About this file
This synopsis announces a federal contract opportunity for the procurement of gas analyzer instruments. The National Oceanic and Atmospheric Administration seeks fourteen analyzer instruments, including twelve Picarro G2401 cavity ring-down spectrometers for measuring carbon dioxide, methane, carbon monoxide and water vapor at surface and tall-tower sites. One Picarro G2401-m aircraft flight-ready spectrometer is required, along with a custom modification of this instrument for high-altitude sampling. Response deadline is August 5, 2022, with evaluation and intended award by mid-August and quarterly deliveries beginning 20 weeks after order. Pricing will be determined at award. The procurement is designated as a total small business set-aside.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
UNITED STATES DEPARTMENT OF COMMERCE
National Oceanic and Atmospheric Administration
ACQUISITION AND GRANTS OFFICE
SYNOPSIS
Analyzer Instruments
SYNOPSIS:
The Analyzer Instruments requirement is hereby synopsized in accordance with FAR 5.207 Preparation and Transmittal of Synopses:
(1) Action Code: N/A
(2) Date: 22 July
(3) Year: 2022
(4) Contracting Office ZIP Code: 80305
(5) Product or Service Code: 6655
(6) Contracting Office Address: 325 Broadway, DSRC, Boulder, CO 80305
(7) Subject: Analyzer Instruments
(8) Proposed Solicitation Number: 1305M322QNRMJ0074
(9) Closing Response Date: 5 August 2022
(10) Contact Point or Contracting Officer:
Contract Specialist: Jade Moreira
Email: jade.moreira@noaa.gov
Contracting Officer: Darrin Moore
Email: darrin.moore@noaa.gov
(11) Contract Award and Solicitation Number:
Solicitation Number: (proposed)
Contract Number: To be determined at time of award.
(12) Contract Award Dollar Amount: To be determined at time of award.
(13) Contract Line Item Number: To be determined at time of award.
(14) Contract Award Date: To be determined at time of award.
(15) Contractor: To be determined at time of award.
(16) Description:
The estimated duration of the Analyzer Instruments contract is one year, with estimated starting date of 15 August 2022 to 14 August 2023, and estimated acquisition of 14 (fourteen) analyzer instruments. The estimated award date is 15 August 2022. For consideration to this award a completion of the solicitation is required. One or more of the items under this acquisition is subject to the World Trade Organization Government Procurement Agreement and Free Trade Agreements.
Analyzer Instruments specifications and evaluations are as follow:
GENERAL TECHNICAL SPECIFICATIONS:
In this procurement the brand name or equal to Picarro instruments should trace gas analyzers for near-simultaneous measurement of CO2, CH4, CO, and H2O. Instruments to be procured must maintain or improve upon the precision of current aircraft and field instrumentation used within the NOAA Global Greenhouse Gas Reference Network:
Picarro G2401: a cavity ring-down spectrometer for measurements of carbon monoxide, carbon dioxide, methane and water vapor. This instrument is to be used to measure ambient air at surface and tall-tower sites.
Picarro G2401-m: an aircraft flight-ready, cavity ring-down spectrometer for measurements of carbon dioxide, methane, carbon monoxide and water vapor. This instrument is to be used to measure ambient air collected with pressurized and unpressurized aircraft. The NOAA GML Carbon Cycle aircraft program plans to expand its airborne trace-gas measurements to include data collected from commercial aircraft.
Custom modification of a G2401-m to have a low-cavity pressure (80-torr) mode of operation with limited software changes for high-altitude, low-pressure measurement of ambient air. The modified analyzer with custom mode is expected to be flight ready, have a 3.5x or better degradation in precision, and require no changes for sample handling.
DETAILED TECHNICAL SPECIFICATIONS:
The G2401 analyzer must be capable of measuring:
● CO2 in bulk air samples to or better than ± 0.05 parts per million (ppm) in 400 ppm
● CH4 in bulk air samples to or better than ±0.5 nmol/mol (ppb) in 1800 ppb
● CO in bulk air samples to or better than ±4 nmol/mol (ppb) in 100 ppb as 1-σ over a 30-second averaging period.
The analyzer must have a measurement range of at least:
● 0 – 1000 ppm CO2
● 0 – 6 ppm CH4
● 0 – 5 ppm CO
The analyzer must be extremely stable and not subject to rapid drift in response to environmental influences such as pressure and temperature changes. The cell temperature and pressure must be controlled to within ±0.01°C and ±0.15 Torr to prevent measurement drift associated with temperature and pressure fluctuations. The analyzer must be able to measure ambient air with a wide range of conditions, from -10 to +45 °C in temperature and 400 – 1330 hPa in pressure. The analyzer should minimize sample gas consumption by having a small sample cell volume and efficient flushing of internal components such as tubing, valves, etc. The analyzer needs to be deployable to the field, as well as in the laboratory.
The G2401-m analyzer must be capable of measuring:
● CO2 in bulk air samples to or better than ± 0.05 parts per million (ppm) in 400 ppm
● CH4 in bulk air samples to or better than ±0.5 nmol/mol (ppb) in 1800 ppb
● CO in bulk air samples to or better than ±4 nmol/mol (ppb) in 100 ppb as 1-σ over a 30-second averaging period.
The analyzer must have a measurement range of at least:
● 0 – 1000 ppm CO2
● 0 – 6 ppm CH4
● 0 – 5 ppm CO
The analyzer must be extremely stable and meet thresholds for drift of ≤ 10 ppb for CO; ≤ 0.2 ppb for CO2; and 2 ppb for CH4 at standard temperature and pressure (STP) over 24 hours. The cell temperature and pressure must be controlled to within ±0.01°C and ±0.15 torr to prevent measurement drift associated with temperature and pressure fluctuations. The analyzer must be able to measure ambient air with a wide range of conditions, from -10 to +45 °C in temperature and 400 – 1330 hPa in pressure. The analyzer should minimize sample gas consumption by having a small sample cell volume and efficient flushing of internal components such as tubing, valves, etc. The analyzer needs to be deployable to pressurized and unpressurized aircraft, as well as in the laboratory and in the field.
Custom modification of a G2401-m (delivery required by February 2023)
● 80-torr cavity pressure mode of operation
● limited software changes
● degradation in precision of 2x or better
● no sample handling changes
All analyzers should be robust, be able to run unattended for months, and require little maintenance.
DELIVERABLE SCHEDULE:
● 4-6 units delivered quarterly, starting with the first set within 20 weeks after order is placed.
● The single G2401-m with custom 80-torr mode is required by February 2023.
(17) Place of Contract Performance: Boulder, CO
(18) Set-aside Status: The Government intends this procurement to be a total small business set-aside award.