible.
• Must be a 19” rack unit.
• Must have a 100-120 AC, 48-68 Hz power supply.
DELIVERY/SHIPPING INFORMATION
Grand Fork Human Nutrition Research Center 2420 2nd Avenue North, Grand Forks, ND 58203 Hours of operation: Monday – Friday 6AM – 6PM; except federal holidays Delivery is to be made within 60 days of award.