PURCHASE SPECIFICATION
Pulse Generator Purchase Specification
3 August 2021
Prepared by
309th MMXG/MXDEB
HILL AFB, UTAH 84056
1.0 INTRODUCTION
The work described herein is to procure two (2) commercial off the shelf Dual Channel Pulse Generator, HP 8160A Compatible ECCN 3A992.a Part Number GP1552AH-M3. These specs shall be used to procure, assemble, test, install, start up, qualify and document a Pulse Generator per the requirements that are contained in this document. The Pulse Generator will be used to align, calibrate, and validate test stands. The Pulse Generator is a critical component of the United
States Air Force (USAF), Minuteman III Intercontinental Ballistic Missile (ICBM) Reentry
Vehicle support equipment. The Pulse Generators currently being used are failing to calibrate and are failing to perform as required. Two new Pulse Generators will be procured and used in
Bay K of Building 100 at Hill Air Force Base, Clearfield, Utah.
2.0 REFERENCE DOCUMENTS
IEEE-488.2 IEEE Standard Codes, Formats, Protocols, and Common Commands for use with
IEEE Std.
3.0 PERFORMANCE REQUIREMENTS
The High Level Objectives for this project include replacing the current Pulse Generators with two new Commercial off the shelf Pulse Generators.
3.1 Functional
3.1.1 The Pulse Generator design shall be 100% form, fit, and function compatibility with
Hewlett Packard 8160/8160A.
3.1.2 Dimensions shall not exceed 3.5" H x 8.4" W x 18" L (8.9 cm x 21.3 cm x 45.7 cm) without rack mount attachments.
3.1.3 Pulse Functions
3.1.3.1 Single: One Pulse at each selected period up to 50 MHz rate.
3.1.3.2 Double: One pair of pulses at each period up to 25 MHz repetition rate. Both pulses have the same selected width; the position of the second pulse shall be able to be set by a delay control.
3.1.4 Operating modes
3.1.4.1 Continuous mode shall provide continuous output at a programmed period rate
3.1.4.2 Triggered mode shall output quiescent until triggered by an internal, external, GPIB, or manual trigger, then generate one cycle at the programmed periodic rate.
3.1.4.3 Gated Mode shall perform as triggered mode with the exception pulses are active for the duration of the gated signal; once the gate deactivates the last waveform’s period continues to completion.
3.1.4.4 Burst mode shall perform as triggered mode. Output is active for a programmed number of cycles from 2 to 999,999. Set by the N Burst function.
3.1.4.5 External Width mode shall trigger duration and rate sets pulse width and repetition.
3.1.5 Timing Characteristics
3.1.5.1 Where Period is common to both channels
3.1.5.1.1 Range of a single pulse shall be 20 ns to 10 s (50 MHz – 0.1Hz repetition rate).
3.1.5.1.2 Range of a double pulse shall be 40 ns to 10 s, paired pulse per period mode (25 Mhz
– 0.1 Hz repetition rate).
3.1.5.1.3 Resolution shall be up to 6 digits, limited to 0.1 ns.
3.1.5.1.4 Accuracy shall be max ±1% of setting, ±1 ns, measured at 50% point of rising edge of sync pulse.
3.1.5.1.5 Jitter shall be ≤ 0.1% of setting + 50 ps, decreasing to 0.01% on slowest range
3.1.5.2 Where Width is independent for each channel
3.1.5.2.1 Range of a single pulse shall be 10 ns to 9.99999 s limited by 8 ns off time.
3.1.5.2.2 Resolution up to 6 digits limited to 0.1 ns.
3.1.5.2.3 Accuracy shall be max of ±2% of setting, ±2 ns measured at 50% point of leading and trailing edges, set to fastest transition time (±2 V).
3.1.5.2.4 Jitter shall be ≤ 0.1 of setting + 50 ps decreasing to 0.005% on slowest range.
3.1.5.2.5 External width shall be able to be determined by the external trigger level and the slope setting.
3.1.5.3 Where Delay is independent for each channel
3.1.5.3.1 Range 0 ns to 9.99999 s, limited by 8 ns off time.
3.1.5.3.2 Resolution up to 6 digits limited to 0.1 ns.
3.1.5.3.3 Accuracy shall be max of ±2% of setting, ±2 ns measured at 50% point of leading edge of sync to 50% point of leading edge of output set to fastest transition time (±2 V).
3.1.5.3.4 Jitter shall be ≤ 0.1 of setting + 50 ps decreasing to 0.005% on slowest range.
3.1.5.4 Where Duty Cycle is independent for each channel
3.1.5.4.1 Range 1 to 99% duty cycle may be programmed with duty function key or through
GPIB function key or through GPIB in single pulse per period mode.
3.1.5.4.2 Resolution 3 digits (0.1%).
3.1.5.4.3 Accuracy will be limited by pulse and width accuracy.
3.1.6 Output Characteristics
3.1.6.1 Amplitude shall be independent for each channel
3.1.6.2 High Level Range shall be between -19.90 V to +20 V. While on a 1kΩ source impedance, shall be able to source above -10 V and below -10 V.
3.1.6.3 Low Level Range shall provide between -10 V to +9.90 V to a 50 Ω load.
3.1.6.4 Amplitude Range shall be 0.1 v to 10 Vpp into 50 Ω load (20 Vpp maximum into an open circuit).
3.1.6.5 Resolution shall be 3 digits limited to 10 mV.
3.1.6.6 Accuracy ±1% of level setting, ±2% of pk-pk amplitude, ±50 mVinto 50 Ω load.
3.1.6.7 Aberrations shall be ≤ 5% + 50 mV into 50 Ω load for pulse levels between ± 5 V.
3.1.6.8 Output Impedance shall be 50 Ω ± 2%, 1 kΩ (high level range).
3.1.6.9 Mode shall have Normal / Compliment available.
3.1.7 Transition Times (Independent for Each channel)
3.1.7.1 Range shall be ≤ 6 ns to 25 ms variable, measured at +5 V to ±2 V and -5 V to -2 V.
Leading and trailing edges settable separately and limited to a 20:1 ratio between settings within the following ranges: 6 ns - 100 ns, 50 ns - 1 µs 500 ns - 10 µs, 5 µs - 100 µs 50 ns - 1.0 µs, 500 µs - 10 ms, 10.1 ms - 25 ms.
3.1.7.2 Resolution shall be 3 digits limited to 0.1 ns.
3.1.7.3 Accuracy shall be ±5% of setting ± 2 ns.
3.1.7.4 Linearity < 5% deviation from a straight line between 10% and 90% points.
3.1.8 Internal Trigger
3.1.8.1 Repetition Range shall be 100 ns to 100 s.
3.1.8.2 Resolution shall be 4 digits limited to 100 ns.
3.1.8.3 Accuracy shall be 0.1% ± 1 ns.
3.1.8.4 Jitter shall be ≤ 0.1% of setting + 50 ps.
3.1.9 Inputs and outputs
3.1.9.1 Trigger input sensitivity shall be 150 mVpp (minimum); DC to 50 MHz.
3.1.9.2 Minimum width shall be 10 ns; 250 mV required for triggering.
3.1.9.3 Maximum rate shall be 50 MHz.
3.1.9.4 Input Impedance shall be 1 MΩ ± 5%.
3.1.9.5 Input Protection shall be ±15 V DC plus peak AC.
3.1.9.6 Range shall be selectable from 9.99 V.
3.1.9.7 Resolution 3 digits limited to 10 mV.
3.1.9.8 Accuracy shall be ±5% of setting ± 25mV.
3.1.9.9 Slope selection shall be Positive or Negative; for triggering and for gating, respectively.
3.1.9.10 Counted burst shall be from 2 – 999,999 cycles per burst.
3.1.9.11 Sync output shall be a TTL level pulse at the programmed period rate of the generator’s output. A 50Ω output impedance, shall be used against short circuit and up to ±15 V accidental input. The high level is ≥2 V into 50 Ω. 3.5 ns typical transition.
3.1.10 General requirements
3.1.10.1 Power requirements shall be 90 V – 256 V; 48 -66, 70 VA max
3.1.10.2 Memory shall be non-volitile and store up to 99 complete panel settings. Las user setup is retained at power down.
3.1.10.3 Weight shall not exceed 10 lbs for non-rack mounted configuration.
3.1.10.4 Dimension requirements are given in section 3.1.2.
3.1.10.5 Operating temperature shall be able to fully operate in the temperature range 32 °F –
122 °F (0 °C to +50 °C).
3.1.10.6 Humidity shall be able to fully operate within the following ranges:
95% RH, 0 °C to +30 °C, 75% RH to +40 °C, 45% RH to +50 °C
3.1.10.7 Pulse generator shall be CE labeled.
3.1.10.8 Calibration interval shall be between 6 and 18 months.
3.1.10.9 Warm up time shall not have a required warm up time that exceeds 25 minutes.
3.1.11 Interface requirements
3.1.12 Interface shall offer a GPIB interface.
3.1.13 Internal shall be IEEE-488.2
3.1.14 Address shall be front panel selectable and range from 0-30.
3.1.15 Documentation requirement
3.1.15.1 End user manual shall be made available in both digital and print options.
3.1.15.2 User Calibration operations shall be provided if needed.
3.1.15.3 PMEL calibration cycle recommendation shall be provided if needed.
Ship To
Caralee Pritchett
US Air Force
7278 4th Street
Bldg 100
582nd MMXS Hill AFB, UT, 84056
USA