36C10G19Q0075-004.docx
DOCX document 31 KB Posted
- Attached to
- NX EQ Pulse Oximeters (VA-19-00032743)
- Solicitation number
- 36C10G19Q0075
About this file
36C10G19Q0075 Attachment D Minimum Salient Characteristics- Pulse Oximeters.docx
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C10G19Q0075-0001001.xlsx | XLSX spreadsheet | |
| 36C10G19Q0075-0001000.docx | DOCX document | |
| 36C10G19Q0075-0001002.docx | DOCX document | |
| 36C10G19Q0075-005.docx | DOCX document | |
| 36C10G19Q0075-006.xlsx | XLSX spreadsheet | |
| 36C10G19Q0075-002.docx | DOCX document | |
| 36C10G19Q0075-001.xlsx | XLSX spreadsheet | |
| 36C10G19Q0075-003.docx | DOCX document | |
| 36C10G19Q0075-000.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
NONIN Brand or Equal
This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires that the vendor indicate each product that is being offered as an “equal” product. For each equal product, the vendor must include a description reflecting the characteristics and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified below. The vendor must clearly identify the item by brand name (if any) and make/model number. The vendor must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer.
VENDORS SUBMITTING AN "OR EQUAL" QUOTE MUST MEET ALL THE MINIMUM SALIENT CHARACTERISTICS LISTED BELOW.
Minimum Salient Characteristics (SC) for Literature Review
| CLIN |
| MFG |
| Part Number |
| Description |
NONIN
EM-9509-3150
WristOx2, Mode 3150, Wrist worn Pulse Oximeter
| SC |
| SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
SC 1
Must be small, lightweight and wearable
Literature Review
SC 2
Must be able to transfer data securely, via Bluetooth wireless or USB
SC3
Must have minimum of 12 hours battery life
SC 4
Must have an accurate response for capturing events in challenging conditions such as motion and low perfusion
SC 5
Must have minimum (1) one-year warranty
SC 6
Must have Oxygen Saturation display range of 0% to 100% SpO2
SC 7
Must have minimum pulse rate range of 25 to 240 pulses per minute
SC 8
Must have a three-digit numeric display and a pulse strength bar graph
| CLIN |
| MFG |
| Part Number |
| Description |
0002-0005
EM-9509-9590
EM-9509-959B
EM-9509-959R
EM-9509-959P
Onyx Vantage 9590 Finger Pulse Oximeter Black Onyx Vantage 9590 Finger Pulse Oximeter Blue Onyx Vantage 9590 Finger Pulse Oximeter Red Onyx Vantage 9590 Finger Pulse Oximeter Purple
| SC |
| SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
Must be able to read SpO2 and pulse rate from any angle, in all lighting conditions, day or night
Must supply pulse quality indicator for quick assessment of pulse quality
Must automatically adjust to each patient from pediatric to adult
Must turn on and off automatically with insertion of digit
Must allow for minimum of 1800 spot checks or minimum of 30 hours of continuous operation, using two 2-AAA alkaline batteries
Must have a minimum 2-year warranty
Must have Oxygen Saturation display range of 0% to 100% SpO2
Must have a minimum pulse rate range of 30 to 240 pulses per minute
SC 9
Must have an oxygen saturation declared accuracy of 70-100%
| CLIN |
| MFG |
| Part Number |
| Description |
EM-9510-7500
7500 Series Tabletop Pulse Oximeter
| SC |
| SALIENT CHARACTERISTICS |
| METHOD OF EVALUATION |
Must be able to monitor neonate to adult patients in acute care to homecare environments
Must be flexible to be used as standalone device or interface with other equipment
Must have color coded Pulse Strength Bar Graph with audible tone with each tone
Must have a Pulse Quality Indicator
Must have a Sensor Fault Indicator that indicates when a sensor has become disconnected
Must have a min. of 70 hours memory
Must have at least a one-year warranty
Must have Oxygen Saturation display range of 1% to 100% SpO2
SC 9
Must have pulse rate range of 25 to 240 pulses per minute
SC 10
Must have an oxygen saturation declared accuracy of 70-100%