This attachment provides technical specifications for a solicitation seeking an infrared interferometer for atmospheric sounding applications. Specifically, it requests a Fourier Transform Spectrometer (FTS) spectroradiometer complete system from brand LR Tech, Inc. or equal. The FTS controller must include three optimized scan speeds including 32KHz (1 scan/sec.), 100KHz (2.8 scans/sec.), and 375KHz (10 scans/sec.). It must also meet mirror speed stability requirements of 5% rms at 100KHz and 1% rms at 0.9987. The system must be capable of operating within an ambient to 60C temperature range and provide radiance uncertainty. The related solicitation was issued by the Department of Commerce National Oceanic and Atmospheric Administration seeking a spectroradiometer with number 1305M320QNRMJ0045.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ 1305M320QNRMJ0045.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 1 Specifications
TECHNICAL SPECIFICATIONS:
CLIN 0001:
Brand name or equal to LR Tech, Inc. ASSIST II: infrared interferometer for atmospheric sounding application
To include:
-Fourier Transform Spectrometer (FTS) spectroradiometer complete system.
FTS controller with 3 optimized scan speeds (spectral resolution= 1cm-1): 32KHz (1 scan/sec.), 100KHz (2.8 scans/sec.), 375KHz (10 scan/sec.)
Mirror speed stability: 5% rms at 100KHz, 1% rms at <50KHz -Interferometer
Maximum optical path difference (OPD) of 1 cm Spectral sampling at least 0.5 cm-1 Spectral range: at least 525 to 5,000 cm-1 Single plate KBr beamsplitter (self compensated);
Maximum beam divergence of 60 milliradians full angle;
Throughput: at least 0.00805 sr cm2 HeNe laser metrology system (for digital sampling and mirror velocity control) Sampling rate (laser fringe rate: at least 32khz to 100 kHz, computer selectable) Field of View (45 mrad or better) Noise Equivalent Spectral Radiances (NESR): For ambient target (25℃), Resolution
16cm-1, and Obs. Time 1 sec.:
MCT: 2.5 x10-9 W/cm2 sr cm-1 Spectral band: 900 to 1100 cm-1 InSb: 2.5 x 10-10 W/cm2 sr cm-1 Spectral band: 2000 to 2300 cm-1
-Radiometric calibration Units Two (2) high-emissivity blackbodies operating at different temperatures, viewed at regular intervals to calibrate the observed sky spectra.
High precision temperature controller which can regulate the temperature of two channels independently (blackbodies) with cavity temperature measured at least at 3 different locations
Must have NIST calibration transferability certificates for each blackbody.
Effective Emissivity > 0.9987 Temperature Range, ambient to 60°C Radiance uncertainty <±0.005 K Temperature Stability at least 0.03°C over 120 sec.
Temperature Gradient <0.35°C A third high emissivity blackbody and controller for maintenance and inter-validation with the same specifications listed above, and also with a NIST calibration transferability certificate.
-Data acquisition computer
Attachment 1 Specifications pre-configured/ pre-loaded, specific to instrument Radiometric software (data acquisition, processing and display, and instrument control) Instrument continuous monitoring system (sensors and software)
-Ruggedized and Temperature controlled environmental enclosure.
Rain sensor controlled automatic hatch to protect the foreoptics from precipitation.
-Cables and power supplies (all necessary)
-Rugged transport cases
CLIN 0002:
On-site installation and training. On-site training including:
Instrument installation;
Software training;
Maintenance (basic and first line diagnostic training).
Warranty: 1 year except on source elements, optical windows and desiccant.
REQUIREMENTS:
Weight must allow the instrument to be manually deployed at remote fieldsites without the use of special equipment (forklifts, cranes, etc.). The maximum lifting weight required should be less than 75 kg.