Schedule of Line Items
| HTME VISN | Requisition Number in Action | Equipment Type | Station | Consolidation Period | Benchmark (BM) or Technical Specification (TS) |
| VISN 19: VA Rocky Mountain Network | 436B10015 | XR US | 436 MISSOULA, MT | MAY 2021 | TS |
| VISN 22G: VA Southwest Health Care Network | 501B10017 | XR US | 501 ALBUQUERQUE, NM | MAY 2021 | BM |
| VISN 22G: VA Southwest Health Care Network | 501B10018 | XR US | 501 ALBUQUERQUE, NM | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 506B10028 | XR US | 506 ANN ARBOR, MI | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 506B10049 | NM SPECT CT | 506 ANN ARBOR, MI | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 508B12006 | XR US | 508 DECATUR, GA | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 508B12008 | XR US | 508 DECATUR, GA | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 509B12003 | NM PET CT | 509 AUGUSTA, GA | MAY 2021 | TS |
| VISN 7: VA Southeast Health Care Network | 509B12005 | XR PORT C-ARM | 509 AUGUSTA, GA | MAY 2021 | TS |
| VISN 7: VA Southeast Health Care Network | 509B12008 | XR US | 509 AUGUSTA, GA | MAY 2021 | TS |
| VISN 10: VA Healthcare System of Ohio | 515B10017 | NM BONE DEN | 515 BATTLE CREEK, MI | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 515B10018 | XR US | 515 BATTLE CREEK, MI | MAY 2021 | BM |
| VISN 8: VA Sunshine Healthcare Network | 516B10009 | WORKSTATION | 516 ST PETERSBURG, F | MAY 2021 | BM |
| VISN 8: VA Sunshine Healthcare Network | 516B10010 | XR RAD | 516 ST PETERSBURG, F | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 516B10011 | RT LINAC | 516 ST PETERSBURG, F | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 516B10012 | XR PORT C-ARM | 516 ST PETERSBURG, F | MAY 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 519B10010 | NM SPECT CT | 519 BIG SPRING, TX | MAY 2021 | TS |
| VISN 16: South Central VA Health Care Network | 520B10050 | XR US | 520 BILOXI, MS | MAY 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 528B14580 | XR PORT C-ARM | 528 SYRACUSE, NY | MAY 2021 | TS |
| VISN 2: VA Healthcare Network Upstate New York | 528B14581 | XR PORT C-ARM | 528 SYRACUSE, NY | MAY 2021 | TS |
| VISN 2: VA Healthcare Network Upstate New York | 528B14582 | XR PORT C-ARM | 528 SYRACUSE, NY | MAY 2021 | TS |
| VISN 2: VA Healthcare Network Upstate New York | 528B14583 | XR PORT C-ARM | 528 SYRACUSE, NY | MAY 2021 | TS |
| VISN 4: VA Stars & Stripes Healthcare Network | 529B12012 | XR US | 529 BUTLER, PA | MAY 2021 | TS |
| VISN 4: VA Stars & Stripes Healthcare Network | 529B12013 | XR US | 529 BUTLER, PA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 531B10148 | WORKSTATION | 531 BOISE, ID | MAY 2021 | BM |
| VISN 20: VA Northwest Network | 531B10163 | XR US | 531 BOISE, ID | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 534B12025 | XR PORT C-ARM | 534 CHARLESTON, SC | MAY 2021 | TS |
| VISN 12: VA Great Lakes Health Care System | 537B18003 | XR US | 537 CHICAGO, IL | MAY 2021 | BM |
| VISN 12: VA Great Lakes Health Care System | 537B18004 | XR US | 537 CHICAGO, IL | MAY 2021 | TS |
| VISN 10: VA Healthcare System of Ohio | 539B11027 | XR US | 539 CINCINNATI, OH | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 541B13031 | XR US | 541 CLEVELAND, OH | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 541B13032 | XR US | 541 CLEVELAND, OH | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 541B93028 | XR US | 541 CLEVELAND, OH | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 544B12028 | NM SPECT | 544 COLUMBIA, SC | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 544B12033 | XR US | 544 COLUMBIA, SC | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 544B12049 | XR C-ARM HYBRID | 544 COLUMBIA, SC | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 546B10017 | XR CT | 546 MIAMI, FL | MAY 2021 | TS |
| VISN 10: VA Healthcare System of Ohio | 553B10041 | XR US | 553 DETROIT, MI | MAY 2021 | BM |
| VISN 19: VA Rocky Mountain Network | 554B10001 | XR PORT C-ARM | 554 AURORA, CO | MAY 2021 | TS |
| VISN 7: VA Southeast Health Care Network | 557B12007 | XR US | 557 DUBLIN, GA | MAY 2021 | BM |
| VISN 7: VA Southeast Health Care Network | 557B12008 | XR US | 557 DUBLIN, GA | MAY 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 561B10008 | XR US | 561 EAST ORANGE, NJ | MAY 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 561B10015 | XR RF | 561 EAST ORANGE, NJ | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 565B10026 | XR PORT RAD | 565 FAYETTEVILLE, NC | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 565B10031 | XR US | 565 FAYETTEVILLE, NC | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 570B10020 | XR PORT RAD | 570 FRESNO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 570B10021 | XR US | 570 FRESNO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 570B10024 | XR US | 570 FRESNO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 570B10026 | XR PORT C-ARM | 570 FRESNO, CA | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15035 | XR US | 573 JACKSONVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15036 | XR US | 573 JACKSONVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15037 | XR PORT C-ARM | 573 VILLAGES, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15038 | XR US | 573 GAINESVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15039 | XR US | 573 LAKE CITY, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15040 | XR US | 573 GAINESVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15041 | NM BONE DEN | 573 GAINESVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15042 | XR MRI | 573 GAINESVILLE, FL | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 573B15043 | XR C-ARM | 573 GAINESVILLE, FL | MAY 2021 | TS |
| VISN 16: South Central VA Health Care Network | 580B10031 | XR US | 580 HOUSTON, TX | MAY 2021 | BM |
| VISN 15: VA Heartland Network | 589B12019 | XR US | 589 LEAVENWORTH, KS | MAY 2021 | BM |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10002 | XR C-ARM | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10003 | XR US | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10004 | XR PORT C-ARM | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10005 | XR US | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10006 | XR US | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10007 | XR US | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10008 | XR RAD | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10009 | XR MAMMO | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 590B10010 | XR RAD | 590 HAMPTON, VA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 593B10026 | XR PORT C-ARM | 593 LAS VEGAS, NV | MAY 2021 | TS |
| VISN 9: VA Mid South Healthcare Network | 596B10077 | XR US | 596 LEXINGTON, KY | MAY 2021 | BM |
| VISN 22L: VA Desert Pacific Healthcare Network | 600B10071 | RT LINAC | 600 LONG BEACH, CA | MAY 2021 | BM |
| VISN 9: VA Mid South Healthcare Network | 603B13030 | WORKSTATION | 603 LOUISVILLE, KY | MAY 2021 | BM |
| VISN 12: VA Great Lakes Health Care System | 607B10007 | XR US | 607 MADISON, WI | MAY 2021 | BM |
| VISN 21: VA Sierra Pacific Network | 612B18001 | XR PORT C-ARM | 612 MATHER, CA | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19001 | XR PORT C-ARM | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19002 | XR US | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19003 | XR US | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19005 | XR RAD | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19006 | XR RAD | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 23: VA Midwest Health Care Network | 618B19007 | XR RAD | 618 MINNEAPOLIS, MN | MAY 2021 | TS |
| VISN 19: VA Rocky Mountain Network | 623B10037 | XR US | 623 MUSKOGEE, OK | MAY 2021 | BM |
| VISN 19: VA Rocky Mountain Network | 623B10038 | NM BONE DEN | 623 MUSKOGEE, OK | MAY 2021 | BM |
| VISN 19: VA Rocky Mountain Network | 623B10040 | XR US | 623 MUSKOGEE, OK | MAY 2021 | BM |
| VISN 9: VA Mid South Healthcare Network | 626B10032 | XR US | 626 NASHVILLE, TN | MAY 2021 | BM |
| VISN 9: VA Mid South Healthcare Network | 626B13021 | XR US | 626 NASHVILLE, TN | MAY 2021 | BM |
| VISN 16: South Central VA Health Care Network | 629B11015 | XR PORT RAD | 629 NEW ORLEANS, LA | MAY 2021 | TS |
| VISN 16: South Central VA Health Care Network | 629B11016 | XR C-ARM | 629 NEW ORLEANS, LA | MAY 2021 | TS |
| VISN 16: South Central VA Health Care Network | 629B11017 | XR US | 629 NEW ORLEANS, LA | MAY 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 630B10013 | XR US | 630 NEW YORK, NY | MAY 2021 | BM |
| VISN 2: VA Healthcare Network Upstate New York | 630B10017 | XR US | 630 NEW YORK, NY | MAY 2021 | BM |
| VISN 23: VA Midwest Health Care Network | 636B14015 | XR US | 636 IOWA CITY, IA | MAY 2021 | BM |
| VISN 23: VA Midwest Health Care Network | 636B14016 | XR US | 636 IOWA CITY, IA | MAY 2021 | BM |
| VISN 23: VA Midwest Health Care Network | 636B14017 | XR RAD | 636 IOWA CITY, IA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 637B10021 | XR US | 637 ASHEVILLE, NC | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 637B10022 | XR C-ARM | 637 ASHEVILLE, NC | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14011 | XR US | 640 PALO ALTO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14012 | XR US | 640 PALO ALTO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14013 | XR RAD | 640 PALO ALTO, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 640B14014 | XR RAD | 640 PALO ALTO, CA | MAY 2021 | TS |
| VISN 22G: VA Southwest Health Care Network | 644B18081 | XR US | 644 PHOENIX, AZ | MAY 2021 | BM |
| VISN 22G: VA Southwest Health Care Network | 644B18082 | XR US | 644 PHOENIX, AZ | MAY 2021 | BM |
| VISN 22G: VA Southwest Health Care Network | 644B18086 | WORKSTATION | 644 PHOENIX, AZ | MAY 2021 | BM |
| VISN 22G: VA Southwest Health Care Network | 644B18093 | XR US | 644 PHOENIX, AZ | MAY 2021 | BM |
| VISN 22G: VA Southwest Health Care Network | 644B18094 | XR US | 644 PHOENIX, AZ | MAY 2021 | BM |
| VISN 4: VA Stars & Stripes Healthcare Network | 646B12004 | XR US | 646 PITTSBURGH, PA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 648B12168 | XR US | 648 PORTLAND, OR | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 652B00025 | XR US | 652 RICHMOND, VA | MAY 2021 | BM |
| VISN 6: VA Mid-Atlantic Health Care Network | 658B10001 | XR US | 658 SALEM, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 658B10002 | XR US | 658 SALEM, VA | MAY 2021 | TS |
| VISN 6: VA Mid-Atlantic Health Care Network | 659B10004 | XR PORT RAD | 659 SALISBURY, NC | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 662B10006 | XR C-ARM | 662 SAN FRAN, CA | MAY 2021 | TS |
| VISN 21: VA Sierra Pacific Network | 662B10009 | XR US | 662 SAN FRAN, CA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13000 | XR US | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13001 | XR US | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13002 | XR PORT C-ARM | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13003 | XR US | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13004 | XR PORT RAD | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13005 | XR US | 663 SEATTLE, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 663B13008 | XR MRI | 663 EVERETT, WA | MAY 2021 | TS |
| VISN 20: VA Northwest Network | 668B13000 | XR HYBRID | 668 SPOKANE, WA | MAY 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 674B00180 | XR CT | 674 WACO, TX | MAY 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 674B00191 | XR RAD | 674 TEMPLE, TX | MAY 2021 | TS |
| VISN 8: VA Sunshine Healthcare Network | 675B17003 | XR MAMMO | 675 ORLANDO, FL | MAY 2021 | BM |
| VISN 8: VA Sunshine Healthcare Network | 675B17004 | XR MAMMO | 675 ORLANDO, FL | MAY 2021 | BM |
| VISN 12: VA Great Lakes Health Care System | 676B10002 | XR US | 676 TOMAH, WI | MAY 2021 | BM |
| VISN 4: VA Stars & Stripes Healthcare Network | 693B10002 | XR US | 693 WILKES-BARRE, PA | MAY 2021 | TS |
| VISN 17: VA Heart of Texas Health Care Network | 740B10029 | XR MRI | 740 HARLINGEN, TX | MAY 2021 | TS |
| VISN 10: VA Healthcare System of Ohio | 757B19013 | XR US | 757 COLUMBUS, OH | MAY 2021 | BM |
| VISN 10: VA Healthcare System of Ohio | 757B19014 | XR US | 757 COLUMBUS, OH | MAY 2021 | BM |
Q&A
| VISN | Requisition Number | QUESTIONS | ANSWERS |
| NAC-HTME | 01_Due Date | Would the VA please consider an extension to the August 11th closing date of 2 - 3 weeks. | Deadline for Vendor to submit quotes to NAC extended to 8/25/21 |
| NAC-HTME | 02_Several | Several POs in Zip folders missing from schedule of line items/shouldn't be included. | Please see POs removed tab for those POs that were removed, due to being inadvertently added to zip folders. |
| VISN 07 | 509B12003 | Line 51. Medrad CT Injector interface - Is this just adding an interface computer and software for an existing injector or is this requesting adding a new injector? If this request is for a new injector, would you want floor or ceiling mount? |
Line 54. Intego System - Do you want this to also include the following?
*Radiation Dose Management System (RDMS) software:
* WLIST: Wireless worklist import. Will import patient worklist detail into Intego system.
* IRPACS: Wireless infusion export detail. Will export patient infusion detail to PACS system.
* RDPACS: Wireless Radiation Dose export. Will export patient radiation dose detail to PACS system.
Line 57 CT Simulation Lasers - What color light would you like (Red, Green or Blue)? Do you have a preference for LAP DoradoNova or Gammex? Wall, Bridge or Post installation? Does Movable indicate “Laser Steering” such as in Radiation oncology CT SIM applications? (Green, LAP, Wall is common). Line 51: No additional Injector, only one which comes with the system.
Line 54: Yes Line 57: No CT Sim Lasers are needed as this function will never be used at CNVAMC VISN 07 509B12003 1. Unit physical specifications
d. patient portal diameter Question: please change to 70 cm to allow for full competition
f. minimum 360 degree rotation time Question: please change to .35 s to allow for full competition
h. number of crystals Question: please change to 24,000 pixel crystals to allow for full competition
i. minimum size of crystals Question: please change to 3.95mm x 5.3mm crystal dimension to allow for full competition
m. minimum time of flight Question: please change to 385 psec to allow for full competition
t. coincidence timing window Question: please change to 5.25ns to allow for full competition
v. required reconstruction algorithms Question: “FORE” is a Siemens trade name. Please specifiy “Fore algorithm”
w. X-Ray generator output Question: Please change to 72kw to allow full competition
x. mA range Question: please change to 600mA to allow full competition 1. d. 70cm
f. .35s
h. 24,000 pixel crystals
i. 3.95 mm x 5.3 mm
m. 385 psec.
t. 5.25ns
v. Fore Algorithm
w. 72kw
x. 600mA
VA Requirement: System must be able to perform auto QC.
Trade In included: Biograph PET-CT
| VISN 07 | 509B12008 | XR US Augusta | ||
| 1. | Request 5.f. Recommend removal of sterilization trays. Most probes are disinfected via Trophon, which doesn't require a tray. | |||
| 2. | Request 8.a. Frequency Range 2 - 8 MHz be revised to 3 - 8 MHz. | |||
| 1. | Sterilization trays – Our facility prefers to use methods other than Trophon. Trays requested. | |||
| 2. | Frequency change – accept recommendations to change to 3-8 MHz | |||
| VISN 08 | 516B10010 | • | Benchmark requests (1) detector for the wall stand and (1) detector for table each at a size of 17” x 17”. Benchmark also requests an additional (2) wireless detectors be provided at size of 14”x17” and 10” x 12” not including the detectors in the table and wall stand. | |
| - | Are 4 total detectors being requested as minimum (1) 17”x17”, (1) 17”x17”, (1) 14”x17” & (1) 10”x12”? | Only (1) wireless detector needed for table 14”x17” and one wireless detector needed for wall stand 14”x17” | ||
| VISN 08 | 516B10010 | 1c - Please confirm that a total of 4 detectors are required. |
1c - Please confirm desired detector sizes are 17"x17" (qty 2), 14"x17" (qty 1), 10"x12" (qty 1) 2e - Please remove; unduly restrictive, as only 1 vendor has 4.0 lp/mm 2f - Please remove; unduly restricive, as only 1 vendor has 100um detector 2k - Please reduce to 700 lbs 2n - please provide justification for system weighing less than 700kg 1c Total of two 14”x17” detectors needed 2e- Specification removed per request 2f- Specification removed per request 2k- reduced to 700 lbs
| VISN 08 | 516B10010 | i. Minimum image plate height from floor [in] 2", assuming in reference to detector in wall bucky, may not be possible due to distance between bottom of bucky and bottom of detector within bucky. | |
| i. distance is from detector in wall bucky to floor. Change minimum to 2.7” | |||
| VISN 08 | 516B10010 | Section 2. Unit specifications j. minimum tilt of imaging surface in degrees +130/-175. Question: Please clarify this specification as it does not seem to indicate standard usage of the system. | |
| Section 2. Unit specifications i. Minimum image plate height from floor. Question: Please change specification to 2.7 inches to allow full competition. | Sect 2- Specification removed |
Sect 2- i. Change specification to 2.7 inches.
| VISN 08 | 516B10010 | • | Any additional grips or positioning devices needed for Table or Wallstand? | |
| • | Describe location of existing GE DR room needing trade-in (floor level, distance to freight elevator, distance to loading dock for removal) | No standard devices should come with the unit. |
Existing equipment for trade-in GE (2nd floor) Distance to freight elevator-not available Distance to loading dock-not available
| VISN 17 | 519B10010 | 1) | What type of cardiac software is “preferred”. (not listed in the technical specifications). | |
| 2) | What type of collimators are required to perform your daily exams? (not listed in the technical specifications). Example : Low Energy High Resolution, pinhole, High Energy and Medium Energy General Purpose collimators. | |||
| 3) | PET/SPECT Brain software is listed in the technical specifications. Are PET normalized databases required? Are normalized databases for SPECT required? Ictal/Interictal comparison? | |||
| 4) | The type of collimator changers are not listed or described in the technical specifications. Are automatic collimator changers required? | |||
| 5) | The type of quality control is not listed or described in the technical specifications. Is automatic quality control required? | 1. | Cedars. This has been amended in the tech specs) | |
| 2. | Requesting low energy, high resolution, parallel hole collimator(included in tech specs). Energy range information please see 1f | |||
| 3. | Vendor list all offerings | |||
| 4. | Automatic collimator changers are preferred. Technical specifications modified for clarity | |||
| 5. | Automatic quality control. Tech specs modified for clarity | |||
| VISN 17 | 519B10010 | 6h. Touch Screen Patient positioning monitor. Q: Please change spec to, "in room patient positioning to allow for fair competition. | 6h. Requirement to remain as is for workflow requirements | |
| VISN 04 | 529B12012 | XR US Butler | ||
| 1. | Request 5.k. Recommend removal or move to Value Add of Ability to enter standby mode or sleep mode. | |||
| Specification was removed and updated tech spec/compliance matrix has been provided | ||||
| VISN 20 | 531B10148 | 1. | Benchmark is vender specific; is it the expectation for OEM to post process for Dual Energy GE CT scanners? | |
| 2. | Benchmark is vender specific ; is the VA looking for another OEM to interface with their current post processing software? | 1) No. |
2) There are currently no plans to interface the bone densitometry system with the post processing software.
| NAC-HTME | 534B12036 | Please confirm this line item has been removed from the consolidation. | Was NOT part of consolidation. | |
| VISN 12 | 537B18004 | XR US Chicago, IL | ||
| 1. | Request 7.j., k., and l. Be removed or moved to Added Value as these are Advanced Cardiac features. | |||
| 2. | Can you verify items 9.a., b. and c. are required for each system? | |||
| 1. This is a clinical requirement and will not be removed. 2. This is a clinical requirement and features will not be removed. | ||||
| VISN 07 | 544B12028 | Q: Does this obligation include a Tade-In? If so, please can you please provide the system specifications (ie, make, model, serial number) | Yes for trade-in / demolition. Existing system is a Philips Skylight gamma camera system, S/N: K06030067. First installed 4/12/2006 | |
| VISN 08 | 546B10017 | Q: Please confirm the types of cardiac studies you will be performing (ie, calcium scoring, CCTA, etc) |
Added Value, e. Lithium batteries. Question: please provide an explanation for “lithium batteries” or remove this line as it is not applicable to the computed tomography system requested in the obligation.
4. Advanced applications Question: Since existing AW Server with thin client will be used on VA computer, will you require that all vendors quote new post-processing software licenses for the advanced applications listed (CT Fluoroscopy, CT Perfusion, CT Cardiac, Lung CAD, CT Colonoscopy)? Types of studies that will be performed are CCTA, Coronary Disease, Calcium Score, TAVR, Atrial Appendage, Prospective and Retrospective studies.
Removed Lithium batteries line.
4. Answer; Depending on the vendor selected.
| VISN 19 | 554B10001 | 1. | Per specification line 1A under technical requirements, please modify the minimum detector size to 30 cm x 30 cm to allow for competition. | |
| 2. | Per specification line 20 under technical requirements, please modify the minimum range to start at 100 degrees to allow for competition. | |||
| 3. | Per specification line 26 under technical requirements, please modify the range to 645-880 lbs to allow for competition. | |||
| 4. | Per specification line 28 under technical requirements, please modify the range to Minimum matrix pixels 1280 pixels x 1024 to maximum 1548x1524 pixels to allow for competition. | 1. Current specification provides more coverage and numerous c-arms meet spec. |
2. Current spec provides needed flexibility in operating field and numerous c-arms meet spec.
3. Lighter weight increases mobilities in smaller operating space and numerous c-arms meet spec.
4. Current minimum matrix size provides necessary resolution and numerous c-arms meet spec.
| VISN 02 | 561B10008 | Is there a trade-in for this obligation number and if so, what is the model, serial number and probe types being traded in? | Yes, GE LOGIQ P5 SN 2615SU2 | |
| VISN 02 | 561B10008 | Is Windows 10 a requirement for the operating system upon delivery? | A supported operating system is required. | |
| VISN 02 | 561B10008 | What is the intended clinical use for this system? | Intended use – Prostate biopsies and lesion visualization in Surgical Service Urology Section | |
| VISN 02 | 561B10008 | Is the power assist functionality to help with quick access to system use or to be used as battery powered operation? | Quick Access to system. Functionality allows for moving the system without a complete system shutdown and boot up power cycle. | |
| VISN 02 | 561B10015 | 1. | Is this for an “in room conventional RF system” (with spot film device) or a “remote RF system”? | |
| a. | If this is a remote room, all manufactures have a fixed detector in the table and the VA is asking for a wireless table detector. |
| 2. | Regarding the DR panel configuration,, please answer these questions |
| a. | How many DR panels are you requesting? |
| b. | What size DR panel do you want in the wallstand |
| c. | If this is a remote system you are requesting,, if so,, our system comes with 17x17 fixed DR panel in the table |
| d. | Do you want a wireless DR panel? |
| i. | If so, I assume 14x17 size,, correct? |
| 3. | 1f. – should that be “what is the maximum patient table load capacity”? |
| a. | No vendor has a minimum weight capacity of 700 lbs. |
4. 1p. Minimum SID range [in] "40 – 72 – are you referring to the radiographic tube to wallstand or the Fluoro tower to table SID?
| 5. | 1t. – are you referring to a Fluoro Minimum Imaging Tower field of view [in] 4 x 4? |
| a. | I do not believe any vendor offers a 4 x 4 Fluoro size. |
6. 2cc. - Ability to capture live video,,, are you referring to a 3rd party video capture system i.e. TIMS or are you looking for an inherent video functionality like “cine loop”
| 7. | 2n. – please clarify “Removable lead curtains with counterweights for when curtains are not in use”. |
| a. | Are you referring to a mobile shield? If this is a remote room system, no vendor has removable lead curtains. Lead curtains are for the Fluoro spot film device. |
8. 2pp. – you are asking for removable table detector but all vendor’s remote systems come with a fixed table detector. Are you referring to a wireless detector for tabletop and stretcher work?
9. 2rr. – how many batteries do you want in total? Our wireless DR panel comes with 2 batteries standard.
10. 2w. – what is “automatic tube crane, protocol-based movement”?
11. 2z. - can you clarify what the “Quality control tracking package” is?
12. Do you want a compression cone built into the system? 1. Remote RF System
a. VANJ is asking for a wireless detector in addition to a fixed table detector. There are manufacturers in the marketplace that have wireless detectors.
2. a. 2 DR panels
b. Minimum 14 x 17
c. This is not a question.
d. Yes
i. Yes
3. 1f. Yes, Tech Specs is adjusted. The maximum weight capacity is now 700 lbs. There is no minimum weight capacity.
4. 1p. Referring to the radiographic tube to wall stand.
5. 1t. Updated Tech specs to be optional for Minimum Imaging Tower FOV
6. 2cc. Inherent video functionality like “cine loop”
7. 2n. Tech Specs have been updated
a. Not all procedures will be done exclusively remote. There is a requirement for lead shields.
8. 2pp. Yes
9. 2rr. Minimum of 2 batteries
10. 2w. Pre-programmed undulation, automatic positioning based on preset protocols
11. 2z. Tracks repeat analysis which is a requirement
12. This is not a requirement
| VISN 02 | 561B10015 | a. | We need more information about the trade-in: does it have an existing wireless detector, KW, monitor suspension, WS, etc.? | |
| b. | There seems to be an inconsistency between the tech specs and the compliance matrix. We can’t determine if the customer wants remote or nearby. For configuration they have floor mount or overhead. This would be for a RAD room, not the RF which is the request here. So as you read more into the RFO…it does make mention of the to Removable lead curtains with counter weights for when curtains are not in use -- which would be a nearby room. But in the excel, it makes reference to config as an overhead tube but also further down makes the same mention of the lead curtains which would never be on a remote system. Please clarify the requirement. | a. Trade in has all except WS. |
b. The requirement for the system is to have remote functionality but still include lead acrylic shield on ceiling suspension for all procedures. Not all procedures will be done remotely.
| VISN 02 | 561B10015 | Please change benchmark for items 1.a. and 1.b to allow 14x17" detectors for competition | Tech Specs adjusted |
| VISN 02 | 561B10015 | Please change benchmark for item 1.f to 600 lbs to allow for competition | Maximum weight capacity is 700 pounds. Minimum weight capacity is removed. |
| VISN 02 | 561B10015 | For item 1.p., is this SID related to the overhead tube or tube to the wall stand? | Tube to wall stand |
| VISN 02 | 561B10015 | For item 1.q and 1.r, please change measurement to 19 inches to allow for competition | Minimum monitor size to remain. |
| VISN 02 | 561B10015 | Please remove benchmark 2.w. to allow for competition | Specification is a clinical requirement and will not be removed. |
| VISN 02 | 561B10015 | Please adjust benchmark 2.ddd to 26 inches to allow for competition | Specifications for monitor sizes are the minimum requirements |
| VISN 02 | 561B10015 | Is this request for a remote fluoroscopy control room or tableside fluoroscopy control? | Request is for both. |
| VISN 02 | 561B10015 | 1. Unit physical specifications |
a. Minimum wall detector size [in x in]: 15 x 15. Question: Please adjust specification to be 14x17 to allow for fair competition
p. Please update to 45 – 71 inches (115 cm to 180 cm) to allow for fair competition.
v. Please adjust spec to 18.5" to allow for fair competition.
2. Additional specifications
hh. Auto tracking of the tube and detector during vertical adjustment at the wall stand and/or tube. Question: Please remove or make this requirement optional to allow for fair competition.
ii. Tilt-able detector holder with inherent grid (-20 degrees to 90 degrees). Question: Please remove or make this requirement optional to allow for fair competition.
ddd. Ability to lower tabletop height to at least 18” from the floor in the horizontal position at stated maximum weight capacity. Question: Please increase to 19” to allow for fair competition.
3. Software Requirements
a. Monitoring of Dosage- The system must be able to supply a technology that allows for monitoring and tracking of radiation dose provided to a patient. DAP accuracy must be 99% reproducible with <2% deviation.
Question: Please change this specification to <5% deviation to allow for full competition. 1. a. Technical specifications adjusted to 14x17 for competition
b. Minimum is 72 inches and technical specifications to remain as is.
c. Specification is a clinical requirement and will not be adjusted.
2. hh. Specification is a clinical requirement and will not be removed
ii. Adjusted technical specification to optional ddd. Minimum monitor size to remain
3. a. Updated technical specification to <5% deviation.
VISN 06 565B10026 1a - To ensure maximum competition, please change to 32kW 1d - To ensure maximum competition, please change to 0.1 - 320. There are no portable x-ray exams that should ever require 500 mAs given the efficiency (DQE and TQF) of today's detectors.
1f - To ensure maximum competition, please change to 130cm 1h - Please remove specification; given most, if not all, vendors have detector charging mechanisms either on the cart, within the detector bins, or via external chargers, battery life without recharge is not a necessary spec. Additionally, site has included spec 2g, which requires detector battery charging on the unit. Further, as all batteries degrade over time, the accuracy of this specification cannot be verified, nor are kVp and mAs consistent throughout all exposures, making the spec obsolete.
1j - To ensure maximum competition, please change to -90/+90. Only one vendor features -90/+180.
11 - To ensure maximum competition, please change to 19" or less.
1m - Please clarify or remove. Is this width, height, length?
2o - To ensure maximum competition, please remove this requirement; only one vendor features this software. 1a – reduced to 32 kw 1d changed to 0.1 – 320 1f changed to 130cm 1h reduced to one hour 1j changed to -90/+90 1L changed to 19” 1m 22 inches wide 2o tube and detector alignment is up and coming and is a clinical requirement VISN 06 565B10026 1. Unit Physical specifications, a. High frequency generator power [kW]: 40. Is there a clinical reason this requirement is 40kW? Can this be changed to 32 kW as this appears to be a lock out spec eliminating other vendors from competing.
Same section, c. Tube maximum kVp [kVp]: 150 kVp. Is there a clinical reason this is set to 150 kVp? Can this be amended to 133 kVp? This appears to be a lockout spec eliminating other vendors from competing.
Same section, d. Generator mAs range [mAs]: 0.1 - 500. Is there a clinical reason this is set 0.1 - 500 mAs? This appears to be a lockout spec eliminating other vendors from competing. Can this be changed to 50-400mAs?
Same section, f. Maximum tube reach: 134cm. Is there a clinical reason this is set at 134cm? Can this be changed to 120 cm? This appears to be a lockout spec eliminating other vendors from competing.
Same section, i. Maximum charge time from no charge [hr]: 4 hrs. Is there a clinical reason this is set at 4 hrs charge time? This appears to be a lockout spec eliminating other vendors from competing. Can this be changed to 8 hours?
Same section, l. Minimum display monitor size [in]: 21 inches. Can this be changed to 19 inches? This appears to be a lockout spec eliminating other vendors ffrom competing.
New section - 2. System wireless detector configuration, e. Maximum pixel size [µm]: 140. Is there a clinical reason this is set at 140? This appears to be a lockout spec eliminating other vendors from competing. Can this be changed to 150?
Section 2. System wireless detector configuration, f. Minimum detector battery life without recharge [exposure/hour]: 500/4. Can this be updated/changed to 400/4? This appears to be a lockout spec favoring one vendor and eliminating other vendors from competing. 1a – reduced to 32 kw 1c remains 150kVp.
some service calibrations go past clinical use.
1d changed to 0.1 – 320 1f changed to 130cm 1h reduced to one hour 1i increased to 8 hours 1j changed to -90/+90 1L changed to 19” 1m 22 inches wide 2e increased to 150 2f changed to 400/4 VISN 06 565B10026 1. Unit physical specifications
l. minimum display monitor size – Question: please change to minimum display size of 19 inches to allow for full competition
a. high frequency generator power – Question: please change to 30kw generator to allow for full competition
b. Generator kv range – Question: please change to 50-125 kV to allow for full competition
f. maximum tube reach – Question: please change to 110 cm to allow for full competition
i. maximum time from no charge – Question: please change to 4.5 hours to allow for full competition
3. Additional Specifications
c. collapsible column – Please make collapsible column optional to allow full competition
l. display monitor at tube head – Question: please make optional to allow for full competition
o. tube and detector alignment software – Question: please make optional to allow for full competition 1a – reduced to 32 kw 1b changed to 40 -125 1c remains 150kVp.
some service calibrations go past clinical use.
1d changed to 0.1 – 320 1f changed to 130cm 1h reduced to one hour 1i increased to 8 hours 1j changed to -90/+90 1L changed to 19” 1m 22 inches wide 2e increased to 150 2f changed to 400/4 VISN 21 570B10020 Specifications Matrix Questions i. and j. refer to a fixed detecor. Is this what is intented for the portable DR or do you want a wireless detector? Question does not pertain to 570B10020.
In 570B10020 under Physical specifications requirement (i) refers to Minimum SID while requirement (j) refers to Minimum tube reach. Under Additional Specifications requirement (i) refers to Ability to move portable from tube head and requirement (j) refers to Ability to charge detector battery on unit.
No changes will be made.
| VISN 21 | 570B10020 | 3y - To ensure maximum competition, please remove this requirement; only one vendor features this software. | 3y. Request denied. More than one vendors meet this requirement. |
| VISN 21 | 570B10020 | 1b. generator kv range of 40-150, is 40-133kv acceptable? 1n and 2e. max pixl size of 139x139, is 150micron acceptable as 150 mis the standard and not restrictive with no loss of image quality? |
1q. focal spot size .6-1.2, is .7-1.3 acceptable?
3k. trickle charge method for detector battery, we provide 2 slot battery chargerand extra batteries, is this acceptable if you are referring to on board trickle charging of detector? 1b. Range is acceptable. Tech spec revised.
1n & 2e Range is acceptable. Tech spec revised.
1q. Range is acceptable, Technical specifications have been updated, range increased.
3k. Request Denied, this is not an acceptable option.
VISN 21 570B10020 1a. Please change generator frequency to 30kw to allow fair competition.
1b. Please update generator KV range to 50-125KV to allow fair competition 1g. Please update to 170 GB to allow fair competition 1p. Please update to 140KUH to allow fair competition 1s. please update collimation light field accuracy to 2% to align with industry standards 1v. Please update display monitor to 15” to allow fair competition
2f. update detector distributor limit to 330 to allow fair competition
3d. Please make collapsible column optional to allow fair competition 3i. Please make ability to move portable from tube head to optional as there are safety concerns 3u. Please make Display Monitor on Tube Head optional to allow fair competition 3y. Is the goal to provide proper patient alignment with tube and detector? If so, please revise requirement to read “vendor to provide quality assurance software to maintain proper patient positioning with tube and detector” 1a. Request accepted. Technical specification has been updated.
1b. Request denied. More than one vendors meet this requirement.
1g. Request denied. More than one vendors meet this requirement.
1p. Request accepted. Technical specification has been updated.
1s. Request accepted. Technical specification has been updated.
1v. Request accepted. Technical specification has been updated.
2f. Request denied. More than one vendors meet this requirement.
3d. Request denied. More than one vendors meet this requirement.
3i. Request denied. More than one vendors meet this requirement.
3u. Request denied. More than one vendors meet this requirement.
3y. Request denied. More than one vendors meet this requirement.
| VISN 08 | 573B15035 | XR US NFSG VHS - Jacksonville OPC | |
| 1. | Request 1.a. Recommend minimum screen size 24in be revised to 22in | ||
| 2. | Request 1.c. Recommend minimum number of split screens from 6 be revised to 2. | ||
| 3. | Request 1.d. Recommend minimum monitor rotation from 360 degrees be revised to 90 degrees either direction. | ||
| 4. | Request 1.f. Recommend monitor height range from floor 17 inches be revised to 55-66 inches. | ||
| 5. | Request 1.g. Recommend minimum control panel rotation 180 degrees be revised to 130 degrees. | ||
| 6. | Request 1.j. Recommend minimum battery life 30 minutes be revised to 20 minutes. | ||
| 7. | Request 1.i. Recommend maximum equipment weight 230 lbs be revised to 280 lbs. | ||
| 8. | Request 8.b. Description is vendor specific- recommend “10-3v curved array transducer (11.5 curvature, 130-degree FOV)" be revised to "Intracavity Array Transducer". | ||
| 9. | Request 8.c. Description is vendor specific- recommend “12-3 linear array transducer (ERGO)" and Frequency Range 12 - 3 MHz be revised to "Linear Array Transducer" and Frequency Range 11 - 3 MHz. | ||
| 10. | Request 8.d. Description is vendor specific- recommend “18-4 EMT ultra-broadband, linear multi-row array transducer" and Frequency Range 18 - 4 MHz be revised to "Linear Array Transducer” and Frequency Range 16 – 4 MHz. | 1. | Revised per request. |
| 2. | Revised per request. | ||
| 3. | Revised per request. | ||
| 4. | Revised per request. | ||
| 5. | Revised per request. | ||
| 6. | Revised per request. | ||
| 7. | Revised per request. | ||
| 8. | Revised per request. | ||
| 9. | Revised to - Linear Array Transducer. Original frequency of 12-3 MHz remains | ||
| 10. | Revised to – Linear Array Transducer. Original frequency of 18-4 MHz remains. | ||
| VISN 08 | 573B15036 | XR US NFSG VHS - Jacksonville OPC | |
| 1. | Request 1.a. Recommend minimum screen size 24in be revised to 22in | ||
| 2. | Request 1.c. Recommend minimum number of split screens from 6 be revised to 2. | ||
| 3. | Request 1.d. Recommend minimum monitor rotation from 360 degrees be revised to 90 degrees either direction. | ||
| 4. | Request 1.f. Recommend monitor height range from floor 17 inches be revised to 55-66 inches. | ||
| 5. | Request 1.g. Recommend minimum control panel rotation 180 degrees be revised to 130 degrees. | ||
| 6. | Request 1.j. Recommend minimum battery life 30 minutes be revised to 20 minutes. | ||
| 7. | Request 1.i. Recommend maximum equipment weight 230 lbs be revised to 280 lbs. | ||
| 8. | Request 8.b. Description is vendor specific- recommend “10-3v curved array transducer (11.5 curvature, 130-degree FOV)" be revised to "Intracavity Array Transducer". | ||
| 9. | Request 8.c. Description is vendor specific- recommend “12-3 linear array transducer (ERGO)" and Frequency Range 12 - 3 MHz be revised to "Linear Array Transducer" and Frequency Range 11 - 3 MHz. | ||
| 10. | Request 8.d. Description is vendor specific- recommend “18-4 EMT ultra-broadband, linear multi-row array transducer" and Frequency Range 18 - 4 MHz be revised to "Linear Array Transducer” and Frequency Range 16 – 4 MHz. | 1. | Revised per request. |
| 2. | Revised per request. | ||
| 3. | Revised per request. | ||
| 4. | Revised per request. | ||
| 5. | Revised per request. | ||
| 6. | Revised per request. | ||
| 7. | Revised per request. | ||
| 8. | Revised per request. | ||
| 9. | Revised to - Linear Array Transducer. Original frequency of 12-3 MHz remains | ||
| 10. | Revised to –Linear Array Transducer. Original frequency of 18-4 MHz remains | ||
| VISN 08 | 573B15038 | XR US Gainesville | |
| 1. | Request 1.h. Recommend minimum image storage 1TB be revised to 0.5TB | ||
| 2. | Request 1.i. Recommend minimum battery life 30min be removed. | ||
| 3. | Request 1.j. Recommend extended battery life 60min be removed. | ||
| 4. | Request 1.l. Recommend maximum equipment weight 230 lbs be revised to 264 lbs. | ||
| 5. | Request 2.n. Description "Micro Flow Imaging" is vendor specific- recommend removal or clinical clarification. | ||
| 6. | Request 5.j. Recommend removal of shear and strain based elastography. Typically this not used in this department. | ||
| 7. | Request 8.a. Recommend Frequency Range 2 - 5 MHz be revised to 2 MHz. | ||
| 8. | Request 8.b. Recommend Frequency Range 8 - 2 MHz be revised to 8 - 3 MHz. | ||
| 9. | Request 8.c. Recommend Frequency Range 5 - 1 MHz be revised to 4 - 1 MHz.. | ||
| 10. | Request 8.d. Recommend Frequency Range 5 - 1 MHz be revised to 4 - 1 MHz. | 1. | Revised per request |
| 2. | Technical Specification to remain as is | ||
| 3. | Technical Specification to remain as is | ||
| 4. | Revised per request | ||
| 5. | Item removed from Technical Specification | ||
| 6. | Item removed from Technical Specification | ||
| 7. | Frequency range remains | ||
| 8. | Frequency range remains | ||
| 9. | Frequency range remains | ||
| 10. | Frequency range remains | ||
| VISN 08 | 573B15039 | XR US NFSG VHS - Lake City | |
| 1. | Request 1.a. Recommend minimum screen size 24in be revised to 22in | ||
| 2. | Request 1.c. Recommend minimum number of split screens from 6 be revised to 2. | ||
| 3. | Request 1.d. Recommend minimum monitor rotation from 360 degrees be revised to 90 degrees either direction. | ||
| 4. | Request 1.f. Recommend monitor height range from floor 17 inches be revised to 55-66 inches. | ||
| 5. | Request 1.g. Recommend minimum control panel rotation 180 degrees be revised to 130 degrees. | ||
| 6. | Request 1.k. Recommend minimum battery life 30 minutes be revised to 20 minutes. | ||
| 7. | Request 1.i. Recommend maximum equipment weight 230 lbs be revised to 280 lbs. | ||
| 8. | Request 8.b. Description is vendor specific- recommend “10-3v curved array transducer (11.5 curvature, 130-degree FOV)" be revised to "Intracavity Array Transducer". | ||
| 9. | Request 8.c. Description is vendor specific- recommend “12-3 linear array transducer (ERGO)" and Frequency Range 12 - 3 MHz be revised to "Linear Array Transducer" and Frequency Range 11 - 3 MHz. | ||
| 10. | Request 8.d. Description is vendor specific- recommend “18-4 EMT ultra-broadband, linear multi-row array transducer" and Frequency Range 18 - 4 MHz be revised to "Linear Array Transducer” and Frequency Range 16 – 4 MHz. | 1. Revised per request. |
2. Revised per request.
3. Revised per request.
4. Removed
5. Revised per request.
6. Revised per request.
7. Revised per request.
8. Revised per request.
9. Revised to - Linear Array Transducer. Original frequency of 12-3 MHz remains
10. Revised to – Linear Array Transducer. Original frequency of 18-4 MHz remains
| VISN 08 | 573B15040 | XR US NFSG VHS - GNV Primary Care OPC | |
| 1. | Request 1.a. Recommend minimum screen size 24in be revised to 22in | ||
| 2. | Request 1.c. Recommend minimum number of split screens from 6 be revised to 2. | ||
| 3. | Request 1.d. Recommend minimum monitor rotation from 360 degrees be revised to 90 degrees either direction. | ||
| 4. | Request 1.f. Recommend monitor height range from floor 17 inches be revised to 55-66 inches. | ||
| 5. | Request 1.g. Recommend minimum control panel rotation 180 degrees be revised to 130 degrees. | ||
| 6. | Request 1.k. Recommend minimum battery life 30 minutes be revised to 20 minutes. | ||
| 7. | Request 1.i. Recommend maximum equipment weight 230 lbs be revised to 280 lbs. | ||
| 8. | Request 8.b. Description is vendor specific- recommend “10-3v curved array transducer (11.5 curvature, 130-degree FOV)" be revised to "Intracavity Array Transducer". | ||
| 9. | Request 8.c. Description is vendor specific- recommend “12-3 linear array transducer (ERGO)" and Frequency Range 12 - 3 MHz be revised to "Linear Array Transducer" and Frequency Range 11 - 3 MHz. | ||
| 10. | Request 8.d. Description is vendor specific- recommend “18-4 EMT ultra-broadband, linear multi-row array transducer" and Frequency Range 18 - 4 MHz be revised to "Linear Array Transducer” and Frequency Range 16 – 4 MHz. | 1. Revised per request. |
2. Revised per request.
3. Revised per request.
4. Removed
5. Revised per request.
6. Revised per request.
7. Revised per request.
8. Revised per request.
9. Revised to - Linear Array Transducer. Original frequency of 12-3 MHz remains.
10. Revised to – Linear Array Transducer. Original frequency of 18-4 MHz remains VISN 08 573B15042 2. Table Requirements
d. motorized – Question: please change to “table facilitates safe and easy patient transport throughout facility” to allow for fair competition.
3. Additional Specifications
t. MRI-compatible injector Ceiling mounted – Question: please remove the requirement for ceiling mount. All commercially available MRI injectors are pedestal mounted 2 d. Requirement remains- no revision necessary 3 t. Requirement remains- no revision necessary VISN 16 580B10031 Can you verify the quantities as two Ultrasound Systems are listed as well as 3 ireless Connections and 5 Batteries. Qty 2 systems.
Each system must be capable of FIPS compliant wireless connectivity. The specific qty and part numbers will vary from manufacturer to manufacturer.
Qty 5 between 2 systems is preferred however, vendor must provide at least 2 per system for minimum quantity of 4.
| VISN 15 | 589B12019 | Is a battery required for the unit? | Yes, a battery is required. | |
| VISN 06 | 590B10005 | XR US Hampton, VA | ||
| 1. | Request 1.c. Recommend minimum control panel rotation from 360 degrees be revised to 90 degrees either direction. | |||
| 2. | Request 8.d. Frequency Range 4 - 18 MHz be revised to Frequency Range 4 – 16 MHz. | 1. Reduced to 180 |
2. Reduced to 4 – 16 Mhz
| VISN 06 | 590B10006 | XR US Hampton, VA | |
| 1. | Request 1.c. Recommend minimum control panel rotation from 360 degrees be revised to 90 degrees either direction. | ||
| 2. | Request 8.d. Frequency Range 4 - 18 MHz be revised to Frequency Range 4 – 16 MHz. | 1. Reduced to 180 |
2. Reduced to 4 – 16 Mhz VISN 06 590B10008 2d - Please remove or provide clarification for why limiting mA at 100 kVp is required. Most systems feature 800 mA at 100 kVp to provide the most flexible techniques.
2e - Please remove or reduce to 150um; unduly restrictive, as only 1 vendor has 100um.
2f - Please reduce to 700 lbs; 881 lbs is incredibly specific and overkill 4c - Please provide the generic requirement for "AutoGrid"; is this equivalent to a non-physical grid 2d changed to 800 2e increased to 150 2f reduced to 600 4c yes, equivalent to a non-physical grid VISN 06 590B10008 2. Unit Physical Specifications d. Maximum radiographic mA exposure @ 100 kVp: 630. Is this a lockout spec? Is there a clinical reason the maximum is set at 630? Can this be updated to 800 so as not to specifically lock out vendors?
Also, same section: 2. Unit Physical Specs, f. Minimum patient table load capacity (static) [lbs]: 881 lbs. What is the clinical reason the minimu patient table load capacity (static) is set at 881 lbs? This appears to also be a lockout spec eliminating other vendors from competing. Can this patient weight limit be changed to 600 lbs?
2d changed to 800 2f reduced to 600 VISN 06 590B10008 4. Software Requirements
a. Stitching – ability to stitch multiple long-axis images with options for automated and manual process. Question: Do you want Stitching on both table and wall stand?
Question: Please provide the following information to determine an accurate trade-in value: Software Level, List of Coils and Channels, Software Level, Year of Installation, and Projected De-Installation date.
Question: Is there a trade-in? Please provide the following information to determine an accurate trade-in value: Year of Installation, number of detectors, and Projected De-Installation date.
Question: Would you like Separate 72” and 40” focus grids included in addition to midrange grid? Stitching on just wall stand NO trade in Just Midrange grid
| VISN 06 | 590B10009 | Technical Requirements,3. Additional specifications, d. Stereotactic/tomosystheis guided biopsy, standard or optional? | Standard |
| VISN 06 | 590B10009 | 3. Additional specifications |
Q: What is your preferred vacuum assisted device?
Q: Do you require a Chair?
Q: Is a device needed for imaging core samples?
5. Security/Connectivity Requirements Q: What is the version of your McKesson PACs and are you planning to read off of PACs? 3. Unsure of what is meant by vacuum assisted device.
No chair Yes to core sample
5. Will read from MRS 8/9 VISN 06 590B10010 2d - Please remove or provide clarification for why limiting mA at 100 kVp is required. Most systems feature 800 mA at 100 kVp to provide the most flexible techniques.
2e - Please remove or reduce to 150um; unduly restrictive, as only 1 vendor has 100um.
2f - Please reduce to 700 lbs; 881 lbs is incredibly specific and overkill 4c - Please provide the generic requirement for "AutoGrid"; is this equivalent to a non-physical grid 2d changed to 800 2e increased to 150 2f reduced to 600 4c yes, equivalent to a non-physical grid VISN 06 590B10010 Question: Do you want 17”x17” Detector (wall stand) included in the quote?
Question: Do you want the wall stand to accommodate the use of a mobile table?
Question: Would you like Separate 72” and 40” focus grids included in addition to midrange grid?
Question: Is there a trade-in? Please provide the following information to determine an accurate trade-in value: Year of Installation, number of detectors, and Projected De-Installation date 14 x 17 only No mobile table Just Midrange NO Trade in
| VISN 21 | 593B10026 | Please provide serial number and manufacturer date for the Ziehm trade-in system. | Serial No. is 20216. Procurement date was 10/01/2021. |
| NAC-HTME | 596B10077 | Not benchmark | Changed to tech spec. |
| VISN 09 | 596B10077 | 1.What department will the system be used in ? |
2.Will this do cardiac procedures?
3.Can you provide a list of studies?
4.Are you setting up demos for vendors? 1. Vascular Lab (Radiology/ Surgery)
2. No
3. See tech specs for study packages
4. No, please provide all applicable brochures and/or videos
| VISN 23 | 618B19005 | 1b(c) - please increase to 8 lbs in order to promote maximum competition | |
| 2k - please clarify what "single-sided suspended table swivel range" means? Why is this a minimum requirement? | 1b(c)-Detector weight can be increased to 8 lbs. |
2k- Cantilever mounted X-ray tabletop that's capable of pivoting between 0 and 90 degrees on the pedestal base. Sometimes referred to as a “Diving Board” table.
Why is this a minimum requirement?
A radiographic Table that pivots 90 degrees is beneficial in an outpatient clinic setting for patients that are not able to ambulate independently and the x-ray table needs to move to adjust for the patient lifting equipment.