Summary of Modifications/Changes in this Update

This Summary of Changes is for information only.

It is not a part of the referenced document, and should not be used for project documentation.

U.S. Department of Veterans Affairs ♦ Office of Construction & Facilities Management

DATE OF THIS VERSION:

January 1, 2017

TITLE OF DOCUMENT:

Ceiling Mounted Patient Lift System, 11 73 00

DATE OF VERSION BEING SUPERSEDED (old):

September 1, 2014

DESCRIPTION OF DOCUMENT (previous title, number, other identifying data):

Ceiling Mounted Patient Lift System, 11 73 00

SUMMARY OF CHANGES IN THIS VERSION:

1. Section 1.3. Quality Assurance: C. Certification of compliance with VA requirements shall be provided by an independent third party, Inspector of Record (IOR).

2. Section 1.4. Submittals: A Shop Drawings, Product Data, and Samples:

a. Shop Drawings shall show structural supports for ceiling mounted lift and provide structural calculations for attachments to structure.

b. A set of stamped and signed drawings shall be provided by a registered structural engineer and shall meet all code requirements for the jurisdiction having authority.

c. Shop drawings shall show obstructions and coordinate their relocation.

d. Manufacturer shall provide BIM for clash detection at the request of the VA or General Contractor.

3. Section 1.5 Applicable Publications:

a. D. International Electromagnetic Commission: has been changed to 60601-1-2 (2015) as of April 1st 2017.

b. E. VA Patient Safety Alert AL14-07: a editable PDF checklist is now provided on VHA’s Patient Safety Website.

4. Section 2.1 Ceiling Track System:

a. As per ISO 10535 standards, installed rail shall be tested for 1.5 times greater than the motor’s weight capacity.

5. Section 2.2 Lifting Unit:

a. B. Lifting Capacity shall be 550 lbs for non-bariatric lifts, and 1,000 lbs for bariatric lifts.

6. Section 2.4 Batteries:

a. B. for non-bariatric lifts a minimum of 40 transfers. For Bariatric lifts a minimum of 50 transfers.

7. Section 2.6 Straps and Sling:

a. Straps and Slings shall meet ISO 10535, Appendix A Guidelines.

8. Section 3.4 Inspection:

a. Periodic Inspection shall be provided by manufacturer on a yearly basis in compliance with ISO 10535.