Atch 1 - NMR_Specifications.pdf
PDF 419 KB Posted
- Solicitation number
- FA9101-22-Q-B005
View the file
Other files
| File | Type | Posted |
|---|---|---|
| FA9101-22-Q-B005.pdf | ||
| FA9101-22-Q-B005 - Brand Name JA-Final-Redacted-Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Justification for Nuclear Magnetic Resonance (NMR) Hydrogen Analyzer
The existing NMR Hydrogen Analyzer is used to determine the hydrogen concentration of hydrocarbon fuels for heating value calculations. The current instrument is over 20 years old, no longer supported by the vendor, and is operating on a Windows XP computer. Failure of the computer will make the instrument inoperable, as the software will not run on a higher version of Windows.
Specifications for NMR:
Compliant with ASTM Method D7171
Minimum of 20 MHz, 0.47 Tesla magnet.
Benchtop instrument.
Capable of calibration with pure standards or samples and reference values supplied by the user.
Solvent free determination of hydrogen in aviation fuels.
Includes sample preparation aluminum block for sample conditioning.
Includes glass sample tubes and PTFE plugs (or equivalent) to minimize evaporation.
Include instrument software, core license, computer with Windows 10 operating system, monitor, and accessories – keyboard, mouse, printer, and Ethernet cable.
Software capable of exporting to EXCEL or ASCII file.
Other Requirements to include in RFQ:
• Include instrument installation, set-up and training.
File details come from the government source that posted it. Updated .