This Statement of Work (SOW) details the procurement of a KingFisher™ Flex 96 DW Head and Heating Block processor for an aquatic animal health research unit. The equipment is required to improve high-throughput DNA/RNA extractions from pathogens affecting aquatic species including shrimp, crawfish, crustaceans, and finfish, supporting critical research in the American aquaculture industry.

The specifications include a benchtop instrument with medium to high throughput, capable of processing 24 or 96 samples per run using various plate formats (PCR, skirted, deep-well) with volumes ranging from 20-5,000 μL. Key features include BindIt software for customizable protocols, heating/cooling ranges from 5°C above ambient to 115°C, and standalone functionality. The procurement includes VWR pricing and a comprehensive AB assurance plan covering installation, training, on-site response, maintenance, remote monitoring, and priority technical support.

View the file

Other files

File Type Posted
RFQ_1232SA25Q0760.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work (SOW)

Project: Kingfisher Flex – 96 DW Head and Heating Block

GENERAL INFORMATION

1.0 Scope of Work:

We seek to obtain a KingFisher™ Flex 96 DW Head and Heating Block processor.

The requested equipment will provide highly versatile, automated processing for DNA, RNA, proteins and cell samples of up to 96 samples per run. The Automated KingFisher™ Flex processor will aid to improve high throughput DNA/RNA extractions collected from pathogens affecting shrimp, crawfish, crustaceans and finfish samples to address issues concerning mission critical research supporting the American aquaculture industry.

2.0 Background:

The aquatic animal health research unit is requesting to purchase the KingFisher™ Flex processor to improve high throughput DNA/RNA extractions collected from pathogens affecting shrimp, crawfish crustacean and finfish samples to address issues concerning mission critical research. We have provided VWR pricing for our market research. The specifications are the unit supplied by VWR (cat no. 95041-912, cat

no. MISCH-TECHSERV (Smart start kingfisher,1 day installation and training), cat no.

MISC-TECHSERV-EX (AB assurance Plan includes: 2-day on-site response, planned maintenance, high priority remote service center support, remote instrument monitoring and diagnostics, on-site service, labor on-site service, parts on-site service, travel factory certified replacement parts, engineer labor and travel included for repair visits, priority technical phone support).

2.1 Specifications:

1. Instrument size - Benchtop

2. Throughput level - Medium to high

3. Processing volume by plate format - PCR plate (20–100 μL), skirted (or similar skirted PCR plate), 20–200 μL: 96-well plate, 50–1,000 μL: 96 deep-well plate, 200–5,000 μL: 24 deep-well plate

4. Samples per run – 24 or 96

5. Software included – BindIt

6. Customizable protocols – Yes, with (BinIt software)

7. Heating/cooling ranges – From 5ºC above ambient temperature to 115ºC.

8. Ultrviolet lamps – No.

9. Standalone use – Yes.

10. Size – 68 x 60 x 38 cm; 26.77 x 23.6 x 14.96 in. (W x D x H)

11. Weight – 28 kg; 62 lbs