Attachment B

RFQ: 36C24818Q9798

Line Item Description Qty Unit

Unit Price

Total Price

Patient Heater-Cooler System for Cardiopulmonary procedures

To include all required connection cables. EA 2

Total Cost

Salient Characteristics:

• Provides 3 circuits for fast and efficient heating and cooling

• Compressor-based cooling and 3 independent water tanks that eliminate the need for ice and water supply in the operating room and provide separate settings for warm cardioplegia, cold cardioplegia and patient temperature.

• Temperature settings must be able to be adjusted to the tenth decimal point for accurate temperature control

• A water evacuation feature must allow pulling back water from the lines into the tanks at the end of a procedure

• Operating Conditions: +10˚C/50˚F through +35˚C/95˚F

• Storage Temperature: 0˚C/32˚F through +40˚C/104˚F

• Minimum level patient circuits 4.5 l (11.6 l total volume)

• Maximum level patient circuits 6.5 l (13.8 l total volume)

• Water flow patient circuits 15 - 17 l

• Water flow cardioplegia circuit 9 - 11 l

• Water pressure in the circuits 0.7 bar max./-0.3 bar min.