SALIENT CHARACTERISTICS

For

NON-EXPENDABLE EQUIPMENT:

LITHOTRIPTERS

Prepared By:

Strategic Acquisition Center Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400 Fredericksburg, VA 22408

1. SCOPE OF WORK:

The Veterans Health Administration (VHA) Non-Expendable Equipment National Program identified Lithotripters as a candidate for VHA-wide (otherwise referred to as “national”) single award Requirements contract.

Lithotripters are used in a medical procedure called lithotripsy, which involves the destruction of hardened masses like kidney stones or gallstones. The period of performance is for one (1) twelve (12) month base period with two (2) twelve (12) month option periods.

2. REQUIREMENT:

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires that the offeror indicate each product that is being offered as an “equal” product. For each equal product, the offeror must include a description reflecting the characteristics and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The offeror also must clearly identify the item by brand name (if any) and make/model number. Finally, the offeror must include descriptive literature, such as illustrations, drawings, or a clear reference to previously furnished descriptive data or information available to the contracting officer, and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

SC Salient Characteristics (SC) Literature Review

Lithotripters

CLIN
Brand
Part Number
Description
0001
Olympus
SPL-S
ShockPulse-SE Lithotripsy System - includes Generator, Transducer, Nose Cone, Power Cord,

Cleaning Stylet & Wrench

0002
Olympus
SPL-T
ShockPulse Lithotripsy Transducer
0003
Olympus
SPL-FS
ShockPulse Lithotripsy Footswitch
0004
Olympus
SPL-PDBX376
ShockPulse Lithotripsy Probe, Sterile Single Use, 3.76 mm (3/Box)
0005
Olympus
SPL-PDBX340
ShockPulse Lithotripsy Probe, Sterile Single Use, 3.40 mm (3/Box)
0006
Olympus
SPL-PDBX183
ShockPulse Lithotripsy Probe, Sterile Single Use, 1.83 mm (3/Box) - Includes Cleaning Stylet — Small
0007
Olympus
SPL-PDBX150
ShockPulse Lithotripsy Probe, Sterile Single Use, 1.50 mm (3/Box) - Includes Cleaning Stylet — Small
0008
Olympus
SPL-PDBX097
ShockPulse Lithotripsy Probe, Sterile Single Use, 0.97 mm (3/Box)

All SCs (1-8) apply to CLIN 1.

CLINS 2-8 are not necessarily tied to the SCs so they are included in a separate table at the end of this document with the Simplified explanation for each line item.

SC
APPLICABLE CLIN(s)
SALIENT CHARACTERISTICS
METHOD OF EVALUATION
SC 1
0001
System capable of producing ultrasonic wave energy and intermittent shockwave (ballistic/mechanical) energy.
Literature Review
SC 2
0001
System able to fragment stones with a broad range of hardness and size
Literature Review
SC 3
0001
System able to simultaneously suction through the probes to evacuate debris
Literature Review
SC 4
0001
System must be able to use both variable ultrasonic and mechanical energy to rapidly fragments stones of all types with minimal damage to surrounding tissue
Literature Review
SC 5
0001
System operable via single watertight (excluding plug) footswitch and hand controls
Literature Review
SC 6
0001
System able to utilize sterile single use disposable probes
Literature Review
SC 7
0001
Voltage Requirement: 100–240 VAC ±10%
Literature Review
SC 8
0001
Operating and Storage Environment:

Operating Environment Ambient Temperature 10 – 40°C Relative Humidity 30% – 70% Air Pressure 70 – 106 kPa Storage Environment Ambient Temperature -25 – +55°C Relative Humidity 10% – 85% Air Pressure 70 – 106 kPa Literature Review image1.png