36C25019Q1306-002.pdf
PDF 2 MB Posted
- Attached to
- ROTARY CLAW MEDICAL VACUUM PUMP
- Solicitation number
- 36C25019Q1306
About this file
36C25019Q1306 Powerex Medical Vacuum Specifications.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| -17983.docx | DOCX document | |
| 36C25019Q1306-000.docx | DOCX document | |
| 36C25019Q1306-003.pdf | ||
| 36C25019Q1306-001.docx | DOCX document | |
| 36C25019Q1306-002.pdf | ||
| 36C25019Q1306-003.pdf | ||
| 36C25019Q1306-001.docx | DOCX document | |
| 36C25019Q1306-000.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Powerex • 150 Production Drive • Harrison, OH 45030 • Phone: 888-769-7979 •Fax: 513-367-3125
BASE MODEL NUMBER: CVPT0504B
SYSTEM CAPACITY: 76 SCFM @ 19” Hg HORSEPOWER: 15 HP Total (5 HP x 3) RECEIVER SIZE/CONFIGURATION: 120 Gal./Vertical ELECTRICAL: 460V, 3PH, 60Hz.
(MUST SPECIFY VOLTAGE WHEN ORDERING.)
*System performance shown with one pump module in reserve per NFPA 99
GENERAL DESCRIPTION
The Powerex medical vacuum system is designed to create a suction system to remove unwanted fluids or gases from hospital working areas. The medical vacuum system package is compliant with the NFPA 99 requirements for medical vacuum and is operationally tested prior to shipment. The package shall include multiple oil-free rotary claw vacuum pumps and associated equipment, one ASME tank and one medical control panel. All components shall be completely pre-piped and pre-wired to single-point service connections. System shall be provided with intake and discharge flexible connectors. All interconnecting piping and wiring shall be completed and operationally tested prior to shipment. Vibration isolation pads are included with the system.
DETAILED PRODUCT DESCRIPTION:
OIL-LESS ROTARY CLAW VACUUM PUMP
The medical vacuum pumps shall be of the air-cooled oil-less rotary claw design. Each claw vacuum pump is equipped with a set of opposing rotors with frictionless operation, eliminating the need to service or replace wearing parts. Each oil-less rotary claw vacuum pump shall be equipped with:
TEFC, standard efficiency electric motor with coupling Inlet filter Vibration isolation pads Check valve
VACUUM RECEIVER
The system shall include an ASME rated air receiver rated for 200 PSIG and full vacuum. The tank shall be equipped with a vacuum gauge, a sight glass, by-pass valves, and a manual drain.
PREMIUM NFPA CONTROL PANEL
The control system provides automatic lead/lag activation based on demand and automatic alternation of all pumps in order to equalize the amount of usage among the available pumps. The control panel shall include:
A color touch screen panel which displays the operating status of the unit. The touch screen will display pump status, pump faults, pump run hours, system pressure, system alarms and service alert notifications for the pumps.
Building automation communication gateway, through a gateway server card with BacNet® protocol and Web server features. Web server features include email notifications in case the system is in alarm or has achieved one of its maintenance intervals and requires service.
Ethernet port for connection to BacNet® server or direct connection to facility Ethernet for viewing of system operations and status via device IP-address.
A PLC controller that includes logic programmed to appropriately manage start-stop, lead-lag and other pump operations.
UL508A listed control panel in a NEMA 12 enclosure.
rscully Highlight
Powerex • 150 Production Drive • Harrison, OH 45030 • Phone: 888-769-7979 •Fax: 513-367-3125
VARIABLE FREQUENCY DRIVE (VFD):
Variable Frequency Drive (VFD) control improves efficiency over a conventional “on/off” demand based system by more closely matching the pump speed to the changing load requirements. The control system calculates the frequency to speed up or slow down the pump based on changes to the demand verses the set-point. Pump speeds matched to demand will result in lower amp draws, prolong motor life, prolong bearing life, and a reduction in energy cost. All VFD systems come standard with a HMI control panel; our Powerex Building Management Integrator (PBMI) is an available upgrade. This option is not available with Basic Controls.
SYSTEM CONNECTIONS
The system is supplied with a remote intake connection and flexible connectors for both intake and discharge. All piping complies with current NFPA 99 requirements.
Claw Vacuum Package 4-7.5HP with Premium Controls
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
(T) 1.888.769.7979 • (F) 513.367.3125 • www.powerexinc.com
01/2017
Dimensions Model Dim. A Dim. B Dim. C Inlet Outlet (x3)
CVPT0504A 66" 57” 87” 2" 1"
CVPT0504B 66" 57” 87” 2" 1"
CVPT0755A 66" 57” 87” 2" 1"
CVPT0755B 66" 57” 87” 2" 1"
Claw Vacuum Package with Premium Controls
Model HP SCFM @ 19” Hg each
NFPA
System
Capacity(1)
Tank Size (gal) BTU/Hr (2) dB(A)
Level (3)
System F.L.A. System Weight (lbs)208V 230V 460V
CVPT0504A 4 (3) 29 58 120V 17,340 82 32 29.6 15.8 2,650
CVPT0504B 5 (3) 38 76 120V 21,676 82 39.2 36.2 19.1 2,800
CVPT0755A 6.4 (3) 52 104 200V 27,744 82 50 49.4 25.7 3,550
CVPT0755B 7.5 (3) 65 130 200V 32,512 82 68 55.4 28.7 3,750
Notes:
1 – System Capacity is shown with one or more pumps in reserve per NFPA 99.
2 – BTU/Hr levels are shown with reserve pump(s) on standby.
3 – dB(A) levels are shown with one pump in reserve per NFPA 99.
4 – 3 Year Limited Warranty.
rscully Highlight rscully Highlight rscully Highlight rscully Highlight rscully Highlight rscully Highlight
Line
U
AI
I
S1
AI
Analogue input signal selector (U/I)
1 2
ON
In0 1 2 3 4 5 6 71 12131415
OutPw r
R un
Er r
St at 0 71 2 3 54 6 1110
+24 0v DC 0 1 2 3 54 6 7 8 9 10 11 12 13 Com
NL G 0 10 2 13 4 5 6 37 2 10 11
Com Com Com Com
Line
Medical Vacuum Systems
Model SCFM @ 19" Hg HP Tank
(Gal.)
Duplex Enclosure
MCED050 33 5 120
MCED070 52.8 7.5 small 120
MCED075 67.8 7.5 big 120
Triplex Enclosure
MCET050 66 5 200
MCET070 105.6 7.5 small 200
MCET075 135.6 7.5 big 200
Quadplex Enclosure
MCEQ050 99 5 200
MCEQ070 158.4 7.5 small 200
MCEQ075 203.4 7.5 big 200
• Most energy efficient technology
• Oil-free compression chamber
• Extremely low maintenance
• No water and sewage costs
• Suitable for dedicated waste anesthesia (WAGD) systems
• Air-cooled design
BENEFITS
• NEW Enclosed Systems
• Premium Warranty
• Expandable Systems
• Variable Frequency Drive
• Oxygen Assured Pumps
• Internal Tank Lining
• Non-HMI Controls
• Alternate Tank Sizes &
Custom Configurations
Model SCFM @ 19" Hg HP Tank
(Gal.)
Duplex Modular Skidmount
CVPD0504A° 29 4 120
CVPD0504B° 38 5 120
CVPD0754A° 52 6.4 120
CVPD0754B° 65 7.5 120
CVPD1005° 87 10 200
CVPD1505° 129 15 200
Triplex Modular Skidmount
CVPT0504A° 58 4 120
CVPT0504B° 76 5 120
CVPT0755A° 104 6.4 200
CVPT0755B° 130 7.5 200
CVPT1005° 174 10 200
CVPT1505° 258 15 200
Quadplex Modular Skidmount
CVPQ0505A° 87 4 200
CVPQ0505B° 114 5 200
CVPQ0755A° 156 6.4 200
CVPQ0755B° 195 7.5 200
CVPQ1005° 261 10 200
CVPQ1505° 387 15 200
Pentaplex Modular Skidmount
CVPP1505 516 15 200
DRY CLAW
Inside the pump housing, two claw shaped rotors take in air as they rotate in opposite directions. The air is compressed by the rotors, then discharged through a silencer to atmosphere. The pumping chamber is dry. There is no contact between the rotors or the cylinder wall eliminating internal wear and parts to replace. The rotors are synchronized by gears requiring a small amount of oil in the gear housing. Gear oil change requirements are minimal at approximately 20,000 hours.
AVAILABLE OPTIONS
SYSTEMS SPECIFICATIONS
CVPT0755 with VFD
Contact factory for alternate tank sizes and expandable systems °System is OSHPD (seismic) certified; Certification No. OSP-3093-10
Model SCFM @ 19" Hg HP Tank
(Gal.)
Duplex Horizontal Tankmount
CVTD0203 16 2 80
CVTD0303 21 3 80
CVTD0504A 29 4 120
CVTD0504B 38 5 120
Duplex Vertical Tankmount
CVTD0203V 16 2 80
CVTD0303V 21 3 80
CVTD0504AV 29 4 120
CVTD0504BV 38 5 120
Duplex Tank-over Skidmount
CVPDT0502A 29 4 60
CVPDT0502B 38 5 60
CVPDT0752A 52 6.4 60
CVPDT0752B 65 7.5 60
rscully Highlight rscully Highlight rscully Highlight rscully Highlight rscully Highlight rscully Highlight rscully Cross-Out rscully Cross-Out
Cross-Out
MEDICAL VACUUM SYSTEMS OPTIONS
STANDARD PREMIUM CONTROL PANEL
• PLC controller and 6" color HMI (Human Machine Interface) touch screen displays the operating status of the unit
• UL508A listed and labeled, NEMA 4/12 enclosure
• Visual and audible alarms for:
• System overload trip
• High temperature conditions
• Service intervals
• Panel door includes:
• Visual and audible alarms
• HOA switch for each pump
• Building automation gateway with BacNet® protocol
• Email notifications for alarms and service alerts.
OPTIONAL NON-HMI CONTROL PANEL
• UL508A listed and labeled, NEMA 4/12 enclosure.
• Timed lead/lag alternation.
• Visual and audible reserve in use alarm.
• Redundant control circuit transformers.
• Dry contacts for remote alarm monitoring.
• Each vacuum pump provided with:
• Hand/Off/Auto selector switch.
• Magnetic starter with 3 leg overload protection.
• Visual and audible high temperature shut down.
• Hour meter.
• Fully compliant with the latest edition of
NFPA 99
• Multiple vacuum pumps
• ASME receiver
• UL Listed control panel (HMI or Non-HMI)
• All interconnecting piping and wiring
• Vibration isolation pads and flex connectors
• Manufactured in ISO 9001:2008 UL certified facilities
SYSTEM DESIGN FEATURES CONFIGURATION TYPES
Horizontal TankmountVertical TankmountModular SkidmountTank-over Skidmount rscully Highlight rscully Highlight rscully Cross-Out
Cross-Out
SR4325
03/2018
The leader in oil-free air technology
Powerex manufactures industry-leading comprehensive medical air and vacuum systems, from source equipment to pipeline
For more information contact your local Powerex distributor or Powerex sales representative today
150 Production Drive • Harrison, OH 45030 • USA T 1.888.769.7979 • F 513-367-3125 • www.powerexinc.com
| Submittal Cover Sheet |
| Spec |
| CVPT5-7.5 PREMIUM |
| Slide Number 1 |
| CVPT0504BB-VFD_D1 |
| PXMI-V369V F1 VFD MED VAC TRI |
| Medical Vacuum - pages (SR4325) 2018 |
File details come from the government source that posted it.