Attachment B
RFQ: 36C24818Q9798
Line Item Description Qty Unit
Unit Price
Total Price
Patient Heater-Cooler System for Cardiopulmonary procedures
To include all required connection cables. EA 2
Total Cost
Salient Characteristics:
• Provides 3 circuits for fast and efficient heating and cooling
• Compressor-based cooling and 3 independent water tanks that eliminate the need for ice and water supply in the operating room and provide separate settings for warm cardioplegia, cold cardioplegia and patient temperature.
• Temperature settings must be able to be adjusted to the tenth decimal point for accurate temperature control
• A water evacuation feature must allow pulling back water from the lines into the tanks at the end of a procedure
• Operating Conditions: +10˚C/50˚F through +35˚C/95˚F
• Storage Temperature: 0˚C/32˚F through +40˚C/104˚F
• Minimum level patient circuits 4.5 l (11.6 l total volume)
• Maximum level patient circuits 6.5 l (13.8 l total volume)
• Water flow patient circuits 15 - 17 l
• Water flow cardioplegia circuit 9 - 11 l
• Water pressure in the circuits 0.7 bar max./-0.3 bar min.