36C10G18Q0134-002.docx

DOCX document 24 KB Posted

Attached to
NX Dermatomes
Solicitation number
36C10G18Q0134
Issued by

About this file

36C10G18Q0134 36C10G-18-Q-0134 - Attachment B - Salient Characteristics.docx

View the file

Other files

File Type Posted
36C10G18Q0134-006.docx DOCX document
36C10G18Q0134-007.xls XLS spreadsheet
36C10G18Q0134-004.docx DOCX document
36C10G18Q0134-003.docx DOCX document
36C10G18Q0134-001.xlsx XLSX spreadsheet
36C10G18Q0134-005.docx DOCX document
36C10G18Q0134-000.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires that the offeror indicate each product that is being offered as an “equal” product. For each equal product, the offeror must include a description reflecting the characteristics and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The offeror also must identify the item by brand name (if any) and make/model number. Finally, the offeror must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer, and describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

SC Salient Characteristics (SC) Literature Review Skin Graft Mesher

SC 1Dermatomes ratios must be available in 1:1, 1.5:1, 2:1, 3:1, and 4:1
SC 2Dermatomes must contain a convenient hinge design to allow the cutters to be changed quickly or removed for cleaning
SC 3Dermatomes must have a built-in comb that minimizes the chances of grafts sticking and wrapping around the cutters
SC 4Dermatomes must have an Autoclave case that makes sterilization and transportation simple
SC 5Dermatomes must have Lube less gears help to ensure a smooth performance, time after time
CLIN
Manufacturer (Brand)
Product Number
Description
0002
Zimmer
00-7701-022-00
Skin Graft Mesher Ratchet Handle: A surgical skin grafting tool

Air Dermatome

SC 1Dermatome must contain a smaller diameter handle
SC 2Dermatome must be lightweight for easier maneuverability
SC 3Dermatome must have a power source with a hose that exhausts no less than 10’ (3 m) away from surgical site
SC 4Dermatome must have a consistent source of regulated medical grade air or nitrogen for optimal performance
SC 5Dermatome cutting range has to be no less than from 0.000” - 0.030” (0.00 mm - 0.76 mm)
SC 6Dermatome must have a safety lever to help prevent accidental activation
SC 7Dermatome instruments must have the ability to be sterilized by steam sterilization
SC 8Dermatome dimensions should be no less than 8.5 in. (21.6 cm) Length, 5.18 in. (13.16 Cm) width
CLIN
Manufacturer (Brand)
Product Number
Description
0013
Zimmer
00-8801-000-00
Zimmer Biomet Air Dermatome Complete Kit

Electric Dermatome

SC 1Dermatomes must be useable anywhere in the OR where there is an electrical outlet
SC 2Dermatomes must utilize a power source no less than 4.3 amp capability
SC 3Dermatomes cutting range has to be no less than from 0.000” - 0.030” (0.00 mm - 0.76 mm)
SC 4Dermatome must have a safety lever to help prevent accidental activation
SC 5Dermatome instruments must have the ability to be sterilized by steam sterilization
CLIN
Manufacturer (Brand)
Product Number
Description
0025
Zimmer
00-8821-000-00
Zimmer Biomet Electric Dermatome Complete Kit

MeshGraft II Tissue Expansion System

SC 1Dermatomes ratios must be available in 1.5:1, 3:1, 6:1, and 9:1
SC 2Dermatomes must consist of a ratchet handle, continuous feed roller, a cutting roller, and a guidance plateau.
CLIN
Manufacturer (Brand)
Product Number
Description
0034
Zimmer
00-2195-000-00
MeshGraft II Complete

File details come from the government source that posted it.